Locking Screw Insertion Path Guidance System in Minimally Invasive Fracture Reduction Surgery

被引:0
|
作者
Choi Y.-H. [1 ]
Lee S.-J. [2 ]
Oh C.-W. [3 ]
Bang H.H. [1 ]
Lee H.W. [1 ]
Park C.W. [1 ]
机构
[1] Institute of Medical Device and Robot, Kyungpook National University
[2] Department of Orthopaedic Surgery, Gyeongsang National University College of Medicine and Gyeongsang National University Changwon Hospital
[3] Department of Orthopaedic Surgery, School of Medicine, Kyungpook National University, Kyungpook National University Hospital
关键词
C-arm fluoroscopy; intramedullary nailing; locking screw guide; minimal invasive fracture surgery;
D O I
10.5302/J.ICROS.2023.23.0014
中图分类号
学科分类号
摘要
In this paper, we propose a system to detect the location and direction of locking screw insertion using only a fluoroscopic image and guide screw insertion path, with a device that combines a SCARA robot arm and a conical remote center-of-motion (RCM) mechanism. In the proposed system, the target location and direction for screw insertion were derived from the fluoroscopic image of an intramedullary nail. The insertion guidance device comprise the SCARA robot arm and the RCM mechanism. The insertion guidance device calculated the current indicated path through forward kinematics analysis, and the target path and the current path were transmitted to the orthopedic surgeon through the control program. Moreover, the insertion guidance device set the insertion position and direction independently. The orthopedic surgeon first adjusted the SCARA robot arm to match the insertion position value to the target value and then sequentially adjusted the RCM mechanism so that the insertion direction matched the target value. After determining the insertion position and direction, the orthopedic surgeon inserted the surgical instrument through the path guided by the insertion guidance device and fastened the screw to the bone fragments and the intramedullary nail. When the proposed screw insertion guidance system was applied to fracture surgery, the locking screw was fastened with a minimum number fluoroscopic image, thus reducing surgery time and radiation exposure. The validity of the proposed method was confirmed through simulations and experiments using a fracture model. © ICROS 2023.
引用
收藏
页码:458 / 465
页数:7
相关论文
共 50 条
  • [1] REDUCTION IN PEDICLE SCREW PLACEMENT TIME WITH A GUIDEWIRELESS PEDICLE SCREW SYSTEM FOR MINIMALLY INVASIVE SPINE SURGERY
    Greenleaf, R.
    Grossi, P.
    Erb, M.
    El-Sayegh, R.
    Pracyk, J. B.
    Marcini, A.
    VALUE IN HEALTH, 2021, 24 : S126 - S126
  • [2] An Intelligent and Autonomous Endoscopic Guidance System for Minimally Invasive Surgery
    Weede, O.
    Moennich, H.
    Mueller, B.
    Woern, H.
    2011 IEEE INTERNATIONAL CONFERENCE ON ROBOTICS AND AUTOMATION (ICRA), 2011,
  • [3] Minimally invasive sliding hip screw insertion technique
    Khan, Asif Z.
    Parker, Martyn J.
    HIP INTERNATIONAL, 2012, 22 (02) : 214 - 217
  • [4] REDUCTION IN PEDICLE SCREW PLACEMENT TIME WITH A NOVEL GUIDEWIRELESS PEDICLE SCREW SYSTEM FOR MINIMALLY INVASIVE SPINE SURGERY: INITIAL FINDINGS
    Grossi, P.
    Badin, M.
    Erb, M.
    Cirilli, L.
    Malone, D.
    Walsh, M.
    Menzie, A. M.
    Bhattacharyya, S. K.
    VALUE IN HEALTH, 2019, 22 : S223 - S223
  • [5] Navigated and minimally invasive screw osteosynthesis of a talus fracture
    Haida, Dominik M.
    Moehlig, Thorsten
    Huber-Wagner, Stefan
    UNFALLCHIRURGIE, 2025, 128 (01): : 64 - 68
  • [6] Minimally invasive surgery in fracture management
    ZENG Bingfang Department of OrthopaedicsSixth Peoples HospitalShanghai Jiao Tong UniversityShanghai China
    中华医学杂志(英文版), 2008, (15) : 1349 - 1351
  • [7] Minimally invasive surgery in fracture management
    Zeng Bing-fang
    CHINESE MEDICAL JOURNAL, 2008, 121 (15) : 1349 - 1351
  • [8] Use of a 2.0-mm locking plate/screw system for mandibular fracture surgery
    Ellis, E
    Graham, J
    JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 2002, 60 (06) : 642 - 645
  • [9] An Historical Perspective of the Development of Plate and Screw Fixation and Minimally Invasive Fracture Surgery With a Unified Biological Approach
    Russell, Thomas A.
    TECHNIQUES IN ORTHOPAEDICS, 2007, 22 (03) : 186 - 190
  • [10] Guidance for the Treatment of Femoral Neck Fracture with Precise Minimally Invasive Internal Fixation Based on the Orthopaedic Surgery Robot Positioning System
    Wu, Xin-bao
    Wang, Jun-qiang
    Sun, Xu
    Han, Wei
    ORTHOPAEDIC SURGERY, 2019, 11 (03) : 335 - 340