Effects of an extended MRI approval of an implantable spinal cord stimulation device on compliance with manufacturer's recommendations

被引:0
|
作者
Reining, Marco [1 ,2 ]
Winkler, Dirk [3 ]
Kirchhof, Klaus [4 ]
Boettcher, Joachim [5 ]
Kretzschmar, Michael [1 ,6 ]
机构
[1] SRH Wald Klin Gera GmbH, Dept Pain Med & Palliat Care, Str Friedens 122, D-07548 Gera, Germany
[2] SRH Mobile Univ, Riedlingen, Germany
[3] Univ Hosp Leipzig, Dept Neurosurg, Leipzig, Germany
[4] SRH Wald Klin Gera GmbH, Dept Diagnost & Intervent Radiol, Gera, Germany
[5] Friedrich Schiller Univ Jena, Jena, Germany
[6] SRH Univ Appl Sci Heidelberg, Heidelberg, Germany
关键词
health services research; implantable neurostimulators; magnetic resonance imaging; patient safety; spinal cord stimulation; EXPLANTATION; SAFETY;
D O I
10.1111/papr.70022
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
ObjectiveThe current approval for all available spinal cord stimulation (SCS) devices set several limitations for magnetic resonance imaging (MRI). As a result, most of the necessary examinations are not possible within the scope of the restrictive approvals and, if legally permitted, must be carried out off-label. This retrospective subgroup analysis investigates if the currently extended MRI approval of one specific device allows more MRI scans to be conducted within the manufacturer's recommendations.Materials and MethodsTechnical MRI data (saved in the DICOM headers) and medical treatment data for all MRI examinations on patients with Proclaim (R) implantable pulse generators (IPG; Proclaim (R) spinal cord stimulation systems, Abbott Laboratories, Plano, TX, USA) were examined. Due to a major change in our standard operating procedures for MRI scans in 2019, the two time periods (before and after 2019) were separately analyzed.ResultsWe identified 62 MRI scans with the IPG. The entire implanted system was approved for MRI examinations in just over 50% of the cases, regardless of old and new approval. Options for lead placement were expanded in the new approval; however, this did not significantly improve the number of MR conditional devices. By contrast, for a higher specific absorption rate, significantly more scans within the recommendations are possible in Period 2 (p = 0.011). However, the number of possible scans did not reach statistical significance in Period 1 (p = 0.078). No device-related adverse events were noted.ConclusionThe new MRI approval is suitable for performing more scans within the manufacturer's specifications. Cervical leads remain problematic because longer leads are required, and the lower impedances inhibit the MRI mode.
引用
收藏
页数:8
相关论文
共 20 条
  • [1] Device manufacturer receives OK to proceed with spinal cord stimulation clinical trial for chronic pain
    Fitzpatrick, Sean
    EXPERT REVIEW OF MEDICAL DEVICES, 2012, 9 (04) : 321 - 321
  • [2] Perioperative Spinal Cord Stimulation Management: A Clinical Scenario of Device Loss and Recommendations for Anesthesiologists
    Hardman, Matthew I.
    Hagedorn, Jonathan M.
    PAIN MEDICINE, 2020, 21 (04) : 865 - 867
  • [3] Flexible and fully implantable upconversion device for wireless optogenetic stimulation of the spinal cord in behaving animals
    Wang, Ying
    Xie, Kai
    Yue, Haibing
    Chen, Xian
    Luo, Xuan
    Liao, Qinghai
    Liu, Ming
    Wang, Feng
    Shi, Peng
    NANOSCALE, 2020, 12 (04) : 2406 - 2414
  • [6] Potential device interaction of a dual chamber implantable cardioverter defibrillator in a patient with continuous spinal cord stimulation
    Schimpf, R
    Wolpert, C
    Herwig, S
    Schneider, C
    Esmailzadeh, B
    Lüderitz, B
    EUROPACE, 2003, 5 (04): : 397 - 402
  • [7] Exploring the Effects of Spinal Cord Stimulation for Gait Disturbances in Parkinson's Disease
    Vanegas-Arroyave, N.
    Kwei, K.
    Lee, M.
    Prakash, C.
    Branco, L.
    Tarakad, A.
    Taneff, Y.
    Ince, N.
    Viswanathan, A.
    Najafi, B.
    MOVEMENT DISORDERS, 2023, 38 : S57 - S58
  • [8] The effects of functional electrical stimulation leg cycle ergometry training on arterial compliance in individuals with spinal cord injury
    Zbogar, D.
    Eng, J. J.
    Krassioukov, A. V.
    Scott, J. M.
    Esch, B. T. A.
    Warburotn, D. E. R.
    SPINAL CORD, 2008, 46 (11) : 722 - 726
  • [9] The effects of functional electrical stimulation leg cycle ergometry training on arterial compliance in individuals with spinal cord injury
    D Zbogar
    J J Eng
    A V Krassioukov
    J M Scott
    B T A Esch
    D E R Warburton
    Spinal Cord, 2008, 46 : 722 - 726
  • [10] Effects of spinal cord stimulation in a patient with Parkinson's disease and chronic back pain
    Weise, D.
    Winkler, D.
    Meixensberger, J.
    Classen, J.
    JOURNAL OF NEUROLOGY, 2010, 257 : S217 - S217