Measurement of body fluid volume change using multisite impedance measurements

被引:0
|
作者
Patterson, R. [1 ]
Ranganathan, C. [1 ]
Engel, R. [1 ]
Berkseth, R. [1 ]
机构
[1] University of Minnesota Hospital, Department of Physical Medicine and Rehabilitation, Minneapolis, MN 55455, United States
来源
关键词
BIOLOGICAL MATERIALS - DIALYSIS - ELECTRIC MEASUREMENTS - Impedance - MATHEMATICAL MODELS;
D O I
暂无
中图分类号
学科分类号
摘要
Previous electrical impedance studies that predict body fluid volume and body fat used a single measurement site between the wrist and ankle. Calculations based on an electrical model of the thorax suggest that with a single wrist-to-ankle measurement only the electrical characteristics of the legs and arms will contribute significantly to a prediction of the total body fluid volume. In this study, the impedance of the arms, legs and trunk were measured separately at 100 kHz and the values were correlated with weight change that occurred on 11 patients undergoing haemodialysis. Using multiple linear regression analysis combining data from the arms, legs and trunk gave a correlation coefficient of 0. 87, whereas the correlation coefficient for measurements between the wrist and the ankle was 0. 64. These results suggest that multisite impedance measurements will allow more reliable body fluid and body fat determinations to be made.
引用
收藏
页码:33 / 37
相关论文
共 50 条
  • [1] MEASUREMENT OF BODY-FLUID VOLUME CHANGE USING MULTISITE IMPEDANCE MEASUREMENTS
    PATTERSON, R
    RANGANATHAN, C
    ENGEL, R
    BERKSETH, R
    MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1988, 26 (01) : 33 - 37
  • [2] DETERMINATION OF EXTRACELLULAR FLUID VOLUME USING IMPEDANCE MEASUREMENTS
    ESPEJO, MGA
    NEU, J
    HAMILTON, L
    EITZMAN, B
    GIMOTTY, P
    BUCCIARELLI, RL
    CRITICAL CARE MEDICINE, 1989, 17 (04) : 360 - 363
  • [3] Measurement of leg fluid volume using bioelectrical impedance
    Singh, Bhajan
    Yadollahi, Azadeh
    Bradley, T.
    FASEB JOURNAL, 2014, 28 (01):
  • [4] Measurement of body fluid distribution with posture change using segmental bioelectrical impedance analysis
    Nishimura, Naoki
    Sugimoto, Naotoshi
    Ohnishi, Norikazu
    Matsumoto, Takaaki
    Masuo, Yoshihisa
    Iwase, Satoshi
    FASEB JOURNAL, 2017, 31
  • [5] DETERMINATION OF EXTRACELLULAR FLUID VOLUME IN MAN BY MEASUREMENT OF WHOLE BODY IMPEDANCE
    DUCROT, H
    THOMASSET, A
    JOLY, R
    JUNGERS, P
    EYRAUD, C
    LENOIR, J
    PRESSE MEDICALE, 1970, 78 (51): : 2269 - +
  • [6] BODY-FLUID DETERMINATIONS USING MULTIPLE IMPEDANCE MEASUREMENTS
    PATTERSON, R
    IEEE ENGINEERING IN MEDICINE AND BIOLOGY MAGAZINE, 1989, 8 (01): : 16 - 18
  • [8] Body electrical impedance measurements: Body fat or body fluid?
    Patterson, RP
    SECOND JOINT EMBS-BMES CONFERENCE 2002, VOLS 1-3, CONFERENCE PROCEEDINGS: BIOENGINEERING - INTEGRATIVE METHODOLOGIES, NEW TECHNOLOGIES, 2002, : 1746 - 1747
  • [9] THE INFLUENCE OF URINE VOLUME ON BODY IMPEDANCE MEASUREMENT
    Hong, K. H.
    Park, K. S.