Impact of Power and Time in Hepatic Microwave Ablation: Effect of Different Energy Delivery Schemes

被引:0
|
作者
Trujillo, Macarena [1 ]
Najafabadi, Mahtab Ebad [2 ]
Romero, Antonio [3 ]
Prakash, Punit [2 ]
Cornelis, Francois H. [4 ]
机构
[1] Univ Politecn Valencia, Elect Engn Dept, BioMIT, Valencia 46022, Spain
[2] George Washington Univ, Dept Biomed Engn, Washington, DC 20052 USA
[3] Univ Politecn Valencia, Dept Appl Math, Valencia 46022, Spain
[4] Mem Sloan Kettering Canc Ctr, Dept Radiol, New York, NY 10065 USA
关键词
microwave ablation; theoretical model; power delivery scheme; VIVO BOVINE LIVER; DIELECTRIC-PROPERTIES; TEMPERATURE-DEPENDENCE; THERMAL-ABLATION; TISSUE; MODEL; PERFUSION;
D O I
10.3390/s24237706
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Microwave ablation often involves the use of continuous energy-delivery protocols with a fixed power and time. To achieve larger ablation zones, a range of protocols and power levels have been studied in experimental studies. The objective of the present study was to develop and experimentally evaluate the performance of a coupled computational electromagnetic-bioheat transfer model of 2.45 GHz microwave ablation under a variety of continuous and pulsed power delivery schemes. The main aim was to obtain an in-depth knowledge of the influence of energy delivery settings on ablation zone profiles and thermal damage in the peri-ablation zone. In addition to the theoretical model, we evaluated the power delivery schemes using ex vivo experiments and compared them to previously published data from in vivo experiments. The results showed slight differences in terms of the ablation zone size for different power delivery schemes under ex vivo conditions, with the applied energy level being the most important factor that determines ablation zone size; however, under in vivo conditions, applying a high-power pulse prior to and following a longer constant power application (BOOKEND 95 W protocol) presented the most favorable ablation zones. Moreover, the modeling and experimental studies identified threshold applied power and ablation times beyond which increases did not yield substantive increases in ablation zone extents.
引用
收藏
页数:14
相关论文
共 50 条
  • [1] Microwave ablation of the liver in a live porcine model: the impact of power, time and total energy on ablation zone size and shape
    Hui, Terrence Chi Hong
    Brace, Christopher Lee
    Hinshaw, J. Louis
    Quek, Lawrence Han Hwee
    Huang, Ivan Kuang Hsin
    Kwan, Justin
    Lim, Gavin Hock Tai
    Lee, Fred T., Jr.
    Pua, Uei
    INTERNATIONAL JOURNAL OF HYPERTHERMIA, 2020, 37 (01) : 668 - 676
  • [2] Power Delivery Impedance Impact of Power Gating Schemes
    Shekhar, Sameer
    Jain, Amit K.
    Winer, Noam
    2016 IEEE 20TH WORKSHOP ON SIGNAL AND POWER INTEGRITY (SPI), 2016,
  • [3] Microwave ablation energy delivery: Influence of power pulsing on ablation results in an ex vivo and in vivo liver model
    Bedoya, Mariajose
    del Rio, Alejandro Munoz
    Chiang, Jason
    Brace, Christopher L.
    MEDICAL PHYSICS, 2014, 41 (12)
  • [4] High power microwave ablation of normal swine lung: impact of duration of energy delivery on adverse event and heat sink effects
    Kodama, Hiroshi
    Ueshima, Eisuke
    Gao, Song
    Monette, Sebastien
    Paluch, Lee-Ronn
    Howk, Kreg
    Erinjeri, Joseph P.
    Solomon, Stephen B.
    Srimathveeravalli, Govindarajan
    INTERNATIONAL JOURNAL OF HYPERTHERMIA, 2018, 34 (08) : 1186 - 1193
  • [5] Numerical Study on the Effect of Liquid Injection in Hepatic Microwave Ablation
    Tian, Zhen
    Nan, Qun
    Nie, Xiao-hui
    Dong, Tong
    Wang, Rui-rui
    PROCEEDINGS OF THE 2016 INTERNATIONAL CONFERENCE ON BIOMEDICAL AND BIOLOGICAL ENGINEERING, 2016, : 373 - 379
  • [6] Experimental Evaluation of the Heat Sink Effect in Hepatic Microwave Ablation
    Ringe, Kristina I.
    Lutat, Carolin
    Rieder, Christian
    Schenk, Andrea
    Wacker, Frank
    Raatschen, Hans-Juergen
    PLOS ONE, 2015, 10 (07):
  • [7] Impact of microwave ablation on survival rates and recurrence in hepatic malignant tumors
    Shi, Jutian
    Jiang, Yutian
    Li, Jinpeng
    Chen, Hua
    Cong, Ning
    CLINICAL & TRANSLATIONAL ONCOLOGY, 2025, 27 (03): : 1131 - 1141
  • [8] Microwave Synthesis of Zeolites: Effect of Power Delivery
    Gharibeh, Murad
    Tompsett, Geoffrey A.
    Yngvesson, K. Sigfrid
    Couner, W. Curtis
    JOURNAL OF PHYSICAL CHEMISTRY B, 2009, 113 (26): : 8930 - 8940
  • [9] Guidelines for power and time variables for microwave ablation in a porcine liver
    Hope, William W.
    Schmelzer, Thomas M.
    Newcomb, William L.
    Heath, Jessica J.
    Lincourt, Amy E.
    Norton, H. James
    Heniford, B. Todd
    Iannitti, David A.
    JOURNAL OF GASTROINTESTINAL SURGERY, 2008, 12 (03) : 463 - 467
  • [10] Guidelines for Power and Time Variables for Microwave Ablation in a Porcine Liver
    William W. Hope
    Thomas M. Schmelzer
    William L. Newcomb
    Jessica J. Heath
    Amy E. Lincourt
    H. James Norton
    B. Todd Heniford
    David A. Iannitti
    Journal of Gastrointestinal Surgery, 2008, 12 : 463 - 467