Monitoring of blood removal failure by continuous measurement of blood circuit vibration

被引:0
|
作者
Funaba D. [1 ]
Moriya Y. [1 ]
Oku T. [2 ]
Yamauchi S. [2 ]
Motohashi Y. [2 ]
Sato T. [1 ,2 ]
机构
[1] Graduate School of Engineering, Toin University of Yokohama
[2] Department of Clinical Engineering, Faculty of Biological Engineering, Toin University of Yokohama
关键词
Backflow; Blood circuit vibration; Blood removal failure;
D O I
10.11239/jsmbe.Annual59.701
中图分类号
学科分类号
摘要
The possibility of monitoring blood removal failure by continuous measurement of blood circuit vibration due to backflow induced by blood removal failure was investigated. Measurement of the change in blood circuit vibration displacement amplitude at each blood removal pressure showed a proportional relationship between the pressure and the amplitude, thus indicating that blood removal failure can be detected by monitoring the displacement amplitude of blood circuit vibration. © 2021, Japan Soc. of Med. Electronics and Biol. Engineering. All rights reserved.
引用
收藏
页码:701 / 702
页数:1
相关论文
共 50 条
  • [1] BLOOD-VOLUME MONITORING BY ULTRASONIC MEASUREMENT IN THE EXTRACORPOREAL BLOOD CIRCUIT
    SCHNEDITZ, D
    PROBST, W
    KUBISTA, H
    BINSWANGER, U
    NIEREN-UND HOCHDRUCKKRANKHEITEN, 1991, 20 (11) : 649 - 652
  • [2] Development of a Noninvasive Continuous Blood Pressure Measurement and Monitoring System
    Islam, Md. Manirul
    Rafi, Fida Hasan Md.
    Mitul, Abu Farzan
    Ahmad, Mohiuddin
    Rashid, M. A.
    bin Abd Malek, Mohd Fareq
    2012 INTERNATIONAL CONFERENCE ON INFORMATICS, ELECTRONICS & VISION (ICIEV), 2012, : 1085 - 1090
  • [3] Nano-Resonator Readout Circuit for Implantable Continuous Blood Glucose Monitoring
    Han, Kwonsang
    Kim, Hyungseup
    Lee, Byeoncheol
    Kim, Jaesung
    Lee, Hyunjoong
    Ko, Hyoungho
    BASIC & CLINICAL PHARMACOLOGY & TOXICOLOGY, 2019, 124 : 8 - 8
  • [4] CONTINUOUS MONITORING OF BLOOD GLUCOSE
    ROSEVEAR, JW
    CLINICAL CHEMISTRY, 1971, 17 (07) : 628 - &
  • [5] Continuous monitoring of blood glucose
    Robert, JJ
    HORMONE RESEARCH, 2002, 57 : 81 - 84
  • [6] Noninvasive Continuous Blood Pressure Measurement and GPS Position Monitoring of Patients
    Krejcar, Ondrej
    Slanina, Zdenek
    Stambachr, Jan
    Silber, Petr
    Frischer, Robert
    2009 IEEE 70TH VEHICULAR TECHNOLOGY CONFERENCE FALL, VOLS 1-4, 2009, : 3 - +
  • [7] A NEW METHOD FOR CONTINUOUS MONITORING OF BLOOD GLUCOSE BY MEASUREMENT OF DISSOLVED OXYGEN
    KADISH, AH
    HALL, DA
    CLINICAL CHEMISTRY, 1965, 11 (09) : 869 - &
  • [8] Continuous monitoring of circuit breakers using vibration analysis
    Hoidalen, HK
    Runde, M
    IEEE TRANSACTIONS ON POWER DELIVERY, 2005, 20 (04) : 2458 - 2465
  • [9] ADOPTION OF CONTINUOUS BEAT-TO-BEAT NOCTURNAL BLOOD PRESSURE MEASUREMENT IN AMBULATORY BLOOD PRESSURE MONITORING
    Estrada, Sebastian
    Kwon, Younghoon
    Baruch, Martin
    Harrison, Elizabeth
    Williams, Ronald
    Cho, Yeilim
    Gerdt, David
    Bonner, Heather
    Mazimba, Sula
    Berry, Richard
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2019, 73 (09) : 1893 - 1893
  • [10] Blood Pressure Measurement/Monitoring
    Carma, Ayala
    Xin, Tong
    Jing, Fang
    JOURNAL OF CLINICAL HYPERTENSION, 2013, 15