Comparison of forearm and intra-arterial blood pressure measurements according to body and arm positions in obese patients

被引:0
|
作者
Ismailoglu, Elif Gunay [1 ]
Aygun, Hakan [2 ]
Sahan, Seda [1 ]
Ergin, Eda [1 ]
Gectan, Eliz [3 ]
机构
[1] Izmir Bakircay Univ, Nursing Dept, Hlth Sci Fac, Izmir, Turkiye
[2] Izmir Bakircay Univ, Med Fac, Dept Anesthesiol, Izmir, Turkiye
[3] Izmir Bakircay Univ Cigli Training & Res Hosp, Dept Anesthesiol, Izmir, Turkiye
关键词
blood pressure; forearm blood pressure; intensive care; obese patients; upper arm blood pressure; MANAGEMENT; READINGS; CUFF;
D O I
10.1111/jep.14093
中图分类号
R19 [保健组织与事业(卫生事业管理)];
学科分类号
摘要
RationaleNoninvasive blood pressure (BP) monitoring is very important also difficult for accurate diagnosis and monitor of obese patients.AimsThe study aimed to assess the agreement between forearm arm BP and intra-arterial BP values in a different body and arm positions in obese patients.MethodsThe descriptive study was carried out on 60 intensive care patients with a body mass index above 30 kg/m2 who were monitored with invasive radial BP in the general and surgical intensive care units. BP values obtained from the upper arm and forearm with different arm and body positions were compared with intra-arterial BP results. Bland-Altman analysis and correlation coefficient were used for the accuracy of upper and forearm noninvasive BP reading in different positions.ResultsThe best agreement was found between the forearm systolic BP in the supine and half-sitting position with the arm down and intra-arterial BP values. Also, the best agreement in diastolic BP was found between the half-sitting and half-sitting position with the arm down and intra-arterial BP.ConclusionsForearm systolic BP measurement, especially in the supine and half-sitting position with the arm below the heart level position, was best agreement with intra-arterial measurement, regarded as the gold standard. For this reason, it is more appropriate to use forearm BP measurement in obese patients monitored in intensive care to obtain accurate results.
引用
收藏
页数:9
相关论文
共 50 条
  • [1] BLOOD PRESSURE MEASUREMENTS IN OBESE PERSONS - COMPARISON OF INTRA-ARTERIAL AND AUSCULTATORY MEASUREMENTS
    BERLINER, K
    YILDIZ, M
    FUJIY, H
    GARNIER, B
    LEEDH
    [J]. AMERICAN JOURNAL OF CARDIOLOGY, 1961, 7 (01): : 10 - &
  • [2] Blood pressure measurement in obese patients: non-invasive proximal forearm versus direct intra-arterial measurements
    Verkhovsky, A.
    Smit, M.
    Levin, A.
    Coetzee, J. F.
    [J]. SOUTHERN AFRICAN JOURNAL OF ANAESTHESIA AND ANALGESIA, 2018, 24 (03) : 70 - 74
  • [3] Blood pressure measurement in obese patients: comparison between upper arm and forearm measurements
    Pierin, AMG
    Alavarce, DC
    Gusmao, JL
    Halpern, A
    Mion, D
    [J]. BLOOD PRESSURE MONITORING, 2004, 9 (03) : 101 - 105
  • [4] ARTERIAL-PRESSURE IN OBESE PATIENTS - COMPARISON OF MEASUREMENTS AT THE ARM, FOREARM AND WITH A RADIAL ARTERY CATHETER
    BOWEN, M
    CHO, D
    STEVENS, W
    [J]. ANESTHESIOLOGY, 1995, 83 (3A) : A482 - A482
  • [5] INTRA-ARTERIAL BLOOD-PRESSURE MEASUREMENTS
    ZSIGMOND, EK
    [J]. JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1981, 245 (07): : 703 - 703
  • [6] Blood Pressure Measurement in Severely Obese Patients: Validation of the Forearm Approach in Different Arm Positions
    Leblanc, Marie-Eve
    Auclair, Audrey
    Leclerc, Jacinthe
    Bussieres, Jean
    Agharazii, Mohsen
    Hould, Frederic-Simon
    Marceau, Simon
    Brassard, Patrice
    Godbout, Christian
    Grenier, Audrey
    Cloutier, Lyne
    Poirier, Paul
    [J]. AMERICAN JOURNAL OF HYPERTENSION, 2019, 32 (02) : 175 - 185
  • [7] Influence of the arm position on intra-arterial blood pressure measurement
    Netea, RT
    Bijlstra, PJ
    Lenders, JWM
    Smits, P
    Thien, T
    [J]. JOURNAL OF HUMAN HYPERTENSION, 1998, 12 (03) : 157 - 160
  • [8] ERRORS IN INTRA-ARTERIAL BLOOD-PRESSURE MEASUREMENTS
    ZINGG, W
    BAINES, WD
    MATHUR, AN
    [J]. CANADIAN JOURNAL OF SURGERY, 1968, 11 (01) : 93 - &
  • [9] OSCILLOMETRIC AND INTRA-ARTERIAL BLOOD-PRESSURE MEASUREMENTS IN THE NEONATE - A COMPARISON OF METHODS
    LUI, K
    DOYLE, PE
    BUCHANAN, N
    [J]. AUSTRALIAN PAEDIATRIC JOURNAL, 1982, 18 (01): : 32 - 34
  • [10] Comparison of noninvasive oscillometric and intra-arterial blood pressure measurements in hyperacute stroke
    Manios, Efstathios
    Vemmosa, Konstantinos
    Tsivgoulis, Georgios
    Barlas, Gerasimos
    Eleni, Koroboki
    Spengos, Konstantinos
    [J]. BLOOD PRESSURE MONITORING, 2007, 12 (03) : 149 - 156