Performance Characteristics of a New Generation 148-cm Axial Field-of-View uMI Panorama GS PET/CT System with Extended NEMA NU 2-2018 and EARL Standards

被引:3
|
作者
Zhang, Haiqiong [1 ]
Ren, Chao [1 ]
Liu, Yu [1 ]
Yan, Xinchun [1 ]
Liu, Meixi [1 ]
Hao, Zhixin [1 ]
Xing, Haiqun [1 ]
Huo, Li [1 ]
机构
[1] Chinese Acad Med Sci, Peking Union Med Coll Hosp, Dept Nucl Med,Beijing Key Lab Mol Targeted Diag &, State Key Lab Complex Severe & Rare Dis, Beijing, Peoples R China
关键词
NEMA; EARL; uMI Panorama GS PET/CT; LAFOV; time of flight; SCANNER;
D O I
10.2967/jnumed.124.267963
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The uMI Panorama GS PET/CT system is a new long-axial-field-of- view scanner featuring high sensitivity, time-of-flight (TOF) resolution, spatial resolution, and count rate performance. The aim of this study is to assess the PET system on the basis of the National Electrical Manufacturers Association (NEMA) NU 2-2018 and European Association of Nuclear Medicine Research Limited (EARL) standards. Methods: Spatial resolution, count rate performance, sensitivity, accuracy, image quality, TOF resolution, and coregistration accuracy were evaluated following the NEMA NU 2-2018 standard. Additional experiments included energy resolution, 200-cm-long line sources for sensitivity, a 175-cm-long scatter phantom for count rate and TOF resolution, as well as the compliance with the EARL guideline. Moreover, an 18 F-FDG PET patient study was reconstructed with various frame durations. Results: The PET system achieved sub-3-mm trans- axial and axial spatial resolutions at a 1-cm radial offset. The sensitivities with the 70-cm-long and 200-cm-long line sources were observed to be 176.3 and 90.8 kcps/MBq, respectively, at the center of the field of view. The noise-equivalent count rates (NECRs) of the 70-cm-long and 175-cm-long scatter phantoms were measured to be 3.35 Mcps at 57.57 kBq/mL and 2.24 Mcps at 33.27 kBq/mL, respectively. The TOF resolutions for both phantoms were approximately 189 ps at 5.3 kBq/mL and lower than 200 ps below the NECR peaks. The absolute count rate errors of all 34 acquisitions were less than 3% below the NECR peak for the 70-cm-long scatter phantom. With the standard NEMA image quality phantom experiment, the contrast recovery coefficient varied from 68.17% to 94.20% and the background variabilities were all below 2%. The maximum PET/CT coregistration error was 1.33 mm. Regarding EARL compliance, the gaussian filter of 5-mm full width at half maximum could produce acceptable images. The patient data demonstrate visually satisfactory image quality with short frames (less than 1 min). Conclusion: The uMI Panorama GS exhibits spatial resolution and TOF resolution similar to those of the uMI Panorama system (35-cm axial field of view), despite the extended axial field of view. The 148-cm axial coverage, sub-200-ps TOF resolution, high sensitivity, and count rate performances are expected to yield superior image quality and offer new opportunities for various clinical applications.
引用
收藏
页码:1974 / 1982
页数:9
相关论文
共 13 条
  • [1] Preliminary evaluation of the new digital uMI Panorama LAFOV PET/CT system with EARL compliance and extended NEMA NU 2-2018
    Zhang, Haiqiong
    Xu, Hancong
    Xing, Hai-Qun
    Liu, Yu
    Ren, Chao
    Wang, Tong
    Wu, Yifan
    Rong, Haoyang
    Dong, Yun
    Huo, Li
    JOURNAL OF NUCLEAR MEDICINE, 2024, 65
  • [2] NEMA NU 2-2018 performance evaluation of a new generation 30-cm axial field-of-view Discovery MI PET/CT
    Zeimpekis, Konstantinos G.
    Kotasidis, Fotis A.
    Huellner, Martin
    Nemirovsky, Alexandra
    Kaufmann, Philipp A.
    Treyer, Valerie
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2022, 49 (09) : 3023 - 3032
  • [3] Voluntary motion management for uMI Panorama GS, a new 148-cm axial FOV PET/CT
    Zhao, Yizhang
    Zhang, Duo
    Li, Yue
    Liu, Hao
    Sun, Chen
    Yi, Hongyuan
    Ye, Qing
    Zhao, Yumo
    Zhu, Yuchun
    Huo, Li
    Lu, Yihuan
    JOURNAL OF NUCLEAR MEDICINE, 2024, 65
  • [4] NEMA NU 2–2018 performance evaluation of a new generation 30-cm axial field-of-view Discovery MI PET/CT
    Konstantinos G. Zeimpekis
    Fotis A. Kotasidis
    Martin Huellner
    Alexandra Nemirovsky
    Philipp A. Kaufmann
    Valerie Treyer
    European Journal of Nuclear Medicine and Molecular Imaging, 2022, 49 : 3023 - 3032
  • [5] Performance Characteristics of the Biograph Vision Quadra PET/CT System with a Long Axial Field of View Using the NEMA NU 2-2018 Standard
    Prenosil, George A.
    Sari, Hasan
    Furstner, Markus
    Afshar-Oromieh, Ali
    Shi, Kuangyu
    Rominger, Axel
    Hentschel, Michael
    JOURNAL OF NUCLEAR MEDICINE, 2022, 63 (03) : 476 - 484
  • [6] NEMA NU 2-2018 evaluation and image quality optimization of a new generation digital 32-cm axial field-of-view Omni Legend PET-CT using a genetic evolutionary algorithm
    Smith, Rhodri Lyn
    Bartley, Lee
    O'Callaghan, Christopher
    Haberska, Luiza
    Marshall, Chris
    BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2024, 10 (02)
  • [7] Performance Evaluation of the uMI Panorama PET/CT System in Accordance with the National Electrical Manufacturers Association NU 2-2018 Standard
    Li, Guiyu
    Ma, Wenhui
    Li, Xiang
    Yang, Weidong
    Quan, Zhiyong
    Ma, Taoqi
    Wang, Junling
    Wang, Yunya
    Kang, Fei
    Wang, Jing
    JOURNAL OF NUCLEAR MEDICINE, 2024, 65 (04) : 652 - 658
  • [8] Performance Characteristics of a New-Generation Digital Bismuth Germanium Oxide PET/CT System, Omni Legend 32, According to NEMA NU 2-2018 Standards
    Yamagishi, Shin
    Miwa, Kenta
    Kamitaki, Shun
    Anraku, Kouichi
    Sato, Shun
    Yamao, Tensho
    Kubo, Hitoshi
    Miyaji, Noriaki
    Oguchi, Kazuhiro
    JOURNAL OF NUCLEAR MEDICINE, 2023, 64 (12) : 1990 - 1997
  • [9] Performance characteristics of the digital uMI550 PET/CT system according to the NEMA NU2-2018 standard
    Shuguang Chen
    Pengcheng Hu
    Yushen Gu
    Haojun Yu
    Hongcheng Shi
    EJNMMI Physics, 7
  • [10] Performance characteristics of the digital uMI550 PET/CT system according to the NEMA NU2-2018 standard
    Chen, Shuguang
    Hu, Pengcheng
    Gu, Yushen
    Yu, Haojun
    Shi, Hongcheng
    EJNMMI PHYSICS, 2020, 7 (01)