Dose-space mapping study on 2D ion-chamber array in radiotherapy

被引:0
|
作者
Du Q. [1 ,2 ]
Wu Z.-F. [1 ,2 ]
机构
[1] Beijing Key Laboratory of Nuclear Detection & Measurement Technology, Beijing
[2] Institute of Nuclear and New Energy Technology, Tsinghua University, Beijing
来源
| 1875年 / Atomic Energy Press卷 / 50期
关键词
2D ion-chamber array; Dosimetry; Effective point of measurement; Monte Carlo;
D O I
10.7538/yzk.2016.50.10.1875
中图分类号
学科分类号
摘要
The 2D ion-chamber array is widely used in dose distribution measurement in radiotherapy, and its effective point of measurement (EPOM), i.e. the accurate dose-space mapping position related to the dose from ion-chamber, is an important problem in accurate radiotherapy. In this study, Monte Carlo simulations were employed to research the EPOM for 2D ion-chamber array. Both the actual dose distribution along the central axis of photon beam in a water phantom in the absence of the probe and the dose curve from 2D ion-chamber array were simulated to locate the position of the EPOM by correlation analysis. It is found that the EPOM for 2D ion-chamber array is shifted from the front face of the ion-chamber cavity toward the cavity center by less than 1 mm. And the EPOM depends on the ion-chamber layout and the beam characteristics. The EPOM shift increases with the length of the cavity and the energy of the beam, and decreases with the increase of the cavity diameter. © 2016, Editorial Board of Atomic Energy Science and Technology. All right reserved.
引用
收藏
页码:1875 / 1880
页数:5
相关论文
共 11 条
  • [1] Almond P.R., Biggs P.J., Coursey B.M., Et al., AAPM's TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams, Medical Physics, 26, 9, pp. 1847-1870, (1999)
  • [2] Andreo P., Huq M.S., Westermark M., Et al., Protocols for the dosimetry of high-energy photon and electron beams: A comparison of the IAEA TRS-398 and previous international codes of practice, Physics in Medicine and Biology, 47, 17, pp. 3033-3053, (2002)
  • [3] Low D.A., Moran J.M., Dempsey J.F., Et al., Dosimetry tools and techniques for IMRT, Medical Physics, 38, 3, pp. 1313-1338, (2011)
  • [4] Kawrakow I., On the effective point of measurement in megavoltage photon beams, Medical Physics, 33, 6, pp. 1829-1839, (2006)
  • [5] McEwen M.R., Kawrakow I., Ross C.K., The effective point of measurement of ionization chambers and the build-up anomaly in MV X-ray beams, Medical Physics, 35, 3, pp. 950-958, (2008)
  • [6] Tessier F., Kawrakow I., Effective point of measurement of thimble ion chambers in megavoltage photon beams, Medical Physics, 37, 1, pp. 96-107, (2010)
  • [7] Looe H.K., Harder D., Poppe B., Experimental determination of the effective point of measurement for various detectors used in photon and electron beam dosimetry, Physics in Medicine and Biology, 56, 14, pp. 4267-4290, (2011)
  • [8] Legrand C., Hartmann G.H., Karger C.P., Experimental determination of the effective point of measurement for cylindrical ionization chambers in <sup>60</sup>Co gamma radiation, Physics in Medicine and Biology, 57, 11, pp. 3463-3475, (2012)
  • [9] Legrand C., Hartmann G.H., Karger C.P., Experimental determination of the effective point of measurement and the displacement correction factor for cylindrical ionization chambers in a 6 MV photon beam, Physics in Medicine and Biology, 57, 21, pp. 6869-6880, (2012)
  • [10] Wang L.L.W., Rogers D.W.O., Study of the effective point of measurement for ion chambers in electron beams by Monte Carlo simulation, Medical Physics, 36, 6, pp. 2034-2042, (2009)