Evaluating the Safety and Efficacy of a Novel Glaucoma Drainage Device in High-Risk Adult Glaucoma Patients: A One-Year Pilot Study

被引:0
|
作者
Ahmed, Faisal [1 ,2 ]
Normando, Eduardo [1 ,2 ]
Ahmed, Syed [1 ]
Virdee, Simrun [1 ]
Al-Nahrawy, Ahmed [1 ,2 ]
机构
[1] Imperial Coll Healthcare NHS Trust, Western Eye Hosp, Dept Glaucoma, London NW1 5QH, England
[2] Imperial Coll, Imperial Coll Ophthalmol Res Grp ICORG, Dept Surg & Canc, London NW1 5QH, England
关键词
shunt; tube; glaucoma drainage device; BAERVELDT; OUTCOMES; TUBE;
D O I
10.3390/jcm13174996
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: We report on the 12-month safety and efficacy outcomes of a new non-valved glaucoma drainage device, the eyePlate-300 (Rheon Medical, Lausanne, Switzerland), in managing refractory glaucoma. Methods: A retrospective review was conducted on consecutive patients over 18 who underwent glaucoma drainage device (GDD) surgery with the eyePlate-300 after a single glaucoma consultation between February 2020 and April 2021, with at least 12 months of documented post-op follow-up. Results: A total of 16 eyes from 15 patients were included. Complete success was observed in 47% of patients and overall success in 83%. The mean IOP decreased from 31.5 mm Hg to 10.7 mm Hg (67% reduction from baseline), and the number of IOP-lowering drops was reduced from 3.1 to 0.7 at one year. The mean BCVA remained stable. No additional IOP-lowering surgeries were required, and no severe sight-threatening complications were noted. Conclusions: The initial one-year results suggest that the eyePlate could be a safe and effective device for reducing IOP in an ethnically diverse refractory glaucoma population. Further follow-up is necessary to determine the long-term safety and efficacy.
引用
收藏
页数:16
相关论文
共 50 条
  • [1] Safety and Efficacy of the Preserflo® Microshunt in Refractory Glaucoma: A One-Year Study
    Majoulet, Alexandre
    Scemla, Benjamin
    Hamard, Pascale
    Brasnu, Emmanuelle
    Hage, Alexandre
    Baudouin, Christophe
    Labbe, Antoine
    JOURNAL OF CLINICAL MEDICINE, 2022, 11 (23)
  • [2] One-year results on efficacy and safety of a new glaucoma shunt device versus trabeculetomy
    Pillunat, Karin
    Jamke, Melanie
    Herber, Robert
    Haase, Maike
    Pillunat, Lutz
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2023, 64 (08)
  • [3] Efficacy and Safety of an Adjustable Glaucoma Drainage Device (eyeWatch System) for Refractory Glaucoma: A 2-Year Prospective Study
    Roy, Sylvain
    Mermoud, Andre
    JOURNAL OF GLAUCOMA, 2024, 33 (02) : 132 - 138
  • [4] Focusing on patients at high-risk for glaucoma in Brazil: A pilot study
    Guedes, R. A. Paletta
    Guedes, V. M. Paletta
    Chaoubah, A.
    JOURNAL FRANCAIS D OPHTALMOLOGIE, 2009, 32 (09): : 640 - 645
  • [5] One-Year Efficacy and Safety of Omidenepag Isopropyl in Patients with Normal-Tension Glaucoma
    Inoue, Kenji
    Shiokawa, Minako
    Kunimatsu-Sanuki, Shiho
    Nozaki, Norie
    Shimizu, Kosuke
    Ishida, Kyoko
    Tomita, Goji
    JOURNAL OF OCULAR PHARMACOLOGY AND THERAPEUTICS, 2022, 38 (05) : 354 - 358
  • [6] One-Year Efficacy and Safety of the PAUL Glaucoma Implant Using a Standardized Surgical Protocol
    Jose, Patricia
    Barao, Rafael C.
    Teixeira, Filipa J.
    Marques, Raquel E.
    Peschiera, Riccardo
    Barata, Andre
    Pinto, Luis Abegao
    JOURNAL OF GLAUCOMA, 2022, 31 (03) : 201 - 205
  • [7] One-Year Efficacy and Safety of the PAUL Glaucoma Implant Using a Standardized Surgical Protocol
    Jose, Patricia
    Barao, Rafael C.
    Teixeira, Filipa J.
    Marques, Raquel E.
    Peschiera, Riccardo
    Barata, Andre
    Pinto, Luis Abegao
    OBSTETRICS AND GYNECOLOGY, 2022, 139 (03):
  • [8] Efficacy and Safety of the Susanna Glaucoma Drainage Device After 1 Year of Follow-up
    Susanna, Fernanda N.
    Susanna, Bianca N.
    Susanna, Carolina N.
    Nicolela, Marcelo T.
    Susanna, Remo, Jr.
    JOURNAL OF GLAUCOMA, 2021, 30 (05) : E231 - E236
  • [9] One-year result of XEN45 implant for glaucoma: efficacy, safety, and postoperative management
    Tan, S. Z.
    Walkden, A.
    Au, L.
    EYE, 2018, 32 (02) : 324 - 332
  • [10] One-year result of XEN45 implant for glaucoma: efficacy, safety, and postoperative management
    S Z Tan
    A Walkden
    L Au
    Eye, 2018, 32 : 324 - 332