Development of an Ultra-small Mass Spectrometer for Future Lunar and Planetary Exploration

被引:1
|
作者
Kawashima, Oya [1 ]
Saito, Yoshifumi [1 ]
Asamura, Kazushi [1 ]
Sugita, Seiji [2 ]
Kasahara, Satoshi [2 ]
Yokota, Shoichiro [3 ]
Hirahara, Masafumi [4 ]
机构
[1] Japan Aerosp Explorat Agcy ISAS JAXA, Inst Space & Astronaut Sci, Sagamihara, Kanagawa 2525210, Japan
[2] Univ Tokyo, Dept Earth & Planetary Sci, Bunkyo Ku, Tokyo 1130033, Japan
[3] Osaka Univ, Dept Earth & Space Sci, Toyonaka, Osaka 5600043, Japan
[4] Nagoya Univ, Inst Space Earth Environm Res, Nagoya, Aichi 4648601, Japan
基金
日本学术振兴会;
关键词
D O I
10.1109/AERO58975.2024.10521064
中图分类号
V [航空、航天];
学科分类号
08 ; 0825 ;
摘要
Mass spectrometers have been widely employed as payloads for planetary exploration missions. The instruments measure the atomic and molecular compositions in planetary samples and are widely applicable to quantitative/qualitative analysis. Unlike remote sensing, mass spectrometers need to be in close to the target bodies for in-situ analysis, thus developing small mass spectrometers is an important issue in expanding the scope of analytical targets. There are additional possibilities for small mass spectrometers, namely piggyback or secondary payloads for missions. The small mass spectrometers may accompany sample return missions and provide complementary observations (e.g., multi-point; time continuity). In this study, we are developing an ultra-small neutral mass spectrometer of which ion optics fits into a volume of 1U (10x10x10 cm(3)). The mass spectrometer being developed is based on Electron Impact (EI) ionization method and time-offlight (TOF) mass analyzing technique. The EI ion source is newly developed to achieve low power consumption (similar to 0.5 W), high electron emission (>1.0 mA), and high ionization efficiency (similar to 104 nA/Pa for N-2 gas) with small size (similar to 7x3x3 cm(3)). For the mass analysis part, ion trap electrodes are employed to increase TOF while reducing volume. The ion trap electrodes allow ions to pass around the same optics at specific frequency for each m/z, resulting in a small optics size and long TOF. So far, we developed a test model with an overall optics volume of similar to 7x7x8 cm(3) and a weight of similar to 0.3 kg. As for performance with a singlelap, the mass resolution was m/.m similar to 50, the sensitivity was similar to 1x10(-6) (counts/s)/(particle/cc), and the detection limit (i.e., S/N=1) was similar to 1x10(4) particle/cc, respectively. In addition, we have completed a design with an improved mass resolution to obtain mass spectra for more than 2 laps, using a numerical simulation. According to the results of the simulation, the improved mass resolution (m/.m > 1,000) will be achieved with > 100 laps. It should be noted that the new model does not require high voltage pulses; the instrument is expected to be small enough even if the flight-qualified electrical system is included. The timing of ion trap can be arbitrarily determined, making it possible to provide multiple observation modes that take into account the fundamental trade-off between resolution and sensitivity in TOF instruments. Additionally, the new ion source with low power consumption and high emission current is one of the achievements of this work and can be applied to other mass spectrometers as a stand-alone system; the ion source is planned to be used for JAXA/ISRO LUPEX mission.
引用
收藏
页数:9
相关论文
共 50 条
  • [1] The MAss Spectrometer for Planetary EXploration (MASPEX)
    Brockwell, Tim G.
    Waite, J. Hunter
    Lunine, Jonathan I.
    Meech, Karen J.
    Miller, Greg
    Wilson, Paul
    Pickens, Keith
    Roberts, John
    2016 IEEE AEROSPACE CONFERENCE, 2016,
  • [2] Development of mobile robots for lunar/planetary exploration
    Yoshida K.
    Seimitsu Kogaku Kaishi/Journal of the Japan Society for Precision Engineering, 2011, 77 (01): : 12 - 15
  • [3] Permittivity sensor development for lunar and planetary surface exploration
    Gscheidle, Christian
    Witzel, Thilo
    Smolka, Alexander
    Reiss, Philipp
    FRONTIERS IN SPACE TECHNOLOGIES, 2024, 4
  • [4] Mass Spectrometry of ultra-small gold nanoparticles
    Hewitt, Michael
    Johnson, Grant
    Hernandez, Heriberto
    ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 2019, 257
  • [5] DEVELOPMENT OF THE ULTRA-SMALL SATELLITE AND LAUNCH VEHICLE
    HAYASHI, T
    MATSUO, H
    YOKOYAMA, K
    MURAKAMI, T
    ONOJIMA, N
    YAMAMOTO, H
    ORII, T
    KIMURA, A
    ORBITAL MECHANICS AND MISSION DESIGN, 1989, 69 : 59 - 71
  • [6] MIRA: A new high sensitivity, multicollector mass spectrometer for isotopologue and ultra-small sample stable isotope analysis
    Dennis, P. F.
    GEOCHIMICA ET COSMOCHIMICA ACTA, 2008, 72 (12) : A210 - A210
  • [7] Design Space Exploration of TAGE Branch Predictor with Ultra-Small RAM
    Zhou, Chaobing
    Huang, Libo
    Li, Zhisheng
    Zhang, Tan
    Dou, Qiang
    PROCEEDINGS OF THE GREAT LAKES SYMPOSIUM ON VLSI 2017 (GLSVLSI' 17), 2017, : 281 - 286
  • [8] Accelerator mass spectrometry of ultra-small samples with applications in the biosciences
    Salehpour, Mehran
    Hakansson, Karl
    Possnert, Goran
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2013, 294 : 97 - 103
  • [9] Mass spectrometer instrumentation for landers on small bodies and planetary moons
    McEntire, R
    Managadze, G
    Rosenbauer, H
    Cheng, A
    Murchie, S
    Keath, E
    ACTA ASTRONAUTICA, 1996, 38 (4-8) : 377 - 384
  • [10] Development of ultra-small sized servo actuator with brushless DC motor, planetary gear drive and optical rotary encoder
    Hori, K
    Miyagawa, T
    Ito, K
    INTERNATIONAL JOURNAL OF THE JAPAN SOCIETY FOR PRECISION ENGINEERING, 1997, 31 (01): : 1 - 5