On the use of new vs. heated sample carriers for luminescence measurements

被引:0
|
作者
Discher, Michael [1 ]
Bassinet, Celine [2 ]
Kim, Hyoungtaek [3 ]
机构
[1] Paris Lodron Univ Salzburg, Dept Environm & Biodivers, Salzburg, Austria
[2] Inst Radioprotect & Surete Nucleaire IRSN, F-92260 Fontenay Aux Roses, France
[3] Korea Atom Energy Res Inst, Radiat Safety Management Div, Yuseong, Daejeon, South Korea
基金
新加坡国家研究基金会;
关键词
Luminescence equipment; Luminescence reader; Sample cup; Sample carrier; Heat exposure; ELECTRONIC COMPONENTS; MOBILE PHONES; THERMOLUMINESCENCE;
D O I
10.1016/j.radmeas.2024.107136
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
To ensure precise luminescence measurements, it is crucial to carry out measurements on clean sample carriers. Moreover, it is essential to ensure that the light emitted by the studied sample is reproducibly reflected towards the detector, particularly when the sample exhibits dim luminescence. This study reports observations made by three laboratories working in the field of retrospective dosimetry. The study discusses the results obtained using new sample carriers (made of stainless steel), instead of sample carriers previously heated to a temperature of 450 degrees C. The article describes the cleaning protocols routinely used by each laboratory and the treatment with distilled water that effectively removed spurious signals. When measuring TL signals of resistors using new sample carriers, a signal reduction of 10% was observed in the second measurement compared to the initial TL. For display glass samples from a mobile phone, a signal reduction between 15% and 32% was observed. However, there was negligible change in reproducibility when using the previously heated sample carriers. The reflectance of the sample carrier surface was indirectly measured using breakthrough signal measurements using violet or blue LEDs. A signal reduction of 12-14 % was observed when heating was applied for 20 new, unheated sample carriers using violet LEDs. The impact on dose estimation is discussed, and some recommendations are provided to guarantee precise luminescence measurements.
引用
收藏
页数:9
相关论文
共 50 条
  • [1] Gelation of single heated vs. double heated whey protein isolate
    Glibowski, P.
    Mleko, S.
    Wesolowska-Trojanowska, M.
    INTERNATIONAL DAIRY JOURNAL, 2006, 16 (09) : 1113 - 1118
  • [2] Measurements of the moisture states on laboratory samples vs. measurements on new concrete bridges
    Nykanen, E
    MOISTURE MEASUREMENT IN CONCRETE CONSTRUCTIONS EXPOSED TO TEMPERATURE AND MOISTURE VARIATIONS, 1997, 174 : 66 - 72
  • [3] Opacity measurements of a radiatively heated boron sample
    Merdji, H
    Missalla, T
    Gilleron, F
    Chenais-Popovics, C
    Gauthier, JC
    Renaudin, P
    Gary, S
    Bruneau, J
    JOURNAL OF QUANTITATIVE SPECTROSCOPY & RADIATIVE TRANSFER, 1997, 58 (4-6): : 783 - 789
  • [4] LUMINESCENCE OF GAINASP-INP INJECTION HETEROLASERS WITH HEATED DISCHARGE CARRIERS
    TOLSTIKHIN, VI
    PISMA V ZHURNAL TEKHNICHESKOI FIZIKI, 1992, 18 (14): : 1 - 7
  • [5] Carriers and counters - P systems with carriers vs. (Blind) counter automata
    Hoogeboom, HJ
    DEVELOPMENTS IN LANGUAGE THEORY, 2003, 2450 : 140 - 151
  • [6] Metric vs. English measurements
    Anderson, AR
    JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1942, 118 : 999 - 999
  • [7] Gemmological and Spectroscopic Features of Untreated vs. Heated Amber
    Wang, Yamei
    Yang, Mingxing
    Nie, Shufang
    Liu, Fen
    JOURNAL OF GEMMOLOGY, 2017, 35 (06) : 530 - 542
  • [8] Luminescence in porous silicon - surface vs. bulk
    Narasimhan, K.L.
    Banerjee, S.
    Journal of Porous Materials, 2000, 7 (01) : 283 - 286
  • [9] Luminescence in Porous Silicon—Surface vs. Bulk
    K.L. Narasimhan
    S. Banerjee
    Journal of Porous Materials, 2000, 7 : 283 - 286
  • [10] AUTOMATED SAMPLE DEOXYGENATION FOR IMPROVED LUMINESCENCE MEASUREMENTS
    ROLLIE, ME
    PATONAY, G
    WARNER, IM
    ANALYTICAL CHEMISTRY, 1987, 59 (01) : 180 - 184