Measuring the Pupillary Light Reflex Using Portable Instruments in Applied Settings

被引:0
|
作者
Gray, Nicola S. [1 ,2 ]
Price, Menna [1 ]
Pink, Jennifer [1 ]
O'Connor, Chris [3 ]
Antunes, Ana [4 ]
Snowden, Robert J. [4 ]
机构
[1] Swansea Univ, Dept Psychol, Swansea SA2 8PP, Wales
[2] Swansea Bay Univ Hlth Board, Bridgend CF31 4LN, Wales
[3] Aneurin Bevan Univ Hlth Board, Newport NP18 3XQ, Wales
[4] Cardiff Univ, Sch Psychol, Cardiff CF10 3AT, Wales
基金
英国工程与自然科学研究理事会; 英国经济与社会研究理事会;
关键词
age; anxiety; parasympathetic nervous system; pupillometry; trauma; HEART-RATE-VARIABILITY; AUTONOMIC NERVOUS-SYSTEM; INFRARED PUPILLOMETRY; DYNAMIC PUPILLOMETRY; STATE/TRAIT ANXIETY; STRESS; DISORDER; AGE; DEPRESSION;
D O I
10.3390/vision8040060
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
The early components of the pupillary light reflex (PLR) are governed by the parasympathetic nervous system. The use of cheap, portable pupillometry devices may allow for the testing of parasympathetic-system health in field settings. We examined the reliability of two portable instruments for measuring the PLR and their sensitivity to individual differences known to modulate the PLR. Parameters of the PLR were measured in a community sample (N = 108) in a variety of field settings. Measurements were taken using a commercial pupillometer (NeuroLight, IDMED) and an iPhone using the Reflex Pro PLR analyser (Brightlamp). The parameters of baseline pupil diameter, constriction latency, amplitude and relative amplitude of constriction, and constriction velocity were measured. Individual differences related to age, levels of anxiety, and post-traumatic stress disorder (PTSD) symptomology were assessed. Some measures could not be attained using the iPhone under these field conditions. The reliability of the measures was high, save for the measurement of contraction latency which was particularly unreliable for the iPhone system. The parameters of the PLR showed the same internal relationships as those established in laboratory-based measurements. Age was negatively correlated with all the reliable PLR parameters for both systems. Effects of anxiety and PTSD symptomology were also apparent. The study demonstrated that a hand-held portable infrared pupillometer can be used successfully to measure the PLR parameters under field settings and can be used to examine individual differences. This may allow these devices to be used in workplaces, sports fields, roadsides, etc., to examine parasympathetic activity where needed.
引用
收藏
页数:11
相关论文
共 50 条
  • [1] A PORTABLE PUPILOMETER SYSTEM FOR MEASURING PUPILLARY SIZE AND LIGHT REFLEX
    RADZIUS, A
    WELCH, P
    CONE, EJ
    HENNINGFIELD, JE
    BEHAVIOR RESEARCH METHODS INSTRUMENTS & COMPUTERS, 1989, 21 (06): : 611 - 618
  • [2] Automated Pupillary Light Reflex Test on a Portable Platform
    Temel, Dogancan
    Mathew, Melvin J.
    AlRegib, Ghassan
    Khalifa, Yousuf M.
    2019 INTERNATIONAL SYMPOSIUM ON MEDICAL ROBOTICS (ISMR), 2019,
  • [3] Effect of locally applied prazosin on the kinetics of the pupillary light reflex
    Mortlock, SA
    Langley, RW
    Bradshaw, CM
    Szabadi, E
    BRITISH JOURNAL OF CLINICAL PHARMACOLOGY, 1996, 42 (02) : 236 - 238
  • [4] rPLR: an imaging system for measuring pupillary light reflex at a distance
    Dinalankara, Dinalankara M. R.
    Miles, Judith H.
    Yao, Gang
    APPLIED OPTICS, 2014, 53 (32) : 7787 - 7795
  • [5] Assessment of pupillary light reflex using a smartphone application
    Shin, Young Duck
    Bae, Jin Ho
    Kwon, Eun Jung
    Kim, Hyeon Tae
    Lee, Tae-Soo
    Choi, Young Jin
    EXPERIMENTAL AND THERAPEUTIC MEDICINE, 2016, 12 (02) : 720 - 724
  • [6] Quantitative pupillary light reflex
    Lehrfeld, L
    ARCHIVES OF NEUROLOGY AND PSYCHIATRY, 1931, 25 (01): : 159 - 162
  • [7] PUPILLARY LIGHT REFLEX IN AMBLYOPIA
    KASE, M
    NAGATA, R
    YOSHIDA, A
    HANADA, I
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 1984, 25 (04) : 467 - 471
  • [8] The pupillary light reflex in migraine
    Harle, DE
    Wolffsohn, JS
    Evans, BJW
    OPHTHALMIC AND PHYSIOLOGICAL OPTICS, 2005, 25 (03) : 240 - 245
  • [9] The pupillary light reflex pathway
    Wilhelm, H
    Kardon, RH
    NEURO-OPHTHALMOLOGY, 1997, 17 (02) : 59 - 62
  • [10] Relation of pupillary light reflex to light intensity
    Antonis Tzambazakis
    Dimitrios Fotiou
    Aggelis Fotiou
    Dimitrios Tsiptsios
    Maria Nakou
    Annals of General Psychiatry, 5 (Suppl 1)