The prediction of drop size from intravenous infusion controllers

被引:0
|
作者
Hillman, M.R. [1 ]
机构
[1] Bath Institute of Medical Engineering, Wolfson Centre, Royal United Hospital, Combe Park, Bath BA1 3NG, United Kingdom
来源
关键词
Biomedical Engineering--Drug Therapy - Flow of Fluids--Measurements - Liquids--Drop Formation;
D O I
暂无
中图分类号
学科分类号
摘要
In intravenous therapy, gravity-fed drip controllers regulate the flow of fluid in terms of drops per minute, though the total amount of fluid to be infused is specified in ml/h. To convert from one to the other it is regularly, erroneously, assumed that drop size is constant. The work described here investigates the relationship of drop size with respect to drip rate and fluid properties. The results are presented in an easy-to-use graphical form.