Instrumented Mouthguard Decoupling Affects Measured Head Kinematic Accuracy

被引:0
|
作者
Gellner, Ryan A. [1 ]
Begonia, Mark T. [1 ]
Wood, Matthew [1 ]
Rockwell, Lewis [1 ]
Geiman, Taylor [1 ]
Jung, Caitlyn [1 ]
Rowson, Steve [1 ]
机构
[1] Virginia Tech, Biomed Engn & Mech, Blacksburg, VA 24061 USA
关键词
Instrumented mouthguard; Boil-and-bite; Decoupling; Measurement error; Head; Impact; Kinematics; IMPACT EXPOSURE; MAGNITUDE;
D O I
10.1007/s10439-024-03550-9
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Many recent studies have used boil-and-bite style instrumented mouthguards to measure head kinematics during impact in sports. Instrumented mouthguards promise greater accuracy than their predecessors because of their superior ability to couple directly to the skull. These mouthguards have been validated in the lab and on the field, but little is known about the effects of decoupling during impact. Decoupling can occur for various reasons, such as poor initial fit, wear-and-tear, or excessive impact forces. To understand how decoupling influences measured kinematic error, we fit a boil-and-bite instrumented mouthguard to a 3D-printed dentition mounted to a National Operating Committee on Standards for Athletic Equipment (NOCSAE) headform. We also instrumented the headform with linear accelerometers and angular rate sensors at its center of gravity (CG). We performed a series of pendulum impact tests, varying impactor face and impact direction. We measured linear acceleration and angular velocity, and we calculated angular acceleration from the mouthguard and the headform CG. We created decoupling conditions by varying the gap between the lower jaw and the bottom face of the mouthguard. We tested three gap conditions: 0 mm (control), 1.6 mm, and 4.8 mm. Mouthguard measurements were transformed to the CG and compared to the reference measurements. We found that gap condition, impact duration, and impact direction significantly influenced mouthguard measurement error. Error was higher for larger gaps and in frontal (front and front boss) conditions. Higher errors were also found in padded conditions, but the mouthguards did not collect all rigid impacts due to inherent limitations. We present characteristic decoupling time history curves for each kinematic measurement. Exemplary frequency spectra indicating characteristic decoupling frequencies are also described. Researchers using boil-and-bite instrumented mouthguards should be aware of their limitations when interpreting results and should seek to address decoupling through advanced post-processing techniques when possible.
引用
收藏
页数:18
相关论文
共 31 条
  • [1] Head impact classification using an instrumented mouthguard
    Wu, Lyndia
    Zarnescu, Livia
    Nangia, Vaibhav
    Cam, Bruce
    Malone, Maria
    Hammoor, Bradley
    Bui, Kevin
    Camarillo, David
    [J]. BRAIN INJURY, 2014, 28 (5-6) : 841 - 842
  • [2] AN INSTRUMENTED MOUTHGUARD TO MEASURE HEAD ACCELERATIONS DUE TO IMPACT
    Birmingham, Grant
    Smith, Lilan
    Brock, Jennifer M.
    Lund, John
    Paris, Anthony J.
    [J]. PROCEEDINGS OF THE ASME SUMMER BIOENGINEERING CONFERENCE - 2013, PT B, 2014,
  • [3] On-Field Performance of an Instrumented Mouthguard for Detecting Head Impacts in American Football
    Lee F. Gabler
    Samuel H. Huddleston
    Nathan Z. Dau
    David J. Lessley
    Kristy B. Arbogast
    Xavier Thompson
    Jacob E. Resch
    Jeff R. Crandall
    [J]. Annals of Biomedical Engineering, 2020, 48 : 2599 - 2612
  • [4] An Instrumented Mouthguard for Measuring Linear and Angular Head Impact Kinematics in American Football
    David B. Camarillo
    Pete B. Shull
    James Mattson
    Rebecca Shultz
    Daniel Garza
    [J]. Annals of Biomedical Engineering, 2013, 41 : 1939 - 1949
  • [5] Comparison of head impact measurements via an instrumented mouthguard and an anthropometric testing device
    Desney G. Greybe
    Christopher M. Jones
    M. Rowan Brown
    Elisabeth M. P. Williams
    [J]. Sports Engineering, 2020, 23
  • [6] An Instrumented Mouthguard for Measuring Linear and Angular Head Impact Kinematics in American Football
    Camarillo, David B.
    Shull, Pete B.
    Mattson, James
    Shultz, Rebecca
    Garza, Daniel
    [J]. ANNALS OF BIOMEDICAL ENGINEERING, 2013, 41 (09) : 1939 - 1949
  • [7] On-Field Performance of an Instrumented Mouthguard for Detecting Head Impacts in American Football
    Gabler, Lee F.
    Huddleston, Samuel H.
    Dau, Nathan Z.
    Lessley, David J.
    Arbogast, Kristy B.
    Thompson, Xavier
    Resch, Jacob E.
    Crandall, Jeff R.
    [J]. ANNALS OF BIOMEDICAL ENGINEERING, 2020, 48 (11) : 2599 - 2612
  • [8] Comparison of head impact measurements via an instrumented mouthguard and an anthropometric testing device
    Greybe, Desney G.
    Jones, Christopher M.
    Brown, M. Rowan
    Williams, Elisabeth M. P.
    [J]. SPORTS ENGINEERING, 2020, 23 (01)
  • [9] Biofidelity and Limitations of Instrumented Mouthguard Systems for Assessment of Rigid Body Head Kinematics
    Abrams, Mitchell Z.
    Venkatraman, Jay
    Sherman, Donald
    Ortiz-Paparoni, Maria
    Bercaw, Jefferson R.
    Macdonald, Robert E.
    Kait, Jason
    Dimbath, Elizabeth D.
    Pang, Derek Y.
    Gray, Alexandra
    Luck, Jason F.
    Bir, Cynthia A.
    Bass, Cameron R.
    [J]. ANNALS OF BIOMEDICAL ENGINEERING, 2024,
  • [10] A Head Impact Detection System Using SVM Classification and Proximity Sensing in an Instrumented Mouthguard
    Wu, Lyndia C.
    Zarnescu, Livia
    Nangia, Vaibhav
    Cam, Bruce
    Camarillo, David B.
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2014, 61 (11) : 2659 - 2668