Determination of chloride in infant formula and adult/pediatric nutritional formula by potentiometric titration: Single-laboratory validation, first action 2015.07
被引:0
|作者:
机构:
[1] Bolong, Wu
[2] Fengxia, Zhang
[3] Xiaoning, Ma
[4] Fengjuan, Zhou
[5] Brunelle, Sharon L.
关键词:
A potentiometric method for determination of chloride was validated against AOAC Standard Method Performance Requirement (SMPR®) 2014.015. Ten AOAC Stakeholder Panel on Infant Formula and Adult Nutritionals (SPIFAN ) matrixes;
including National Institute of Standards and Technology (NIST) Standard Reference Material (SRM) 1849a;
were tested in duplicate on 6 independent days. The repeatability (RSDr) ranged from 0.43 to 1.34%;
and the intermediate reproducibility (RSDiR) ranged from 0.80 to 3.04%. All results for NIST SRM 1849a were within the range of the certified concentration (701 ± 17 mg/100 g). Recovery was demonstrated with two overspike levels;
50;
and;
100%;
in the 10 SPIFAN matrixes. Samples were tested in duplicate on 3 different days;
and all results were within the SMPR requirement of 95 to 105%. The LOQs of the method for powdered products and ready-to-feed or reconstituted products were 20 mg/100 g and 2.2 mg/100 mL;
respectively. A wide analytical range from the LOQ to 99.5% chlorine content can be reached with an appropriate dilution factor;
but in practice;
the upper analytical value observed in routine matrix testing was approximately 1080 mg/100 g in skim milk powder. This is a rapid;
simple;
and reliable chlorine-testing method applicable to infant formula;
adult nutritionals;
and ingredients used in these dairy-based products;
such as skim milk powder;
desalted whey powder;
whey protein powder;
and whole milk powder;