Proton Beam Range and Charge Verification Using Multilayer Faraday Collector

被引:1
|
作者
Yeap, Ping L. [1 ,2 ]
Lew, Kah S. [1 ,3 ]
Koh, Wei Y. C. [1 ]
Chua, Clifford G. A. [1 ]
Wibawa, Andrew [1 ]
Master, Zubin [1 ]
Lee, James C. L. [1 ,3 ]
Park, Sung Y. [1 ,4 ]
Tan, Hong Q. [1 ]
机构
[1] Natl Canc Ctr Singapore, Div Radiat Oncol, 11 Hosp Crescent, Singapore 169610, Singapore
[2] Univ Cambridge, Dept Oncol, Cambridge, England
[3] Nanyang Technol Univ, Div Phys & Appl Phys, Singapore, Singapore
[4] Duke NUS Med Sch, Oncol Acad Clin Programme, Singapore, Singapore
关键词
proton beam therapy; quality assurance; multi-layer Faraday collector; proton range;
D O I
10.1177/15330338241262610
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose A daily quality assurance (QA) check in proton therapy is ensuring that the range of each proton beam energy in water is accurate to 1 mm. This is important for ensuring that the tumor is adequately irradiated while minimizing damage to surrounding healthy tissue. It is also important to verify the total charge collected against the beam model. This work proposes a time-efficient method for verifying the range and total charge of proton beams at different energies using a multilayer Faraday collector (MLFC).Methods We used an MLFC-128-250 MeV comprising 128 layers of thin copper foils separated by thin insulating KaptonTM layers. Protons passing through the collector induce a charge on the metallic foils, which is integrated and measured by a multichannel electrometer. The charge deposition on the foils provides information about the beam range.Results Our results show that the proton beam range obtained using MLFC correlates closely with the range obtained from commissioning water tank measurements for all proton energies. Upon applying a range calibration factor, the maximum deviation is 0.4 g/cm2. The MLFC range showed no dependence on the number of monitor units and the source-to-surface distance. Range measurements collected over multiple weeks exhibited stability. The total charge collected agrees closely with the theoretical charge from the treatment planning system beam model for low- and mid-range energies.Conclusions We have calibrated and commissioned the use of the MLFC to easily verify range and total charge of proton beams. This tool will improve the workflow efficiency of the proton QA.
引用
收藏
页数:9
相关论文
共 50 条
  • [1] Technical Note: Use of commercial multilayer Faraday cup for offline daily beam range verification at the McLaren Proton Therapy Center
    Tesfamicael, Biniam
    Athar, Basit
    Buele, Ana Bejarano
    Kozlyuk, Valeri
    Nichiporov, Dmitri
    Watts, David A.
    Arjomandy, Bijan
    MEDICAL PHYSICS, 2019, 46 (02) : 1049 - 1053
  • [2] IN VIVO PROTON BEAM RANGE VERIFICATION USING SPINE MRI CHANGES
    Gensheimer, Michael F.
    Yock, Torunn I.
    Liebsch, Norbert J.
    Sharp, Gregory C.
    Paganetti, Harald
    Madan, Neel
    Grant, P. Ellen
    Bortfeld, Thomas
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2010, 78 (01): : 268 - 275
  • [3] In Vivo Proton Beam Range Verification Using Spine MRI Changes
    Gensheimer, M. F.
    Yock, T. I.
    Liebsch, N. J.
    Sharp, G. C.
    Madan, N.
    Grant, P. E.
    Paganetti, H.
    Bortfeld, T.
    MEDICAL PHYSICS, 2009, 36 (06)
  • [4] A Simple Technique for Proton Beam Range Verification
    Burgdorf, B.
    Kassaee, A.
    Garver, E.
    MEDICAL PHYSICS, 2015, 42 (06) : 3348 - 3348
  • [5] Proton beam range verification using PET: Comparison of polycarbonate and copper activation
    Cho, Jongmin
    Ibbott, Geoffrey
    Gillin, Michael
    Gonzalez-Lepera, Carlos
    Mawlawi, Osama
    JOURNAL OF NUCLEAR MEDICINE, 2013, 54
  • [6] Feasibility of using O-18 for PET proton beam range verification
    Cho, Jongmin
    Gillin, Michael
    Ibbott, Geoffrey
    Mawlawi, Osama
    JOURNAL OF NUCLEAR MEDICINE, 2011, 52
  • [7] VERIFICATION OF PROTON BEAM RANGE USING PHOTOPOLYMERIZED PMMA-BASED PLASTICS SCINTILLATOR
    Kim, Gwangsoo
    Kim, Hongjoo
    Kim, Sunghwan
    Acta Physica Polonica B, Proceedings Supplement, 2024, 17 (07):
  • [8] Proton Beam Range Verification for Prostate Treatment Using Pelvic-Like Biological Phantoms
    Shao, Wencheng
    Xie, Yunhe
    Zhang, Liyan
    Lu, Hsiao-Ming
    MEDICAL PHYSICS, 2019, 46 (06) : E442 - E442
  • [9] Study of charge distribution in a dust beam using a Faraday cup
    Kausik, S. S.
    Chakraborty, M.
    Dutta, P.
    Kakati, M.
    Saikia, B. K.
    PHYSICS LETTERS A, 2008, 372 (06) : 860 - 865
  • [10] Proton Beam Range Verification using Off-site PET by Imaging Novel Proton-Activated Markers
    Cho, Jongmin
    Ibbott, Geoffrey
    Kerr, Matthew
    Amos, Richard
    Mawlawi, Osama
    2013 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2013,