Linearity calibration and uncertainty evaluation for capacitance displacement sensor

被引:3
|
作者
Ge, Chuan [1 ]
Zhang, De-Fu [1 ]
Li, Peng-Zhi [1 ]
Guo, Kang [1 ]
Li, Pei-Yue [1 ]
Yang, Huai-Jiang [1 ]
机构
[1] State Key Laboratory of Applied Optics, Changchun Institute of Optics, Fine Mechanics and Physics, Chinese Academy of Sciences, Changchun,130033, China
关键词
D O I
10.3788/OPE.20152309.2546
中图分类号
学科分类号
摘要
As the linearity of capacitance sensor in adjusting a lithographic projection objective could not meet the requirement of the adjustment mechanism for accuracy standard, a method to improve the measuring linearity of the capacitance sensor was proposed. A piezoelectric actuator was used to provide displacement feeding and a higher precise laser length interferometer was used to provide displacement feedback to ensure the accuracy of moving control. The fitting factors getting from the high order curve fitting were used to calibrate the linearity of the capacitance sensor online. Then, the uncertainties of the environment, installation mechanism and control were analyzed and evaluated to meet the accuracy of the linearity measurement. Finally, calibration experiments for the sensor were carried out. The experiment results indicate that the calibration method of the linearity proposed reduces the influence of the error on the measurement results. After calibration, the linearity of capacitance sensor increases from 0.04714% to 0.00484%, nearly an order of magnitude. Moreover, the capacitance sensor has high linearity repeatability, the deviation of repeatability is 0.38 nm, and the combined uncertainty of the linearity in the whole stroke is 5.70 nm, which meets the displacement control accuracy of the lithographic projection objective. ©, 2015, Chinese Academy of Sciences. All right reserved.
引用
收藏
页码:2546 / 2552
相关论文
共 50 条
  • [1] LINEARITY OF CAPACITANCE DISPLACEMENT TRANSDUCERS
    GLADWIN, MT
    WOLFE, J
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 1975, 46 (08): : 1099 - 1100
  • [2] Linearity calibration method of the high-precision displacement sensor
    Zhang, Defu
    Ge, Chuan
    Li, Xianling
    Ni, Mingyang
    Guo, Kang
    Li, Pengzhi
    [J]. Yi Qi Yi Biao Xue Bao/Chinese Journal of Scientific Instrument, 2015, 36 (05): : 982 - 988
  • [3] The Measurement Uncertainty Evaluation of Yaw Sensor Calibration
    Tipsuwanporn, Vittaya
    Chinchusak, Wanchai
    Jaturunruangsri, Supaporn
    [J]. 2016 16TH INTERNATIONAL CONFERENCE ON CONTROL, AUTOMATION AND SYSTEMS (ICCAS), 2016, : 555 - 558
  • [4] High-Accuracy Calibration Based on Linearity Adjustment for Eddy Current Displacement Sensor
    Liu, Wei
    Liang, Bing
    Jia, Zhenyuan
    Feng, Di
    Jiang, Xintong
    Li, Xiao
    Zhou, Mengde
    [J]. SENSORS, 2018, 18 (09)
  • [5] Performance evaluation and calibration of capacitance sensor for estimating the salinity of reclaimed land
    Kim, Donggeun
    Son, Younghwan
    Park, Jaesung
    Kim, Taejin
    Jeon, Jihun
    [J]. INTERNATIONAL JOURNAL OF AGRICULTURAL AND BIOLOGICAL ENGINEERING, 2020, 13 (01) : 206 - 210
  • [6] Uncertainty evaluation of slope coefficient of high precision displacement sensor
    Tudic, Vladimir
    [J]. Proceedings ELMAR-2006, 2006, : 231 - 234
  • [7] A CAPACITANCE DISPLACEMENT SENSOR WITH ELASTIC DIAPHRAGM
    ROTH, P
    GMELIN, E
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 1992, 63 (03): : 2051 - 2053
  • [8] Evaluation of Humidity Sensor Calibration Uncertainty by Monte Carlo Method
    Wei, M.
    Wen, C.
    Li, C.
    Miao, J.
    [J]. MAPAN-JOURNAL OF METROLOGY SOCIETY OF INDIA, 2024, 39 (03): : 625 - 635
  • [9] Improved 1 kHz capacitance calibration uncertainty
    Jeffery, AM
    Koffman, AD
    [J]. IMTC 2002: PROCEEDINGS OF THE 19TH IEEE INSTRUMENTATION AND MEASUREMENT TECHNOLOGY CONFERENCE, VOLS 1 & 2, 2002, : 73 - 77
  • [10] POWER SENSOR LINEARITY CALIBRATION WITH AN UNKNOWN ATTENUATOR
    Yhland, K.
    Stenarson, J.
    Wingqvist, C.
    [J]. 2010 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS CPEM, 2010, : 769 - 770