Traceable Calibration of Atmospheric Oxidized Mercury Measurements

被引:3
|
作者
Elgiar, Tyler R. [1 ]
Lyman, Seth N. [1 ,2 ]
Andron, Teodor D. [3 ,4 ]
Gratz, Lynne [5 ]
Hallar, A. Gannet [6 ]
Horvat, Milena [3 ,4 ]
Vijayakumaran Nair, Sreekanth [3 ,4 ]
O'Neil, Trevor [1 ]
Volkamer, Rainer [7 ,8 ]
Zivkovic, Igor [3 ,4 ]
机构
[1] Utah State Univ, Bingham Res Ctr, Vernal, UT 84078 USA
[2] Utah State Univ, Dept Chem & Biochem, Logan, UT 84322 USA
[3] JoZef Stefan Inst, Ljubljana 1000, Slovenia
[4] JoZef Stefan Int Postgrad Sch, Ljubljana 1000, Slovenia
[5] Reed Coll, Portland, OR 97202 USA
[6] Univ Utah, Dept Atmospher Sci, Salt Lake City, UT 84112 USA
[7] Univ Colorado, Dept Chem, Boulder, CO 80309 USA
[8] Univ Colorado, CIRES, Boulder, CO 80309 USA
关键词
atmosphere; mercury; elemental; oxidized; calibration; KCL-COATED DENUDERS; PERMEATION TUBES; PERFORMANCE; TOXICITY; VAPOR; TEMPERATURE; CHEMISTRY; PRESSURE; RELEASE; DEVICES;
D O I
10.1021/acs.est.4c02209
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
Most previous measurements of oxidized mercury were collected using a method now known to be biased low. In this study, a dual-channel system with an oxidized mercury detection limit of 6-12 pg m(-3) was deployed alongside a permeation tube-based automated calibrator at a mountain top site in Steamboat Springs Colorado, USA, in 2021 and 2022. Permeation tubes containing elemental mercury and mercury halides were characterized via an International System of Units (SI)-traceable gravimetric method and gas chromatography/mass spectrometry before deployment in the calibrator. The dual-channel system recovered 97 +/- 4 and 100 +/- 8% (+/- standard deviation) of injected elemental mercury and HgBr2, respectively. Total Hg permeation rates and Hg speciation from the gravimetric method, the chromatography system, the dual-channel system, and an independent SI-traceable measurement method performed at the Jozef Stefan Institute laboratory were all comparable within the respective uncertainties of each method. These are the first measurements of oxidized mercury at low environmental concentrations that have been verified against an SI-traceable calibration system in field conditions while sampling ambient air, and they show that accurate, routinely calibrated oxidized mercury measurements are achievable.
引用
收藏
页码:10706 / 10716
页数:11
相关论文
共 50 条
  • [1] Automated Calibration of Atmospheric Oxidized Mercury Measurements
    Lyrnan, Seth
    Jones, Colleen
    O'Neil, Trevor
    Allen, Tanner
    Miller, Matthieu
    Gustin, Mae Sexauer
    Pierce, Ashley M.
    Luke, Winston
    Ren, Xinrong
    Kelley, Paul
    [J]. ENVIRONMENTAL SCIENCE & TECHNOLOGY, 2016, 50 (23) : 12921 - 12927
  • [2] Application of traceable calibration for gaseous oxidized mercury in air
    Nair, Sreekanth Vijayakumaran
    Gacnik, Jan
    Zivkovic, Igor
    Andron, Teodor Daniel
    Ali, Saeed Waqar
    Kotnik, Joze
    Horvat, Milena
    [J]. ANALYTICA CHIMICA ACTA, 2024, 1288
  • [3] Improvements to the Accuracy of Atmospheric Oxidized Mercury Measurements
    Lyman, Seth N.
    Gratz, Lynne E.
    Dunham-Cheatham, Sarrah M.
    Gustin, Mae Sexauer
    Luippold, Adriel
    [J]. ENVIRONMENTAL SCIENCE & TECHNOLOGY, 2020, 54 (21) : 13379 - 13388
  • [4] Uncertainty Assessment of Gaseous Oxidized Mercury Measurements Collected by Atmospheric Mercury Network
    Cheng, Irene
    Zhang, Leiming
    [J]. ENVIRONMENTAL SCIENCE & TECHNOLOGY, 2017, 51 (02) : 855 - 862
  • [5] Traceable measurements and calibration: a primer on uncertainty analysis
    Csavina, Janae
    Roberti, Joshua A.
    Taylor, Jeffrey R.
    Loescher, Henry W.
    [J]. ECOSPHERE, 2017, 8 (02):
  • [6] Fast Time Resolution Oxidized Mercury Measurements during the Reno Atmospheric Mercury Intercomparison Experiment (RAMIX)
    Ambrose, Jesse L.
    Lyman, Seth N.
    Huang, Jiaoyan
    Gustin, Mae S.
    Jaffe, Daniel A.
    [J]. ENVIRONMENTAL SCIENCE & TECHNOLOGY, 2013, 47 (13) : 7285 - 7294
  • [7] LIDAR MEASUREMENTS OF ATMOSPHERIC MERCURY
    EDNER, H
    SVANBERG, S
    [J]. WATER AIR AND SOIL POLLUTION, 1991, 56 : 131 - 139
  • [8] Calibration of atmospheric hydrogen measurements
    Jordan, A.
    Steinberg, B.
    [J]. ATMOSPHERIC MEASUREMENT TECHNIQUES, 2011, 4 (03) : 509 - 521
  • [9] Traceable Determination of Atmospheric Mercury Using Iodinated Activated Carbon Traps
    Zivkovic, Igor
    Berisha, Sabina
    Kotnik, Joze
    Jagodic, Marta
    Horvat, Milena
    [J]. ATMOSPHERE, 2020, 11 (08)
  • [10] Primary mercury gas standard for the calibration of mercury measurements
    de Krom, Iris
    Bavius, Wijnand
    Ziel, Ruben
    Efremov, Evtim
    van Meer, Davina
    van Otterloo, Peter
    van Andel, Inge
    van Osselen, Deborah
    Heemskerk, Maurice
    van der Veen, Adriaan M. H.
    Dexter, Matthew A.
    Corns, Warren T.
    Ent, Hugo
    [J]. MEASUREMENT, 2021, 169