Neurovascular Imaging with Ultra-High-Resolution Photon-Counting CT: Preliminary Findings on Image-Quality Evaluation

被引:0
|
作者
Toth, Adrienn [1 ]
Chetta, Justin A. [1 ]
Yazdani, Milad [1 ]
Matheus, M. Gisele [1 ]
O'Doherty, Jim [1 ,2 ]
Tipnis, Sameer V. [1 ]
Spampinato, M. Vittoria [1 ]
机构
[1] Med Univ South Carolina, Dept Radiol & Radiol Sci, 209D Clin Sci Bldg,96 Jonathan Lucas St, Charleston, SC 29425 USA
[2] Siemens Med Solut, Malvern, PA USA
关键词
DETECTOR CT; INTRACRANIAL ANEURYSMS; ANGIOGRAPHY; DIAGNOSIS; VOLUME;
D O I
10.3174/ajnr.A8350
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
BACKGROUND AND PURPOSE: The first-generation photon-counting detector (PCD) CT was recently introduced into clinical practice and represents a promising innovation in high-resolution CT imaging. The purpose of this study was to assess the image quality of ultra-high-resolution (UHR) PCD-CT compared with energy-integrating detector (EID)-CT, and to explore different reconstruction kernel sharpness levels for the evaluation of intracranial aneurysms. MATERIALS AND METHODS: Ten patients with intracranial saccular aneurysms, who had previously undergone conventional EID-CT, were prospectively enrolled. CT angiograms were acquired on a clinical dual-source PCD-CT in UHR mode, and reconstructed with four vascular kernels (Bv36, Bv40, Bv44, Bv48). Quantitative and qualitative image quality parameters of the intracranial arteries were evaluated. For the quantitative analysis (image noise, SNR, CNR), regions of interest were manually placed at standard anatomical intracranial and extracranial locations by one author. In addition, vessel border sharpness was evaluated quantitatively. For the qualitative analysis, three blinded neuroradiologists rated PCD-CT and EID-CT image quality for the evaluation of the intracranial vessels (i.e., the aneurysms and nine standard vascular branching locations) on a 5-point Likert-type scale. Additionally, readers independently selected their preferred kernel among the four kernels evaluated on PCD-CT. RESULTS: In terms of quantitative image quality, Bv48, the sharpest kernel, yielded increased image noise, and decreased SNR and CNR parameters compared to Bv36, the smoothest kernel. Compared to EID-CT, the Bv48 kernel offered better quantitative image quality for the evaluation of small intracranial vessels (p < .001). Image quality ratings of the Bv48 were superior to those of the EIDCT, and not significantly different from ratings of the B44 reconstruction kernel. When comparing side-by-side all four PCD-CT reconstruction kernels, readers selected the B48 kernel as the best to visualize the aneurysms in 80% of cases. CONCLUSIONS: UHR PCD-CT provides improved image quality for neurovascular imaging. Although the less sharp kernels provided superior SNR and CNR, the sharpest kernels delivered the best subjective image quality on PCD-CT for the evaluation of intracranial aneurysms.
引用
下载
收藏
页数:8
相关论文
共 50 条
  • [1] Ultra-high-resolution photon-counting CT: "Plus Ultra" in lung imaging?
    Milos, Ruxandra-Iulia
    EUROPEAN RADIOLOGY, 2024, 34 (03) : 1893 - 1894
  • [2] Ultra-high-resolution photon-counting CT: “Plus Ultra” in lung imaging?
    Ruxandra-Iulia Milos
    European Radiology, 2024, 34 : 1893 - 1894
  • [3] Impact of ultra-high-resolution imaging of the lungs on perceived diagnostic image quality using photon-counting CT
    Valerie Van Ballaer
    Adriana Dubbeldam
    Emanuele Muscogiuri
    Lesley Cockmartin
    Hilde Bosmans
    Walter Coudyzer
    Johan Coolen
    Walter de Wever
    European Radiology, 2024, 34 : 1895 - 1904
  • [4] Impact of ultra-high-resolution imaging of the lungs on perceived diagnostic image quality using photon-counting CT
    Van Ballaer, Valerie
    Dubbeldam, Adriana
    Muscogiuri, Emanuele
    Cockmartin, Lesley
    Bosmans, Hilde
    Coudyzer, Walter
    Coolen, Johan
    de Wever, Walter
    EUROPEAN RADIOLOGY, 2024, 34 (03) : 1895 - 1904
  • [5] Ultra-high-resolution photon-counting CT of the lungs: a comparative study
    Van Ballaer, V.
    Dubbeldam, A.
    Cockmartin, L.
    Muscogiuri, E.
    Coolen, J.
    Coudyzer, W.
    De Wever, W.
    EUROPEAN RESPIRATORY JOURNAL, 2022, 60
  • [6] Ultra-High-Resolution Coronary CT Angiography With Photon-Counting Detector CT Feasibility and Image Characterization
    Mergen, Victor
    Sartoretti, Thomas
    Baer-Beck, Matthias
    Schmidt, Bernhard
    Petersilka, Martin
    Wildberger, Joachim Ernst
    Euler, Andre
    Eberhard, Matthias
    Alkadhi, Hatem
    INVESTIGATIVE RADIOLOGY, 2022, 57 (12) : 780 - 788
  • [7] Ultra-High-Resolution Photon-Counting CT Imaging of the Chest A New Era for Morphology and Function
    Remy-Jardin, Martine
    Hutt, Antoine
    Flohr, Thomas
    Faivre, Jean-Baptiste
    Felloni, Paul
    Khung, Suonita
    Remy, Jacques
    INVESTIGATIVE RADIOLOGY, 2023, 58 (07) : 482 - 487
  • [8] Ultra-High-Resolution Photon-Counting Detector CT Arthrography of the Ankle: A Feasibility Study
    Luetkens, Karsten Sebastian
    Grunz, Jan-Peter
    Kunz, Andreas Steven
    Huflage, Henner
    Weissenberger, Manuel
    Hartung, Viktor
    Patzer, Theresa Sophie
    Gruschwitz, Philipp
    Erguen, Sueleyman
    Bley, Thorsten Alexander
    Feldle, Philipp
    DIAGNOSTICS, 2023, 13 (13)
  • [9] The Small Pixel Effect in Ultra-High-Resolution Photon-Counting CT of the Lumbar Spine
    Huflage, Henner
    Hendel, Robin
    Woznicki, Piotr
    Conrads, Nora
    Feldle, Philipp
    Patzer, Theresa Sophie
    Erguen, Suleyman
    Bley, Thorsten Alexander
    Kunz, Andreas Steven
    Grunz, Jan-Peter
    INVESTIGATIVE RADIOLOGY, 2024, 59 (09) : 629 - 634
  • [10] Coronary Stenosis Quantification With Ultra-High-Resolution Photon-Counting Detector CT Angiography
    Eberhard, Matthias
    Candreva, Alessandro
    Rajagopal, Rengarajan
    Mergen, Victor
    Sartoretti, Thomas
    Stahli, Barbara E.
    Templin, Christian
    Manka, Robert
    Alkadhi, Hatem
    JACC-CARDIOVASCULAR IMAGING, 2024, 17 (03) : 342 - 344