Design of a Thomson scattering diagnostic for the SMall Aspect Ratio Tokamak (SMART)

被引:0
|
作者
Kaur, M. [1 ]
Diallo, A. [1 ]
LeBlanc, B. [1 ]
Segado-Fernandez, J. [2 ]
Viezzer, E. [2 ]
Huxford, R. B. [3 ]
Mancini, A. [2 ]
Cruz-Zabala, D. J. [2 ]
Podesta, M. [1 ,4 ]
Berkery, J. W. [1 ]
Garcia-Munoz, M. [2 ]
机构
[1] Princeton Plasma Phys Lab, Princeton, NJ 08540 USA
[2] Univ Seville, Dept Atom Mol & Nucl Phys, Seville, Spain
[3] RBH Opt, Burgess Hill RH15 8HL, W Sussex, England
[4] Ecole Polytech Fed Lausanne, Swiss Plasma Ctr, CH-1015 Lausanne, Switzerland
来源
REVIEW OF SCIENTIFIC INSTRUMENTS | 2024年 / 95卷 / 09期
关键词
TRANSPORT; HEAT;
D O I
10.1063/5.0219308
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
We describe the design of a Thomson scattering (TS) diagnostic to be used on the SMall Aspect Ratio Tokamak (SMART). SMART is a spherical tokamak being commissioned in Spain that aims to explore positive triangularity and negative triangularity plasma scenarios at a low aspect ratio. The SMART TS diagnostic is designed to operate at high spatial resolution, 6 mm scattering length in the low-field side and 9 mm in the high-field side regions, and a wide dynamic range, electron temperature from 1 eV to 1 keV and density from 5x10(18 )m(-3) to 1x10(20) m(-3), to resolve large gradients formed at the plasma edge and in the scrape-off layer (SOL) under different triangularities and low aspect ratios. A 2 J @1064 nm laser will be used that is capable of operating in the burst mode at 1, 2, and 4 kHz to investigate fast phenomena and at 30 Hz to study 1 s (or more) long discharges. The scattered light will be collected over an angular range of 60 degrees - 120 degrees from 28 spatial points in the midplane covering the entire plasma width and the outer midplane SOL. Each scattering signal will be spectrally resolved on five wavelength channels of a polychromator to obtain the electron temperature measurement. We will also present a method to monitor in situ laser alignment in the core during calibrations and plasma operations.
引用
收藏
页数:6
相关论文
共 50 条
  • [1] Design of new Thomson scattering diagnostic system on COMPASS tokamak
    Bilkova, Petra
    Aftanas, Milan
    Boehm, Petr
    Weinzettl, Vladimir
    Sestak, David
    Melich, Radek
    Stoeckel, Jan
    Scannell, Rory
    Walsh, Mike
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2010, 623 (02): : 656 - 659
  • [2] THOMSON SCATTERING DIAGNOSTIC FOR THE TDEV TOKAMAK
    COTE, A
    MICHAUD, D
    RICHARD, N
    NEUFELD, R
    LEGROS, C
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1995, 66 (01): : 487 - 489
  • [3] Design of multipulse Thomson scattering diagnostic for SST-1 tokamak
    Kumar, Ajai
    Chavda, Chhaya
    Saxena, Y. C.
    Singh, Ranjeet
    Thakar, Aruna
    Thomas, Jinto
    Patel, Kiran
    Pandya, Kaushal
    Bedakihale, Vijay
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2007, 78 (04):
  • [4] Multipoint Thomson scattering diagnostic for the tokamak ISTTOK
    Alonso, M.P.
    Wilcock, P.D.
    Varandas, C.A.F.
    Review of Scientific Instruments, 1999, 70 (1 pt 2):
  • [5] THOMSON SCATTERING DIAGNOSTIC FOR THE MICROWAVE TOKAMAK EXPERIMENT
    FOOTE, JH
    BARTER, JD
    SEWALL, NR
    JOLLY, JJ
    SCHLANDER, LF
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1990, 61 (10): : 2861 - 2863
  • [6] Advances of Thomson scattering diagnostic on the TEXTOR tokamak
    Kantor, M. Yu.
    14TH INTERNATIONAL SYMPOSIUM ON LASER-AIDED PLASMA DIAGNOSTICS (LAPD14), 2010, 227
  • [7] Multipoint Thomson scattering diagnostic for the ETE tokamak
    Berni, LA
    Alonso, MP
    Oliveira, RM
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2004, 75 (10): : 3884 - 3886
  • [8] Tangential Thomson scattering diagnostic for the KSTAR tokamak
    Lee, J. H.
    Oh, S.
    Wi, H. M.
    Lee, W. R.
    Kim, K. P.
    Yamada, I.
    Narihara, K.
    Kawahata, K.
    JOURNAL OF INSTRUMENTATION, 2012, 7
  • [9] A multipoint Thomson scattering diagnostic for the tokamak ISTTOK
    Associação Euratom/IST, Centro de Fusão Nuclear, Instituto Superior Técnico, 1096 Lisboa Codex, Portugal
    Rev. Sci. Instrum., 1 II (783-786):
  • [10] A multipoint Thomson scattering diagnostic for the tokamak ISTTOK
    Alonso, MP
    Wilcock, PD
    Varandas, CAF
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1999, 70 (01): : 783 - 786