Comparison of intelligent 4D CT sequence scanning and conventional spiral 4D CT: A first comprehensive phantom study

被引:10
|
作者
Werner, René [1 ]
Szkitsak, Juliane [2 ]
Sentker, Thilo [1 ]
Madesta, Frederic [1 ]
Schwarz, Annette [3 ,4 ]
Fernolendt, Susanne [3 ,4 ]
Vornehm, Marc [3 ,4 ]
Gauer, Tobias [1 ]
Bert, Christoph [2 ]
Hofmann, Christian [4 ]
机构
[1] University Medical Center Hamburg-Eppendorf, Martinistr. 52, Hamburg,20246, Germany
[2] Department of Radiation Oncology, Universitätsklinikum Erlangen, Friederich-Alexander-Universität Erlangen-Nürnberg, Erlangen,91504, Germany
[3] Friederich-Alexander-Universität Erlangen-Nürnberg, Erlangen,91504, Germany
[4] Siemens Healthcare GmbH, Siemensstr. 3, Forchheim,91301, Germany
来源
Physics in Medicine and Biology | 2021年 / 66卷 / 01期
关键词
Radiotherapy - Computerized tomography - Quality control - Image reconstruction - Image quality - Phantoms - Image enhancement;
D O I
10.1088/1361-6560/abc93a
中图分类号
学科分类号
摘要
4D CT imaging is a cornerstone of 4D radiotherapy treatment. Clinical 4D CT data are, however, often affected by severe artifacts. The artifacts are mainly caused by breathing irregularity and retrospective correlation of breathing phase information and acquired projection data, which leads to insufficient projection data coverage to allow for proper reconstruction of 4D CT phase images. The recently introduced 4D CT approach i4DCT (intelligent 4D CT sequence scanning) aims to overcome this problem by breathing signal-driven tube control. The present motion phantom study describes the first in-depth evaluation of i4DCT in a real-world scenario. Twenty-eight 4D CT breathing curves of lung and liver tumor patients with pronounced breathing irregularity were selected to program the motion phantom. For every motion pattern, 4D CT imaging was performed with i4DCT and a conventional spiral 4D CT mode. For qualitative evaluation, the reconstructed 4D CT images were presented to clinical experts, who scored image quality. Further quantitative evaluation was based on established image intensity-based artifact metrics to measure (dis)similarity of neighboring image slices. In addition, beam-on and scan times of the scan modes were analyzed. The expert rating revealed a significantly higher image quality for the i4DCT data. The quantitative evaluation further supported the qualitative: While 20% of the slices of the conventional spiral 4D CT images were found to be artifact-affected, the corresponding fraction was only 4% for i4DCT. The beam-on time (surrogate of imaging dose) did not significantly differ between i4DCT and spiral 4D CT. Overall i4DCT scan times (time between first beam-on and last beam-on event, including scan breaks to compensate for breathing irregularity) were, on average, 53% longer compared to spiral CT. Thus, the results underline that i4DCT significantly improves 4D CT image quality compared to standard spiral CT scanning in the case of breathing irregularity during scanning. © 2021 Institute of Physics and Engineering in Medicine.
引用
收藏
相关论文
共 50 条
  • [1] Comparison of 4D cone beam CT and 4D spiral scan CT using a dedicated chest phantom
    Nill, S.
    Jetter, S.
    Tacking, T.
    Moser, T.
    Remmert, G.
    Biederer, J.
    Fabel-Schulte, M.
    Dinkel, J.
    Oelfke, U.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2006, 81 : S213 - S214
  • [2] Intelligent 4D CT sequence scanning (i4DCT): First scanner prototype implementation and phantom measurements of automated breathing signal-guided 4D CT
    Werner, Rene
    Sentker, Thilo
    Madesta, Frederic
    Schwarz, Annette
    Vornehm, Marc
    Gauer, Tobias
    Hofmann, Christian
    [J]. MEDICAL PHYSICS, 2020, 47 (06) : 2408 - 2412
  • [3] Intelligent 4D CT Sequence Scanning (i4DCT)
    Werner, R.
    Sentker, T.
    Madesta, F.
    Gauer, T.
    Hofmann, C.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2018, 102 (03): : S5 - S5
  • [4] Intelligent 4D CT sequence scanning ( i4DCT): First prototype measurements
    Werner, R.
    Sentker, T.
    Madesta, F.
    Gauer, T.
    Hofmann, C.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2020, 152 : S954 - S955
  • [5] Clinical application of breathing-adapted 4D CT: image quality comparison to conventional 4D CT
    Werner, Rene
    Szkitsak, Juliane
    Madesta, Frederic
    Buettgen, Laura
    Wimmert, Lukas
    Sentker, Thilo
    Fietkau, Rainer
    Haderlein, Marlen
    Bert, Christoph
    Gauer, Tobias
    Hofmann, Christian
    [J]. STRAHLENTHERAPIE UND ONKOLOGIE, 2023, 199 (07) : 686 - 691
  • [6] Clinical application of breathing-adapted 4D CT: image quality comparison to conventional 4D CT
    René Werner
    Juliane Szkitsak
    Frederic Madesta
    Laura Büttgen
    Lukas Wimmert
    Thilo Sentker
    Rainer Fietkau
    Marlen Haderlein
    Christoph Bert
    Tobias Gauer
    Christian Hofmann
    [J]. Strahlentherapie und Onkologie, 2023, 199 : 686 - 691
  • [7] Intelligent 4D CT sequence scanning (i4DCT): Concept and performance evaluation
    Werner, Rene
    Sentker, Thilo
    Madesta, Frederic
    Gauer, Tobias
    Hofmann, Christian
    [J]. MEDICAL PHYSICS, 2019, 46 (08) : 3462 - 3474
  • [8] Intracardiac Flow at 4D CT: Comparison with 4D Flow MRI
    Lantz, Jonas
    Gupta, Vikas
    Henriksson, Lilian
    Karlsson, Matts
    Persson, Anders
    Carlhall, Carl-Johan
    Ebbers, Tino
    [J]. RADIOLOGY, 2018, 289 (01) : 51 - 58
  • [9] 4D CT scanning: Imaging and planning
    Keall, P.
    Langner, U.
    Suh, Y.
    [J]. MEDICAL PHYSICS, 2007, 34 (06) : 2581 - 2581
  • [10] An Approach to Convert 4D Geometry into a 4D CT Scan
    Villard, P. F.
    Beuve, M.
    Shariat, B.
    [J]. WSCG 2006: SHORT PAPERS PROCEEDINGS: 14TH INTERNATIONAL CONFERENCE IN CENTRAL EUROPE ON COMPUTER GRAPHICS, VISUALIZATION AND COMPUTER VISION 2006, 2006, : 163 - 169