Influence of surgeon experience on the incidence of tip fold-over with slim preformed cochlear implant electrodes

被引:0
|
作者
Knoerle, Esther [1 ]
Vazzana, Caterina [1 ]
Stoever, Timo [1 ]
Helbig, Silke [1 ]
机构
[1] Goethe Univ Frankfurt, Dept Otorhinolaryngol Head & Neck Surg, Univ Hosp, Theodor Stern Kai 7, D-60590 Frankfurt, Germany
关键词
Cochlear implantation; Tip fold-over; Surgical experience; Preformed electrode - perimodiolar; Slim modiolar; SPREAD;
D O I
10.1007/s00405-025-09235-w
中图分类号
R76 [耳鼻咽喉科学];
学科分类号
100213 ;
摘要
PurposeTo assess whether increasing experience with implantation of a thin preformed electrode array for perimodiolar insertion reduces the incidence of tip fold-over (TFO).MethodsThe retrospective study included 100 patients who received a cochlear implant (CI) with the Slim Modiolar (SM) electrode array (Cochlear, Sydney, Australia) at a university CI centre between November 2015 and December 2022. Postoperative radiological imaging was performed to verify electrode position. Surgical reports and radiological images were reviewed and the incidence of TFO was analyzed for three experienced CI surgeons. In addition, the incidence of intraoperative measurements showing evidence of electrode malposition and the mean duration of surgery over time were documented.Results129 SM implantations in 100 patients were included. In seven cases (5.4%) TFO was radiologically detected and successfully revised. In eight cases (6.2%), electrophysiological measurements indicated misplacement and the position was corrected during the same surgery. For one surgeon, five out of 67 implantations (7.5%) were affected by TFO, with the frequency of this complication decreasing over time. The average surgery time for all surgeons was 122.2 (+/- 44.2) minutes, with two surgeons showing a decrease over time.ConclusionThe results show a tendency that the SM electrodes can be implanted with a lower complication rate and faster over time. Therefore, it can be assumed that the implantation of the SM electrode requires a certain amount of practice, even for experienced surgeons. As intraoperative electrophysiological measurements detected 71.4% of all radiologically confirmed TFOs, their use is highly recommended.
引用
收藏
页数:8
相关论文
共 16 条
  • [1] Two Discrete Types of Tip Fold-Over in Cochlear Implantation Using Slim Modiolar Electrodes: Influence of Cochlear Duct Length on Tip Fold-Over
    Oh, Heonjeong
    Kim, Yehree
    Lee, Seung Jae
    Carandang, Marge
    Song, Jae-Jin
    Choi, Byung Yoon
    OTOLOGY & NEUROTOLOGY, 2023, 44 (10) : 1011 - 1014
  • [2] Cochlear Implant Electrode Tip Fold-Over: Our Experience With Long and Flexible Electrode
    Hoegerle, Catalina
    Englhard, Anna
    Simon, Florian
    Grueninger, Ivo
    Mlynski, Robert
    Hempel, John-Martin
    Mueller, Joachim
    OTOLOGY & NEUROTOLOGY, 2022, 43 (01) : 64 - 71
  • [3] Review on cochlear implant electrode array tip fold-over and scalar deviation
    Dhanasingh, Anandhan
    Jolly, Claude
    JOURNAL OF OTOLOGY, 2019, 14 (03) : 94 - 100
  • [4] Review on cochlear implant electrode array tip fold-over and scalar deviation
    Anandhan Dhanasingh
    Claude Jolly
    Journal of Otology, 2019, 14 (03) : 94 - 100
  • [5] Tip Fold-over in Cochlear Implantation: Case Series
    Zuniga, M. Geraldine
    Rivas, Alejandro
    Hedley-Williams, Andrea
    Gifford, Rene H.
    Dwyer, Robert
    Dawant, Benoit M.
    Sunderhaus, Linsey W.
    Hovis, Kristen L.
    Wanna, George B.
    Noble, Jack H.
    Labadie, Robert F.
    OTOLOGY & NEUROTOLOGY, 2017, 38 (02) : 199 - 206
  • [6] Identifying Slim Modiolar Electrode Tip Fold-Over With Intracochlear Electrocochleography
    Varghese, Jordan J.
    Walia, Amit
    Lefler, Shannon M.
    Ortmann, Amanda J.
    Shew, Matthew A.
    Durakovic, Nedim
    Wick, Cameron C.
    Herzog, Jacques A.
    Buchman, Craig A.
    OTOLARYNGOLOGY-HEAD AND NECK SURGERY, 2024, 170 (04) : 1124 - 1132
  • [7] Detection of "tip fold-over" of the cochlear implant electrode array with transimpedance matrix (TIM) measurement
    Nagy Roland
    Perenyi Adam
    Dimak Balazs
    Csanady Miklos
    Kiss Jozsef Geza
    Rovo Laszlo
    ORVOSI HETILAP, 2021, 162 (25) : 988 - 996
  • [8] Characterization of Friction and Speed Effects and Methods for Detection of Cochlear Implant Electrode Tip Fold-over
    Pile, Jason
    Simaan, Nabil
    2013 IEEE INTERNATIONAL CONFERENCE ON ROBOTICS AND AUTOMATION (ICRA), 2013, : 4409 - 4414
  • [9] Mechanisms of electrode fold-over in cochlear implant surgery when using a flexible and slim perimodiolar electrode array
    Ramos-Macias, Angel
    De Miguel, Angel R.
    Carlos Falcon-Gonzalez, Juan
    ACTA OTO-LARYNGOLOGICA, 2017, 137 (11) : 1129 - 1135