A new methodology for efficiency calibration of uncharacterized Ge detectors by using characterized Ge detectors

被引:1
|
作者
Barba-Lobo, A. [1 ,3 ]
Exposito-Suarez, V. M. [2 ]
Suarez-Navarro, J. A. [2 ]
Bolivar, J. P. [1 ]
机构
[1] Univ Huelva, Ctr Nat Resources Hlth & Environm RENSMA, Dept Integrated Sci, Radiat Phys & Environm Grp FRYMA, Huelva 21071, Spain
[2] CIEMAT, Dept Environm, Environm Radioact & Radiol Surveillance Unit URAyV, Ave Complutense 40, Madrid 28040, Spain
[3] Univ Gothenburg, Inst Clin Sci, Dept Med Radiat Sci, Sahlgrenska Acad, SE-41345 Gothenburg, Sweden
关键词
LabSOCS; Efficiency calibration; Gamma-ray spectrometry; Characterized Ge detectors; Uncharacterized; Natural radionuclides; GAMMA-SPECTROMETRY; HPGE DETECTORS; SIMULATION; SAMPLES; GEANT4; PB-210;
D O I
10.1016/j.radphyschem.2024.111763
中图分类号
O64 [物理化学(理论化学)、化学物理学];
学科分类号
070304 ; 081704 ;
摘要
The measurement of gamma emitters natural radionuclides is a very relevant subject in environmental radioactivity in many fields. For this, a valid calibration based on the full-energy peak efficiency (FEPE) is essential, which can be carried out for gamma-ray spectrometry using Monte Carlo codes (Geant4, LabSOCS, MNCP, etc.) without needing a source preparation and with much less time consuming compared to experimental methodologies. Therefore, this work aims to develop a methodology for the FEPE determination of uncharacterized Ge detectors by using characterized Ge detectors with known FEPE. For this, a general relationship was found between the simulated FEPE (epsilon(sim)) for characterized Ge detectors by LabSOCS and the experimental one (epsilon(exp)) for uncharacterized Ge detectors obtained for a problem sample (phosphate rock (PR) in our case), where epsilon(sim) and epsilon(exp) were factorized using three correction factors: geometrical, photon self-absorption and true coincidence summing (TCS) effects. Thus, the epsilon(sim)/epsilon(exp) ratio (FEPER) was obtained for two couples of Ge detectors (2 characterized and 1 uncharacterized detectors), using 4 different cylindrical geometries (C) and 2 Marinelli (M) ones, and varying the sample thickness (h), and gamma energy (E-gamma) (from 46 keV (Pb-210) to 1765 keV (Bi-214)). The consistency obtained for the FEPERs for C geometries suggested a constant geometric factor, finding some deviations for M geometries only at E-gamma < 150 keV. Given a pair of detectors, the behavior of the FEPERs was found to be very similar for all geometries, and h at each specific E-gamma. Moreover, a constat relationship between the FEPERs of both detector couples versus E-gamma was found. Then, the influence of the TCS effects was also studied varying the distance between the geometry bottom and detector window. The developed methodology was also validated confirming the consistency of the FEPERs, which were found to be independent on the chemical composition and apparent density of the sample, supporting our theoretical hypothesis. Consequently, given that the FEPER was found to be independent on the sample type and h, using our developed methodology it is possible to obtain epsilon(exp) for any sample, geometry and h using the FEPER, previously obtained at each E-gamma for the sample selected for methodology development (PR in our case), and the epsilon(sim) that is obtained for the desired case.
引用
收藏
页数:10
相关论文
共 50 条
  • [1] A new and general efficiency calibration methodology for Ge detectors using LabSOCS
    Barba-Lobo, A.
    RADIATION MEASUREMENTS, 2025, 181
  • [2] A practical and general methodology for efficiency calibration of coaxial Ge detectors
    Barba-Lobo, A.
    Bolivar, J. P.
    MEASUREMENT, 2022, 197
  • [3] A new efficiency calibration methodology for different atmospheric filter geometries by using coaxial Ge detectors
    Barba-Lobo, A.
    Bolivar, J. P.
    AIR QUALITY ATMOSPHERE AND HEALTH, 2023, 16 (06): : 1207 - 1214
  • [4] A new efficiency calibration methodology for different atmospheric filter geometries by using coaxial Ge detectors
    A. Barba-Lobo
    J. P. Bolívar
    Air Quality, Atmosphere & Health, 2023, 16 : 1207 - 1214
  • [5] NEW METHOD FOR EFFICIENCY CALIBRATION OF GE(LI) DETECTORS
    HNATOWICZ, V
    FISER, M
    NUCLEAR INSTRUMENTS & METHODS, 1978, 150 (02): : 349 - 352
  • [6] EFFICIENCY CALIBRATION OF GE(LI) DETECTORS USING A RADIUM SOURCE
    WALLACE, G
    COOTE, GE
    NUCLEAR INSTRUMENTS & METHODS, 1969, 74 (02): : 353 - &
  • [7] EFFICIENCY CALIBRATION OF GE(LI) DETECTORS IN CLOSE GEOMETRY
    GMUCA, S
    RIBANSKY, I
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH, 1982, 202 (03): : 435 - 437
  • [8] Wide energy range efficiency calibration method for Ge detectors
    Molnár, GL
    Révay, Z
    Belgya, T
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2002, 489 (1-3): : 140 - 159
  • [9] EFFICIENCY CALIBRATION OF GE(LI) DETECTORS, AND THE EFFECTS OF GEOMETRY AND PHYSICAL-PROPERTIES OF SAMPLES ON THE EFFICIENCY OF DETECTORS
    AYCIK, GA
    JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY-ARTICLES, 1992, 157 (01): : 115 - 123
  • [10] LONG-LIVED STANDARDS FOR EFFICIENCY CALIBRATION OF GE(LI) DETECTORS
    AHMAD, I
    WAHLGREN, M
    NUCLEAR INSTRUMENTS & METHODS, 1972, 99 (02): : 333 - &