Research on the remote calibration method of voltage source based on common-view method and consistency evaluation

被引:0
|
作者
Fang L. [1 ,2 ,3 ]
Duan S. [1 ,2 ,3 ]
Tang Z. [1 ,2 ,3 ]
Li Y. [1 ,2 ,3 ]
Liang Y. [1 ,2 ,3 ]
机构
[1] School of Quality and Technical Supervision, Hebei University, Baoding
[2] National & Local Joint Engineering Research Center of Metrology Instrument and System, Baoding
[3] Hebei Key Laboratory of Energy Measurement and Safety Testing Technology, Baoding
关键词
common-view method; consistency evaluation; metrology; remote calibration; uncertainty;
D O I
10.19650/j.cnki.cjsi.J2210160
中图分类号
学科分类号
摘要
In this article, a remote calibration method of voltage source based on the GPS common-view method is proposed to address the problems of traditional voltage source calibration methods. This method is based on the transfer of non-physical standards and can enable the standard to be placed in the laboratory without having to be transferred to the calibrated site for remote calibration. It can solve the problem of introducing additional errors. By designing the voltage source remote calibration system based on the common-view method, the calibration model is formulated. According to the common-view principle, remote comparison of voltage values between the standard side and the calibrated side can be achieved, and remote calibration of the voltage source can be completed. The calibration results of the remote calibration method and the traditional calibration method at the DC voltage 0~1 V calibration point are compared. The results show that the difference between the voltage differences measured by the two methods is within 5. 2×10-5 V. The uncertainty evaluation process of the two methods at the DC voltage 1 V calibration point is given, and the consistency evaluation of the calibration results of the two methods is carried out by the transfer comparison method. The uncertainty evaluation results show that the extended uncertainty of the remote calibration method at the 1 V calibration point is 9. 182 44×10-5 V (k = 2). The calibration results of the remote method are consistent with the traditional results of the remote method. © 2023 Science Press. All rights reserved.
引用
收藏
页码:308 / 318
页数:10
相关论文
共 27 条
  • [1] KOBATA T, KOJIMA M, KAJIKAWA H., Development of remote calibration system for pressure standard [ J ], Measurement, 45, 10, pp. 2482-2485, (2012)
  • [2] MATSUMOTO H, SASAKI K, HIRAI A., Remote calibration of length standards using 47-km-long optical fiber network, Japanese Journal of Applied Physics, 44, 28, pp. L970-L972, (2005)
  • [3] ALBU M, FERRERO A, MIHAI F., Remote calibration using mobile, multiagent technology [ J ], IEEE Transactions on Instrumentation and Measurement, 54, 1, pp. 24-30, (2005)
  • [4] RAOUF M, ALI R, GADELRAB MS., Construction and remote calibration of an automated resistance measuring system [ J], Mapan: Journal of Metrology Society of India, 26, 2, pp. 125-131, (2011)
  • [5] JEBRONI Z, CHADLI H, CHADLI E, Et al., Remote calibration system of a smart electrical energy meter[J], Journal of Electrical Systems, 13, 4, pp. 806-823, (2017)
  • [6] WANG Q, LI H B, WANG H, Et al., A Remote Calibration Device Using Edge Intelligence[J], Sensors, 22, 1, (2022)
  • [7] KONG X Y, ZHANG X P, LU N, Et al., Online smart meter measurement error estimation based on EKF and LMRLS method[J], IEEE Transactions on Smart Grid, 12, 5, pp. 4269-4279, (2021)
  • [8] LOMBARDI M A, NOVICK A N., Remote time calibrations via the NIST time measurement and analysis service, NCSLI Measure, 1, 4, pp. 50-59, (2006)
  • [9] DAVIS J A, STEVENS M, WHIBBERLEY P B, Et al., Commissioning and validation of a GPS common-view time transfer service at NPL [ C], IEEE International Frequency Control Symposium and PDA Exhibition Jointly with the 17th European Frequency and Time Forum, 2003. Proceedings of the 2003, pp. 1025-1031, (2003)
  • [10] ROVERA G D, CHUPIN B, ABGRALL M, Et al., A simple computation technique for improving the short term stability and the robustness of GPS TAIP3 common-views, 2013 Joint European Frequency and Time Forum & International Frequency Control Symposium (EFTF/ IFC), pp. 827-830, (2013)