Maximum respiratory pressure measuring system: Calibration and evaluation of uncertainty

被引:4
|
作者
Ferreira J.L. [1 ]
Pereira N.C. [1 ]
Oliveira Jr. M. [1 ]
Vasconcelos F.H. [1 ]
Parreira V.F. [2 ]
Tierra-Criollo C.J. [1 ]
机构
[1] Department of Electrical Engineering, Federal University of Minas Gerais, Belo Horizonte
[2] Department of Physiotherapy, Federal University of Minas Gerais, Belo Horizonte
来源
Controle y Automacao | 2010年 / 21卷 / 06期
关键词
Calibration; Maximum respiratory pressures measuring system; Measurement uncertainty; Weighted least squares;
D O I
10.1590/S0103-17592010000600004
中图分类号
学科分类号
摘要
The objective of this paper is to present a methodology for the evaluation of uncertainties in the measurements results obtained during the calibration of a digital manovacuometer prototype (DM) with a load cell sensor pressure device incorporated. Calibration curves were obtained for both pressure sensors of the DM using linear regression by weighted least squares method (WLS). Two models were built to evaluate uncertainty. One takes into account the information listed in the sensor datasheet, resulting in the maximum permissible measurement error of the manovacuometer, and the other on the WLS implemented during calibration. Considering a range of ten calibration points, it was found that calibration procedure designed using WLS modeling indicates that the range of measurement uncertainty extends from 0.2 up to 0.5 kPa. This is inside the manufacter range that extends from 1.5 up 3.5 kPa, showing adequacy for use.
引用
收藏
页码:588 / 597
页数:9
相关论文
共 50 条
  • [1] Uncertainty evaluation for the dynamic calibration of pressure transducer
    Li, Qiang
    Wang, Zhongyu
    Wang, Zhuoran
    [J]. Beijing Hangkong Hangtian Daxue Xuebao/Journal of Beijing University of Aeronautics and Astronautics, 2015, 41 (05): : 847 - 856
  • [2] Increasing the Reliability of Evaluation of Expanded Uncertainty in Calibration of Measuring Instruments
    O. A. Botsiura
    I. P. Zakharov
    [J]. Measurement Techniques, 2020, 63 : 414 - 420
  • [3] Increasing the Reliability of Evaluation of Expanded Uncertainty in Calibration of Measuring Instruments
    Botsiura, O. A.
    Zakharov, I. P.
    [J]. MEASUREMENT TECHNIQUES, 2020, 63 (06) : 414 - 420
  • [4] System gain calibration of CCD and evaluation of uncertainty
    Wang, Hongchao
    Liu, Hongyuan
    Wang, Hengfei
    Ying, Chengping
    Wu, Bin
    [J]. Guangxue Xuebao/Acta Optica Sinica, 2015, 35
  • [5] Measuring Uncertainty Analysis of the New Leveling Staff Calibration System
    Baricevic, Sergej
    Staroveski, Tomislav
    Barkovic, Duro
    Zrinjski, Mladen
    [J]. SENSORS, 2023, 23 (14)
  • [6] Construction and uncertainty evaluation of a calibration system for GPS receivers
    Yeh, T. K.
    Wang, C. S.
    Lee, C. W.
    Liou, Y. A.
    [J]. METROLOGIA, 2006, 43 (05) : 451 - 460
  • [7] MAXIMUM-LIKELIHOOD EVALUATION IN THE MULTISCALE MEASURING SYSTEM
    SHEBAKPOLSKY, MF
    [J]. RADIOTEKHNIKA I ELEKTRONIKA, 1984, 29 (01): : 65 - 68
  • [8] Calibration and uncertainty evaluation of double parallel-joint coordinate measuring machine
    [J]. Xia, G.-S. (xiaguisuo@163.com), 1600, Chinese Academy of Sciences (22):
  • [9] Traceability in metrology and uncertainty evaluation of a calibration system for GPS receivers
    Lee, CW
    Yeh, TK
    Peng, GS
    [J]. RECENT DEVELOPMENTS IN TRACEABLE DIMENSIONAL MEASUREMENTS II, 2003, 5190 : 410 - 418
  • [10] Uncertainty evaluation of proposed setup for the calibration of vertical angle measuring systems by using means for the flat angle calibration
    Siaudinyte, Lauryna
    Suh, Ho Suhng
    [J]. MEASUREMENT, 2015, 67 : 177 - 182