Development of a current measurement system for 4itγ ionization chamber

被引:0
|
作者
Chen, Cong [1 ]
Zeng, Weihua [1 ]
Yang, Zhijie [2 ]
Hou, Shengli [1 ]
Liang, Juncheng [2 ]
Liu, Haoran [2 ]
机构
[1] China Univ Geosci Beijing, Sch Geophys & Informat Technol, Beijing 100083, Peoples R China
[2] Natl Inst Metrol NIM, Beijing 100029, Peoples R China
关键词
4it gamma ionization chamber; Current measurement; High-value resistance method;
D O I
10.1016/j.apradiso.2025.111684
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
A current measurement system named ICMS-TIA (Ionization Current Measurement System based on Transimpedance Amplifier) for rapid and accurate measurement of the ionization current of 4it gamma ionization chamber is developed based on an improved high-value resistance I-V conversion method with range switching function. The effective measurement range of the ICMS-TIA is from 50 fA to 50 mu A and the response time of the system is less than 2.8 s. The test shows that the deviation of the current measurement above 1 pA is less than 0.41%, and the uncertainty over a range of current levels is given. In the linearity tests, the maximum deviation is 1.15 %. These tests indicate that this system can be used for current measurement of 4it gamma ionization chamber.
引用
收藏
页数:8
相关论文
共 50 条
  • [2] Automatic system for ionization chamber current measurements
    Brancaccio, F
    Dias, MS
    Koskinas, MF
    APPLIED RADIATION AND ISOTOPES, 2004, 61 (06) : 1339 - 1342
  • [3] An improved leakage current compensation technique for a 4 pi gamma ionization chamber system
    Hino, Y
    Kawada, Y
    Nazaroh
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1996, 369 (2-3): : 392 - 396
  • [4] IONIZATION MEASUREMENT IN A STREAMER CHAMBER
    ASATIANI, TL
    GAZARYAN, KA
    IVANOV, VA
    ZHMYROV, VN
    NAZARYAN, AA
    JETP LETTERS-USSR, 1967, 6 (04): : 83 - &
  • [5] Relative standard system for radon measurement by internal gas ionization chamber
    Zhang, Huaiqin
    Xing, Yu
    Jing, Fanlan
    Hao, Xiuhong
    Yuanzineng Kexue Jishu/Atomic Energy Science and Technology, 1998, 32 (04): : 317 - 322
  • [6] AN IONIZATION CHAMBER AS A CURRENT REFERENCE STANDARD
    KEMP, LAW
    PHYSICS IN MEDICINE AND BIOLOGY, 1962, 7 (03): : 359 - &
  • [7] Development of an a grid ionization chamber
    China Institute of Atomic Energy, Beijing 102413, China
    Hedianzixue Yu Tance Jishue, 2006, 3 (331-334):
  • [8] Development of a Wideband Current Amplifier dedicated to Fission Chamber Measurement
    Bisiach, Danilo
    Barbot, Loic
    De Izarra, Gregoire
    Destouches, Christophe
    Cargnelutti, Manuel
    Zorzut, Sebastjan
    ANIMMA 2021 - ADVANCEMENTS IN NUCLEAR INSTRUMENTATION MEASUREMENT METHODS AND THEIR APPLICATIONS, 2021, 253
  • [9] AN IONIZATION CURRENT INTEGRATOR AND ITS APPLICATION TO MEASUREMENT OF ENERGY DEPENDENCE OF CORNELL THICK-WALLED IONIZATION CHAMBER
    AITKEN, TW
    NUCLEAR INSTRUMENTS & METHODS, 1965, 34 (01): : 160 - &
  • [10] A GENERAL PURPOSE DIRECT CURRENT IONIZATION CHAMBER
    KEENE, JP
    MACKENZIE, LA
    GILBERT, CW
    PHYSICS IN MEDICINE AND BIOLOGY, 1958, 2 (04): : 360 - 373