Biomechanical cyclical loading on cadaveric cervical spines in a corpectomy model

被引:0
|
作者
Cheng B.C. [1 ,2 ]
Cook D.J. [2 ]
Cuchanski M. [2 ]
Pirris S.M. [3 ]
Welch W.C. [3 ]
机构
[1] Dept. of Bioengineering, Univ. of Pittsburgh, Pittsburgh
[2] Dept. of Neurological Surgery, Univ. of Pittsburgh, Pittsburgh
[3] Dept. of Neurosurgery, Univ. of Pennsylvania, Philadelphia
来源
Journal of ASTM International | 2011年 / 8卷 / 07期
关键词
ASTM; Biomechanical study; Bone screw interface; Corpectomy; Cyclical loading; Flexibility protocol; Lateral mass bone screw; Posterior cervical fixation;
D O I
10.1520/JAI103408
中图分类号
学科分类号
摘要
Approved posterior cervical spinal fixation systems have been submitted to the U.S. Food and Drug Administration with results from standardized test protocols from the ASTM. The bench top mechanical studies are designed to minimize biologic and laboratory variability. However, for implant tests such as ASTM F1717 to be more clinically relevant, anatomical and physiologic considerations must be understood. The specific aim of this study was to determine if a human cadaveric cervical spine with a corpectomy and posterior fixation was effective in maintaining stability prior to and following cyclical loading. Six fresh frozen cadaveric human cervical specimens were harvested and prepared. A C5 corpectomy was performed. Posterior cervical instrumentation was implanted from C3 and C7 spanning across the C4-C6 defect. Each specimen followed an established pure moment test protocol to characterize the instrumented spine in flexion extension, lateral bending, and axial torsion at ±2.5 N m and axial compressive loading to 150 N. Subsequently, each specimen was subjected to 10 000 flexion extension cycles. Following the cyclical loading, each specimen was characterized a second time via the same test protocol. Statistical analyses were then performed on the third cycle data between the two pure moment tests. The mean FE bending range of motion (ROM) was 18.0° ±10.7° prior to the 10 000 cyclical bending protocol. Following cyclical loading, the mean ROM measured 22.0° ±19.9°. In axial compression, the mean ROM was 4.1 mm±1.9 mm prior to cycling and 4.2 mm±1.4 mm post-cycling. A statistically significant difference was detected only in the axial torsion mode of loading (p=0.030). Although the ASTM standard provides consistent test methodologies, biomechanical cadaveric testing remains an important step in the validation of all spinal instrumentation. The pure moment biomechanical cadaveric test protocol for the same construct was capable of detecting significant changes pre- and post-flexion extension fatigue cycling only in the axial torsion mode of loading.© 2011 by ASTM International.
引用
收藏
相关论文
共 50 条
  • [1] Biomechanical evaluation of posterior screw fixation in cadaveric cervical spines
    Papagelopoulos, PJ
    Currier, BL
    Neale, PG
    Hokari, Y
    Berglund, LJ
    Larson, DR
    Fisher, DR
    An, KN
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2003, (411) : 13 - 24
  • [2] An improved biomechanical testing protocol for evaluating multilevel cervical instrumentation in a human cadaveric corpectomy model
    DiAngelo, DJ
    Foley, KT
    SPINAL IMPLANTS: ARE WE EVALUATING THEM APPROPRIATELY, 2003, 1431 : 155 - 172
  • [3] Biomechanical study of anterior cervical corpectomy and step-cut grafting with bioabsorbable screws fixation in cadaveric cervical spine model
    Zhang, Jian
    He, Xijing
    Li, Haopeng
    Wang, Dong
    Zhao, Weidong
    Xu, Jiehua
    Lan, Binshang
    Xu, Siyue
    SPINE, 2006, 31 (19) : 2195 - 2201
  • [4] A BIOMECHANICAL EVALUATION OF CERVICAL SPINAL STABILIZATION METHODS IN A BOVINE MODEL - STATIC AND CYCLICAL LOADING
    SUTTERLIN, CE
    MCAFEE, PC
    WARDEN, KE
    REY, RM
    FAREY, ID
    SPINE, 1988, 13 (07) : 795 - 802
  • [5] Biomechanical Evaluation of a Novel Total Cervical Prosthesis in a Single-Level Cervical Subtotal Corpectomy Model: An In Vitro Human Cadaveric Study
    Wu, Zi-xiang
    Han, Bao-jun
    Zhao, Xiong
    Kong, Liang
    Liu, Da
    Cui, Gene
    Lei, Wei
    JOURNAL OF SURGICAL RESEARCH, 2012, 175 (01) : 76 - 81
  • [6] Cervical spine loading characteristics in a cadaveric C5 corpectomy model using a static and dynamic plate
    Reidy, D
    Finkelstein, J
    Nagpurkar, A
    Mousavi, P
    Whyne, C
    JOURNAL OF SPINAL DISORDERS & TECHNIQUES, 2004, 17 (02): : 117 - 122
  • [7] Biomechanical evaluation of a novel biomimetic artificial intervertebral disc in canine cervical cadaveric spines
    Jacobs, Celien A. M.
    Doodkorte, Remco J. P.
    Kamali, S. Amir
    Abdelgawad, Abdelrahman M.
    Ghazanfari, Samaneh
    Jockenhoevel, Stefan
    Arts, J. J. Chris
    Tryfonidou, Marianna A.
    Meij, Bjorn P.
    Ito, Keita
    JOR SPINE, 2023, 6 (02):
  • [8] Biomechanical evaluation of posterior screw fixation in cadaveric cervical spines (June, pg 13, 2003)
    Pupagelopoulos, PJ
    Currier, BL
    Neale, PG
    Hokari, Y
    Berglund, LJ
    Larson, DR
    Fisher, DR
    An, KN
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2003, (414) : 347 - 347
  • [9] BIOMECHANICAL EVALUATION OF CASPAR CERVICAL SCREWS - COMPARATIVE STABILITY UNDER CYCLICAL LOADING
    GALLAGHER, MR
    MAIMAN, DJ
    REINARTZ, J
    PINTAR, F
    YOGANANDAN, N
    APFELBAUM, RI
    SONNTAG, VKH
    NEUROSURGERY, 1993, 33 (06) : 1045 - 1051
  • [10] BIOMECHANICAL CONSEQUENCES OF CERVICAL SPONDYLECTOMY VERSUS CORPECTOMY
    Dogan, Seref
    Baek, Seungwon
    Sonntag, Volker K. H.
    Crawford, Neil R.
    NEUROSURGERY, 2008, 63 (04) : 303 - 308