DISSOLVED OXYGEN MONITORING SYSTEM USING A PULSED ELECTRODE: DESIGN, PERFORMANCE, AND EVALUATION.

被引:0
|
作者
Langdon, Christopher [1 ]
机构
[1] Univ of Rhode Island, Graduate Sch, of Oceanography, RI, USA, Univ of Rhode Island, Graduate Sch of Oceanography, RI, USA
来源
Deep-Sea Research, Part A: Oceanographic Research Papers | 1984年 / 31卷 / 11 A Pt A期
基金
美国国家科学基金会; 美国海洋和大气管理局;
关键词
COMPUTERS; MICROPROCESSOR; -; Applications; ELECTRODES; OCEANOGRAPHY; Instruments; OXYGEN; Monitoring;
D O I
暂无
中图分类号
学科分类号
摘要
An instrument for measuring dissolved oxygen in seawater is described. The device uses a conventional polarographic electrode in connection with chronoamperometry to overcome many of the problems limiting the performance of oxygen electrodes. The polarizing potential is applied as a pulse and the resulting current transient is sampled after a 1. 5- to 3-s delay. Current during the first few seconds is 4 to 8 times the steady-state current and is unaffected by flow rate past the membrane provided a recovery time of 3 min or more is allowed between pulses. A microcomputer controls all phases of the measuring process: pulse generation, data acquisition, reduction, and storage. Software is used to correct sensor output for temperature dependence exactly using the activation energy of permeation and the Arrhenius equation. An activity coefficient, a function of temperature and salinity, is computed to correct for salinity. Results of an intercomparison with Winkler determinations are presented and an example is given for the instrument's use to measure the respiration of copepods in an unstirred solution.
引用
收藏
页码:1357 / 1367
相关论文
共 50 条
  • [1] DISSOLVED-OXYGEN MONITORING-SYSTEM USING A PULSED ELECTRODE - DESIGN, PERFORMANCE, AND EVALUATION
    LANGDON, C
    DEEP-SEA RESEARCH PART A-OCEANOGRAPHIC RESEARCH PAPERS, 1984, 31 (11): : 1357 - 1367
  • [2] A TRANSIENT CURRENT MONITORING AND ELECTRODE CHARACTERIZATION SYSTEM FOR A PULSED OXYGEN-ELECTRODE
    WISE, KD
    SMART, RB
    MANCY, KH
    ANALYTICA CHIMICA ACTA, 1980, 116 (02) : 297 - 305
  • [3] Design of Intelligent Monitoring System for Aquaculture Water Dissolved Oxygen
    Ma, Yinchi
    Ding, Wen
    PROCEEDINGS OF 2018 IEEE 3RD ADVANCED INFORMATION TECHNOLOGY, ELECTRONIC AND AUTOMATION CONTROL CONFERENCE (IAEAC 2018), 2018, : 414 - 418
  • [4] AN ELECTRODE SYSTEM FOR MEASURING DISSOLVED OXYGEN
    CARRITT, DE
    KANWISHER, JW
    ANALYTICAL CHEMISTRY, 1959, 31 (01) : 5 - 9
  • [5] A Design Of Dissolved Oxygen Monitoring System Based On Nb-Iot
    Li Yunfeng
    Zhu Tianpei
    2019 INTERNATIONAL CONFERENCE ON SMART GRID AND ELECTRICAL AUTOMATION (ICSGEA), 2019, : 98 - 101
  • [6] DOxy: A Dissolved Oxygen Monitoring System
    Shaghaghi, Navid
    Fazlollahi, Frankie
    Shrivastav, Tushar
    Graham, Adam
    Mayer, Jesse
    Liu, Brian
    Jiang, Gavin
    Govindaraju, Naveen
    Garg, Sparsh
    Dunigan, Katherine
    Ferguson, Peter
    SENSORS, 2024, 24 (10)
  • [7] BIOPROCESS MONITORING OF DISSOLVED-OXYGEN USING A COMPUTERIZED PULSING MEMBRANE-ELECTRODE
    CHEN, JC
    WANG, HY
    BIOTECHNOLOGY PROGRESS, 1993, 9 (01) : 75 - 80
  • [8] Design and Performance Evaluation of eLoran Monitoring System
    Yang, Chaozhong
    Guo, Xiaohang
    Li, Shifeng
    Hu, Zhaopeng
    APPLIED SCIENCES-BASEL, 2024, 14 (16):
  • [9] FRAMEWORK FOR TRANSPORTATION SYSTEM MANAGEMENT PERFORMANCE EVALUATION.
    Brown-West, Orikaye Gogo
    Transportation Research Record, 1985, : 16 - 20
  • [10] Design of Water Dissolved Oxygen Test System
    Wang Juxiang
    Wang Xiao
    Wu Xin
    MATERIALS, MECHATRONICS AND AUTOMATION, PTS 1-3, 2011, 467-469 : 1189 - +