The meaning of the point of maximum oscillations in cuff pressure in the indirect measurement of blood pressure. II

被引:1
|
作者
Mauck, G.W. [1 ]
Smith, C.R. [1 ]
Geddes, L.A. [1 ]
Bourland, J.D. [1 ]
机构
[1] Dept. Mechan. Engin., Lehigh Univ., Bethlehem, Pa., United States
关键词
PRESSURE MEASUREMENT - Analogies - TUBES - Elasticity;
D O I
暂无
中图分类号
学科分类号
摘要
The cuff pressure for which these oscillations reach a maximum and its relationship to the true mean arterial pressure was investigated using a simple one-dimensional theoretical model of the cuff-arm-artery system. Results from this model indicate that the cuff pressure for maximal oscillation is strongly dependent on compression chamber air volume, pulse pressure, and arterial elasticity. Parallel experimental studies indicate general agreement with the theoretical model. The cuff pressure for maximal oscillations appers to provide a reasonable estimation of the true mean arterial pressure provided compression chamber air volume is kept small.
引用
收藏
页码:28 / 33
相关论文