Full correction of the self-absorption of laser-induced plasma beryllium emissions via sample preparation

被引:0
|
作者
Chen, Hui [1 ,2 ]
Chen, Zhuhai [3 ]
Luo, Ronggen [3 ]
Chen, Yuehua [1 ,2 ]
Wei, Shuolei [1 ,2 ]
Mo, Guanqing [1 ,2 ]
Xu, Tao [1 ,2 ]
机构
[1] Fujian Normal Univ, Coll Phys & Energy, Fuzhou 350117, Peoples R China
[2] Fujian Prov Adv High Field Superconducting Mat & E, Fuzhou 350117, Peoples R China
[3] Zijin Min Grp Co Ltd, Longyan 364200, Peoples R China
关键词
LIBS; Self-absorption effect; Powder mixing; Dilution factor; Beryllium; INDUCED BREAKDOWN SPECTROSCOPY; QUANTITATIVE-ANALYSIS; ELEMENTAL ANALYSIS; SPECTRAL-LINES; LIBS ANALYSIS; CURVES; GROWTH; COEFFICIENTS; REDUCTION; PRESSURE;
D O I
10.1016/j.sab.2024.106958
中图分类号
O433 [光谱学];
学科分类号
0703 ; 070302 ;
摘要
Self-absorption effect is always encountered in laser -induced breakdown spectroscopy (LIBS) and immediately distorts the calibration curves especially for the elemental determination of complex materials. With the aim to provide an expeditious approach for LIBS measurements susceptible to self-absorption effect, an exploratory study for sample preparation by powder mixing is implemented, and the self-absorption of laser -induced plasma beryllium emissions is investigated as a function of the dilution factor. For comparison, boric acid and wax powder are separately used as typical binding agent to mix with beryl powder to produce two sets of pressed pellets with sequential gradient of beryllium content variation. For the resonant lines most prone to selfabsorption of Be II emission doublet at 313.042 nm and 313.107 nm, the beryllium spectral line shapes can be directly regulated and improved in the case of both sets of pellets as the dilution factor increases. In addition, the self-absorption effect for the strongly emitting beryllium spectral lines is further assessed by calculating the self-absorption coefficients based on the radiation transport equation. With regard to the both types of binder as diluent, when the dilution factor for beryl increases to 10 with a weight percent of 0.386% for beryllium in the pellet, the SA values may exponentially increase in general to around 0.8 and eventually asymptotically approaches to 1 without exception. Thus the self-absorption effect of laser -induced beryl plasma emissions can be readily overcome with sample preparation by powder mixing. By contrast with laser pulse irradiance in the present work, the dilution factor for powder mixing plays a dominant role in eliminating self-absorption effect. This research proposes a potentially effective approach to reduce self-absorption in laser -induced plasma emissions.
引用
收藏
页数:7
相关论文
共 50 条
  • [1] Calculation of the Self-absorption of a Laser-induced Magnesium Plasma
    He, J.
    Zhang, Q. G.
    LASERS IN ENGINEERING, 2014, 27 (3-4) : 247 - 255
  • [2] Laser-induced plasma characterization through self-absorption quantification
    Hou, Jiajia
    Zhang, Lei
    Zhao, Yang
    Yan, Xingyu
    Ma, Weiguang
    Dong, Lei
    Yin, Wangbao
    Xiao, Liantuan
    Jia, Suotang
    JOURNAL OF QUANTITATIVE SPECTROSCOPY & RADIATIVE TRANSFER, 2018, 213 : 143 - 148
  • [3] Study of Self-Absorption Effect on Laser-Induced Metal Plasma
    Chen Jin-zhong
    Ma Rui-ling
    Wang Jing
    Li Xu
    Su Hong-xin
    SPECTROSCOPY AND SPECTRAL ANALYSIS, 2014, 34 (09) : 2337 - 2341
  • [4] Calibration-free laser-induced plasma analysis of a metallic alloy with self-absorption correction
    Hamidreza Shirvani-Mahdavi
    Seyede Zahra Shoursheini
    Hamid Gholami
    Ziba Dini-Torkamani
    Sare Ghahari-Korani
    Applied Physics B, 2014, 117 : 823 - 832
  • [5] Calibration-free laser-induced plasma analysis of a metallic alloy with self-absorption correction
    Shirvani-Mahdavi, Hamidreza
    Shoursheini, Seyede Zahra
    Gholami, Hamid
    Dini-Torkamani, Ziba
    Ghahari-Korani, Sare
    APPLIED PHYSICS B-LASERS AND OPTICS, 2014, 117 (03): : 823 - 832
  • [6] Progress in the Correction of Self-Absorption Effect on Laser-Induced Breakdown Spectroscopy
    Deng Fan
    Hu Zhen-lin
    Cui Hao-hao
    Zhang Deng
    Tang Yun
    Zhao Zhi-fang
    Zeng Qing-dong
    Guo Lian-bog
    SPECTROSCOPY AND SPECTRAL ANALYSIS, 2021, 41 (10) : 2989 - 2998
  • [7] Accuracy improvement of single-sample calibration laser-induced breakdown spectroscopy with self-absorption correction
    Deng, Fan
    Hu, Zhenlin
    Zhang, Deng
    Chen, Feng
    Niu, Xuechen
    Nie, Junfei
    Zeng, Qingdong
    Guo, Lianbo
    OPTICS EXPRESS, 2022, 30 (06): : 9256 - 9268
  • [8] Laser-induced plasma characterization using self-absorption quantification method
    Zhao Fa-Gang
    Zhang Yu
    Zhang Lei
    Yin Wang-Bao
    Dong Lei
    Ma Wei-Guang
    Xiao Lian-Tuan
    Jia Suo-Tang
    ACTA PHYSICA SINICA, 2018, 67 (16)
  • [9] Hydrogen Alpha Self-Absorption Effects in Laser-Induced Air Plasma
    Parigger, C. G.
    Swafford, L. D.
    Surmick, D. M.
    Witte, M. J.
    Woods, A. C.
    Gautam, G.
    XXII INTERNATIONAL CONFERENCE ON SPECTRAL LINE SHAPES (ICSLS 2014), 2014, 548
  • [10] Effect of Spectral Line Self-Absorption on the Laser-Induced Plasma Diagnostics
    Ben Ahmed, Jalloul
    Fouad, Fethi
    IEEE TRANSACTIONS ON PLASMA SCIENCE, 2014, 42 (08) : 2073 - 2078