Robust Fourier-based slanted-edge method to measure scatter ratio

被引:0
|
作者
Garland, Lisa M. [1 ,2 ]
Cunningham, Ian A.
机构
[1] Western Univ, Robarts Res Inst, Imaging Res Labs, London, ON, Canada
[2] Western Univ, Dept Med Biophys, London, ON, Canada
基金
加拿大自然科学与工程研究理事会;
关键词
Fourier domain; X-ray imaging; X-ray scatter; MTF;
D O I
10.1002/mp.17765
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background: Patient scatter incident on an x-ray detector reduces radiographic contrast and adds quantum noise, and minimizing scatter is critical in some specialized techniques such as dual-energy and energy-subtraction methods. Existing methods to measure scatter are either labor-intensive (multiple disks) or not appropriate to use in radiography where scatter often exceeds the width of the x-ray beam. Purpose: Develop a method to measure the scatter-to-primary ratio (SPR) that can be used for a wide range of radiographic and mammographic conditions, both with scatter equilibrium (scatter function does not exceed primary-beam width) and without. Methods: Fourier theory is used to show the SPR can be measured from the low-frequency drop (LFD) of the Fourier transform of the derivative of a normalized edge profile. The method was validated both experimentally and by simulation for radiography and mammography under scatter equilibrium and nonequilibrium conditions. Results: The theoretical derivation showed that by normalizing an edge profile with a profile without the edge, scatter equilibrium is not required and the method accommodates a nonuniform primary beam from beam divergence and Heel effect. The method was validated by a simulation study for a range of scatter-LSF widths, primary-beam widths, and image regions of interest used in the analysis. Experimental scatter measurements agreed with a similar edge-method published by Cooper when scatter equilibrium is achieved. Conclusions: A simple and direct method of measuring the SPR obtained with both uniform and nonuniform test phantoms is described. Validated both experimentally and theoretically, it uses the Fourier LFD obtained from a normalized slanted-edge profile and works for a wide range of practical mammographic and radiographic conditions.
引用
收藏
页数:14
相关论文
共 50 条
  • [1] Modified slanted-edge method to measure the modulation transfer function of camera
    Zhang, Haosu
    Li, Chuang
    Duan, Yaxuan
    OPTIK, 2018, 157 : 635 - 643
  • [2] A robust Fourier-based method to measure pulse pressure variability
    Acosta, Sebastian
    Ahmed, Mubbasheer
    Yin, Suellen M.
    Brady, Ken M.
    Penny, Daniel J.
    Rusin, Craig G.
    BIOMEDICAL SIGNAL PROCESSING AND CONTROL, 2020, 60 (60)
  • [3] The moving slanted-edge method: A novel technique to measure the temporal response of a fluoroscopic
    Friedman, S. N.
    Cunningham, I. A.
    MEDICAL PHYSICS, 2006, 33 (07) : 2668 - 2669
  • [4] A new slanted-edge MTF calculation method
    Maidment, AD
    Albert, M
    RADIOLOGY, 1999, 213P : 287 - 287
  • [5] A moving slanted-edge method to measure the temporal modulation transfer function of fluoroscopic systems
    Friedman, S. N.
    Cunningham, I. A.
    MEDICAL PHYSICS, 2008, 35 (06) : 2473 - 2484
  • [6] MTF Measurement by Slanted-Edge Method Based on Improved Zernike Moments
    Zhang, Shuo
    Wang, Fengyan
    Wu, Xiang
    Gao, Kangzhe
    SENSORS, 2023, 23 (01)
  • [7] MTF measurement method based on slanted-edge image simulated by normal distribution
    1Changchun Institute of Optics, Fine Mechanics and Physics, Chinese Academy of Sciences, Changchun
    Jinlin
    130033, China
    不详
    100049, China
    Guangxue Xuebao, 7
  • [8] Application of Tatian's method to slanted-edge MTF measurement
    Burns, PD
    Image Quality and System Performance II, 2005, 5668 : 255 - 261
  • [9] An Improved Slanted-edge Method for Measuring Modulation Transfer Function Based on Edge-preserving Filter
    Nie, Saisai
    Ruan, Xiukai
    Cui, Guihua
    Lu, Weikuan
    FOURTEENTH INTERNATIONAL CONFERENCE ON GRAPHICS AND IMAGE PROCESSING, ICGIP 2022, 2022, 12705
  • [10] Robust edge-spread function construction methods to counter poor sample spacing uniformity in the slanted-edge method
    van den Bergh, F.
    JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 2019, 36 (07) : 1126 - 1136