Effect of three-fold astigmatism on high resolution electron micrographs

被引:0
|
作者
Gatan R and D, 6678 Owens Drive, Pleasanton, CA 94588, United States [1 ]
不详 [2 ]
机构
来源
ULTRAMICROSCOPY | / 1卷 / 103-113期
关键词
Aberrations - Atoms - Computer simulation - Computer software - Crystal lattices - Crystal orientation - Grain boundaries - Imaging techniques - Visibility;
D O I
暂无
中图分类号
学科分类号
摘要
The effect of 3-fold astigmatism G on high resolution bright field imaging is explored through image simulation. Images of isolated atomic columns and of two practical specimens are computed are computed for different 200 keV high resolution electron microscopes using EMS software modified to include a 3-fold astigmatism term. The visibility of the effect of 3-fold astigmatism on image contrast depends on the ratio G/(Cs3/4λ 1/4 ), and is therefore strongest for large G and small Cs. Several images calculated for G = 1 μm and Cs = 0.5 mm depart from the symmetry of the sample in ways that bear a close resemblance to commonly observed image contrast effects. The astigmatism also causes lattice images to be shifted. The shift is different crystal orientations. It therefore needs to be taken into account when one is measuring rigid lattice displacements across grain boundaries.
引用
收藏
相关论文
共 50 条
  • [1] EFFECT OF 3-FOLD ASTIGMATISM ON HIGH-RESOLUTION ELECTRON-MICROGRAPHS
    KRIVANEK, OL
    STADELMANN, PA
    ULTRAMICROSCOPY, 1995, 60 (01) : 103 - 113
  • [2] Correction of three-fold astigmatism for ultra-high-resolution TEM
    Overwijk, MHF
    Bleeker, AJ
    Thust, A
    ULTRAMICROSCOPY, 1997, 67 (1-4) : 163 - 170
  • [3] The effect of three-fold astigmatism on measurements of grain boundary volume expansion by high-resolution transmission electron microscopy
    Merkle, KL
    Csencsits, R
    Rynes, KL
    Withrow, JP
    Stadelmann, PA
    JOURNAL OF MICROSCOPY-OXFORD, 1998, 190 : 204 - 213
  • [4] STEM imaging artifacts with three-fold astigmatism in monolayer transition metal dichalcogenides
    Hong, Jinhua
    Koshino, Masanori
    Xu, Hua
    Feng, Qingliang
    Suenaga, Kazu
    APPLIED PHYSICS LETTERS, 2020, 116 (24)
  • [5] 3-FOLD ASTIGMATISM IN HIGH-RESOLUTION TRANSMISSION ELECTRON-MICROSCOPY
    KRIVANEK, OL
    ULTRAMICROSCOPY, 1994, 55 (04) : 419 - 433
  • [6] IMAGE QUAILTIY IMPROVEMENT OF HIGH RESOLUTION ELECTRON MICROGRAPHS WITH ASTIGMATISM AND COMA ABERRATIONS BY OPTICAL MEANS.
    Shen, X.Q.
    Zheng, S.H.
    Li, F.H.
    Boseck, S.
    Optik (Jena), 1988, 79 (04): : 171 - 176
  • [7] The body and three-fold bonds
    Manchot, W
    JUSTUS LIEBIGS ANNALEN DER CHEMIE, 1912, 387 (1/3): : 257 - 293
  • [8] IMAGE QUALITY IMPROVEMENT OF HIGH-RESOLUTION ELECTRON-MICROGRAPHS WITH ASTIGMATISM AND COMA ABERRATIONS BY OPTICAL MEANS
    SHEN, XQ
    ZHENG, SH
    LI, FH
    BOSECK, S
    OPTIK, 1988, 79 (04): : 171 - 176
  • [9] ON INTERPRETATION OF HIGH RESOLUTION ELECTRON MICROGRAPHS OF MACROMOLECULES
    HAYDON, GB
    JOURNAL OF ULTRASTRUCTURE RESEARCH, 1968, 25 (5-6): : 349 - &
  • [10] Three-fold Adaptivity in Groups of Robots: The Effect of Social Learning
    Heinerman, Jacqueline
    Drupsteen, Dexter
    Eiben, A. E.
    GECCO'15: PROCEEDINGS OF THE 2015 GENETIC AND EVOLUTIONARY COMPUTATION CONFERENCE, 2015, : 177 - 183