REMOVAL OF BASE-LINE WANDER AND POWER-LINE INTERFERENCE FROM THE ECG BY AN EFFICIENT FIR FILTER WITH A REDUCED NUMBER OF TAPS.

被引:0
|
作者
Van Alste, J.A. [1 ]
Schilder, T.S. [1 ]
机构
[1] Twente Univ of Technology, Enschede, Neth, Twente Univ of Technology, Enschede, Neth
关键词
ELECTRIC WAVEFORMS - SIGNAL INTERFERENCE;
D O I
暂无
中图分类号
学科分类号
摘要
Linear phase filtering is proposed for the removal of base-line wander and power-line frequency components in electrocardiograms. In order to reduce the large number of computations required in the digital filtering, the desired filter spectrum was defined periodically. Making use of the property that the spectrum period is 50 Hz, the spectrum can be realized with a considerably reduced number of impulse response coefficients. This, in combination with the necessary impulse response symmetry, leads to a reduction in the number of multiplications per output sample by a factor of 10. A suitable impulse response is designed with a pass-band ripple of less than 0. 5 dB and a high stop-band attenuation. The applicability is demonstrated by applying the filtering to exercise electrocardiograms.
引用
收藏
相关论文
共 50 条
  • [1] REMOVAL OF BASE-LINE WANDER AND POWER-LINE INTERFERENCE FROM THE ECG BY AN EFFICIENT FIR FILTER WITH A REDUCED NUMBER OF TAPS
    VANALSTE, JA
    SCHILDER, TS
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1985, 32 (12) : 1052 - 1060
  • [2] An efficient removal of power-line interference and baseline wander from ECG signals by employing Fourier decomposition technique
    Singhal, Amit
    Singh, Pushpendra
    Fatimah, Binish
    Pachori, Ram Bilas
    BIOMEDICAL SIGNAL PROCESSING AND CONTROL, 2020, 57
  • [3] INTEGER COEFFICIENT BANDPASS FILTER FOR THE SIMULTANEOUS REMOVAL OF BASE-LINE WANDER, 50 AND 100 HZ INTERFERENCE FROM THE ECG
    WARIAR, R
    ESWARAN, C
    MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1991, 29 (03) : 333 - 336
  • [4] Baseline Wander and Power-Line Interference Removal from ECG Signals Using Fourier Decomposition Method
    Singh, Pushpendra
    Srivastava, Ishita
    Singhal, Amit
    Gupta, Anubha
    MACHINE INTELLIGENCE AND SIGNAL ANALYSIS, 2019, 748 : 25 - 36
  • [5] Power-line interference and baseline wander elimination in ECG using VMD and EWT
    Mir, Haroon Yousuf
    Singh, Omkar
    COMPUTER METHODS IN BIOMECHANICS AND BIOMEDICAL ENGINEERING, 2024, 27 (15) : 2111 - 2130
  • [6] Removal of power-line interference from the ECG: a review of the subtraction procedure
    Chavdar Levkov
    Georgy Mihov
    Ratcho Ivanov
    Ivan Daskalov
    Ivaylo Christov
    Ivan Dotsinsky
    BioMedical Engineering OnLine, 4
  • [7] Removal of power-line interference from the ECG: a review of the subtraction procedure
    Levkov, Chavdar
    Mihov, Georgy
    Ivanov, Ratcho
    Daskalov, Ivan
    Christov, Ivaylo
    Dotsinsky, Ivan
    BIOMEDICAL ENGINEERING ONLINE, 2005, 4 (1)
  • [8] AN IMPROVED FILTER FOR ECG BASE-LINE WANDER IN REAL-TIME ISCHEMIA MONITORING
    FRANKEL, RA
    POTTALA, EW
    BAILEY, JJ
    JOURNAL OF ELECTROCARDIOLOGY, 1993, 25 : 156 - 156
  • [9] Removal of the baseline wander and the power line interference from ECG signals using the Median-Kalman filter
    Nandi, Kiron
    Lahiri, Abhijit
    FOUNDATIONS AND FRONTIERS IN COMPUTER, COMMUNICATION AND ELECTRICAL ENGINEERING, 2016, : 415 - 419
  • [10] Quantitative analysis of errors due to power-line interference and base-line drift in detection of onsets and offsets in ECG using wavelets
    J. S. Sahambi
    S. N. Tandon
    R. K. P. Bhatt
    Medical and Biological Engineering and Computing, 1997, 35 : 747 - 751