Iodine concentration, HU accuracy, and metal artifacts evaluation on second-generation dual-layer spectral detector CT images with metal implants: a phantom study

被引:0
|
作者
Zhang, Xinming [1 ,2 ]
Li, Hanwei [1 ]
Wang, Haowen [1 ]
Liu, Xiaomin [3 ]
Deng, Weiwei [3 ]
Zhong, Jianqiu [3 ]
Fan, Litian [1 ]
Ling, Qingqing [1 ]
Hu, Debin [1 ]
Qi, Hongliang [1 ]
Chen, Hongwen [1 ]
机构
[1] Southern Med Univ, Nanfang Hosp, Dept Clin Engn, 1838 Guangzhou Ave North, Guangzhou 510515, Peoples R China
[2] Southern Med Univ, Sch Biomed Engn, Guangzhou, Peoples R China
[3] Philips Healthcare, Clin & Tech Support, Shanghai, Peoples R China
关键词
Dual-layer spectral detector; iodine concentration; phantom study; virtual monoenergetic images; ENERGY CT; ITERATIVE RECONSTRUCTION; REDUCTION SOFTWARE; QUANTIFICATION; PRINCIPLES; KVP; HARDWARE;
D O I
10.1177/02841851241269853
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background: Metal implants may affect the image quality, iodine concentration (IC), and CT Hounsfield unit (HU) quantification accuracy. Purpose: To investigate the quantitative accuracy of IC and HU from dual-layer spectral detector (DLCT) in the presence of metal artifacts. Material and Methods: An experimental cylindrical phantom containing eight iodine inserts and two metal inserts was designed. The phantom underwent scanning at three radiation dose levels and two tube voltage settings. A set of conventional images (CIs), virtual monoenergetic images (VMIs), and iodine concentration maps (ICMs) were generated and measured for all the eight iodine inserts. Quantitative indicators of mean absolute percentage error (MAPE), artifact index (AI), contrast-to-noise ratio (CNR), signal-to-noise ratio (SNR), and standard deviation (SD) on CIs and VMIs were calculated for IC and HU. Subjective score evaluation was also conducted. Results: The MAPEiodine values of all regions of interest across different scanning configurations were all <5%. Almost all APEiodine values were <5%, indicating that metal artifacts had little impact on IC measurements. When the tube voltage was fixed, the SD value of attenuation decreased with the increase of the tube current; this is also true when the tube current was fixed. The middle energy reconstructions seemed to give a good balance between reducing artifacts and improving contrast. Conclusion: VMIs from DLCT can reduce metal artifacts, the accuracy of IC quantification is not sensitive to imaging parameters. In summary, metal implants exhibit minimal impact on image quality and IC quantification accuracy in reconstructed images from DLCT.
引用
收藏
页码:1030 / 1038
页数:9
相关论文
共 39 条
  • [1] Dual-Layer Detector Spectral CT Performance: Quantitative Accuracy and Precision of Iodine Concentration Measurements
    Zhao, X.
    Hameed, T.
    Hutchins, G.
    Liang, Y.
    MEDICAL PHYSICS, 2017, 44 (06) : 3233 - 3233
  • [2] Performance of novel multiparametric second-generation dual-layer spectral detector CT in gouty arthritis
    Li, Meng
    Zhang, Hui
    Liu, Jia-ni
    Zhong, Fei
    Zheng, Song-yuan
    Zhang, Jing
    Chen, Shi-xian
    Lin, Rui-feng
    Zhang, Kang-yu
    Liu, Xiao-min
    Xu, Yi-kai
    Li, Juan
    EUROPEAN RADIOLOGY, 2024,
  • [3] The diagnostic value of second-generation dual-layer spectral detector CT for Stanford type A aortic dissection
    Tan, Cong
    Gui, Shen
    Zhong, Zhilin
    Chen, Xianfeng
    EUROPEAN HEART JOURNAL-CARDIOVASCULAR IMAGING, 2024, 25 (11) : e261 - e261
  • [4] Accuracy of bone mineral density quantification using dual-layer spectral detector CT: a phantom study
    van Hamersvelt, Robbert W.
    Schilham, Arnold M. R.
    Engelke, Klaus
    den Harder, Annemarie M.
    de Keizer, Bart
    Verhaar, Harald J.
    Leiner, Tim
    de Jong, Pim A.
    Willemink, Martin J.
    EUROPEAN RADIOLOGY, 2017, 27 (10) : 4351 - 4359
  • [5] Accuracy of bone mineral density quantification using dual-layer spectral detector CT: a phantom study
    Robbert W. van Hamersvelt
    Arnold M. R. Schilham
    Klaus Engelke
    Annemarie M. den Harder
    Bart de Keizer
    Harald J. Verhaar
    Tim Leiner
    Pim A. de Jong
    Martin J. Willemink
    European Radiology, 2017, 27 : 4351 - 4359
  • [6] Effects of radiation dose levels and spectral iterative reconstruction levels on the accuracy of iodine quantification and virtual monochromatic CT numbers in dual-layer spectral detector CT: an iodine phantom study
    Lu, Xiaomei
    Lu, Zaiming
    Yin, Jiandong
    Gao, Yuying
    Chen, Xingbiao
    Guo, Qiyong
    QUANTITATIVE IMAGING IN MEDICINE AND SURGERY, 2019, 9 (02) : 188 - 200
  • [7] Image quality of conventional images of dual-layer SPECTRAL CT: A phantom study
    van Ommen, Fasco
    Bennink, Edwin
    Vlassenbroek, Alain
    Dankbaar, Jan Willem
    Schilham, Arnold M. R.
    Viergever, Max A.
    de Jong, Hugo W. A. M.
    MEDICAL PHYSICS, 2018, 45 (07) : 3031 - 3042
  • [8] Optimizing coronary CT angiography quality with motion-compensated reconstruction for second-generation dual-layer spectral detector CT
    Liu, Shengmei
    Huang, Linyan
    Gui, Shen
    Pang, Xueting
    Liu, Jing
    Li, Xue
    Wang, Yinqiu
    He, Wenzhang
    Zhang, Xiaodi
    Peng, Liqing
    EUROPEAN RADIOLOGY, 2025, 35 (01) : 381 - 392
  • [9] Feasibility and accuracy of dual-layer spectral detector computed tomography for quantification of gadolinium: a phantom study
    Robbert W. van Hamersvelt
    Martin J. Willemink
    Pim A. de Jong
    Julien Milles
    Alain Vlassenbroek
    Arnold M. R. Schilham
    Tim Leiner
    European Radiology, 2017, 27 : 3677 - 3686
  • [10] Feasibility and accuracy of dual-layer spectral detector computed tomography for quantification of gadolinium: a phantom study
    van Hamersvelt, Robbert W.
    Willemink, Martin J.
    de Jong, Pim A.
    Milles, Julien
    Vlassenbroek, Alain
    Schilham, Arnold M. R.
    Leiner, Tim
    EUROPEAN RADIOLOGY, 2017, 27 (09) : 3677 - 3686