UNDERSTANDING THE BIOMECHANICAL RESPONSE OF PROGRESSIVE THREAD DENTAL IMPLANTS USING MULTI-SCALE FINITE ELEMENT ANALYSIS

被引:0
|
作者
Chakraborty, Arindam [1 ]
Sahare, Kalash Darshan [1 ,2 ]
Majumder, Santanu [1 ]
Chowdhury, Amit Roy [1 ]
机构
[1] Indian Inst Engn Sci & Technol, Dept Aerosp Engn & Appl Mech, Sibpur 711103, W Bengal, India
[2] Indian Inst Technol, Dept Aerosp Engn, Kanpur, Uttar Pradesh, India
关键词
micro finite element analysis; dental implant; thread depth; bone strain; implant stress; MICRO-CT; STRESS-DISTRIBUTION; CORTICAL BONE; ELASTIC PROPERTIES; DESIGN; FIXATION; QUALITY; FE;
D O I
暂无
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
Implant osseointegration is an important factor dictating its long-term efficacy in situ. Along with various biological factors, it is greatly influenced by the mechanical stimulus at the peri-implant bone. The present study aims to understand the biomechanical response of progressive thread dental implants using multi-scale-based finite element analysis employing macro and micro models of bone. mu-CT images of a cadaveric human mandible of its premolar region were obtained, along with CT scan of the same region to generate computational models. Total of six dental implants were designed having regular and progressive thread depths. Three different stages of healing of the bone-implant assembly were simulated parametrically. The biomechanical environment at the peri-implant bone was analyzed considering the 'Mechanostat' hypothesis. The obtained results revealed that bone strain is significantly higher during the initial healing phase when the bone is weakest. During this phase, implant stress and its displacement in both buccolingual and coronoapical directions are also noticeably higher. Also, displacements of progressive thread implants were lower in all the healing phases as compared to the implants with constant thread depth. The observations of this mu FEA study highlights the clinical applicability of a progressive thread dental implant as it generates larger functional surface area, thus engages higher trabeculae and therefore is suitable for weaker bone conditions. Furthermore, by comparing the stress values at bone and implant between the two bone models, the CT-based model having inhomogeneous material was deemed suitable as an alternative to computationally expensive mu FEA.
引用
收藏
页码:31 / 44
页数:14
相关论文
共 50 条
  • [1] UNDERSTANDING THE BIOMECHANICAL RESPONSE OF PROGRESSIVE THREAD DENTAL IMPLANTS USING MULTI-SCALE FINITE ELEMENT ANALYSIS
    Chakraborty A.
    Sahare K.D.
    Majumder S.
    Chowdhury A.R.
    International Journal for Multiscale Computational Engineering, 2024, 22 (4 Special Issue) : 31 - 44
  • [2] Multi-scale Finite Element Analysis on Biomechanical Response of Functionally Graded Dental Implant / Mandible System
    Tang, C. Y.
    Guo, Y. Q.
    Tsui, C. P.
    Gao, B.
    JOURNAL OF THE SERBIAN SOCIETY FOR COMPUTATIONAL MECHANICS, 2007, 1 (01) : 164 - 173
  • [3] Probing the Influence of Hybrid Thread Design on Biomechanical Response of Dental Implants: Finite Element Study and Experimental Validation
    Chakraborty, Arindam
    Sahare, Kalash Darshan
    Datta, Pallab
    Majumder, Santanu
    Roychowdhury, Amit
    Basu, Bikramjit
    JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 2023, 145 (01):
  • [4] Mechanical response of taper dental implants using finite element analysis
    Jain, Vinamra
    Dommeti, Vamsi Krishna
    Nutu, Emil
    Merdji, Ali
    Biswas, Jayanta Kumar
    Roy, Sandipan
    3RD INTERNATIONAL CONFERENCE ON ADVANCES IN MECHANICAL ENGINEERING (ICAME 2020), PTS 1-6, 2020, 912
  • [5] Biomechanical Analysis of Axial Gradient Porous Dental Implants: A Finite Element Analysis
    Zhang, Chunyu
    Wang, Yuehong
    JOURNAL OF FUNCTIONAL BIOMATERIALS, 2023, 14 (12)
  • [6] Multi-scale finite element simulation of progressive damage in composite structures
    Yu, Qing
    Wu, Jer-Fang
    Proceedings of the 25th International Conference on Offshore Mechanics and Arctic Engineering, Vol 3, 2006, : 413 - 422
  • [7] A 3D Finite Element Analysis Of Dental Implants With Varying Thread Angles
    NarendraKumar, U.
    Mathew, Arun Tom
    Iyer, Nikhil
    Rahman, Febin
    Manjubala, I.
    MATERIALS TODAY-PROCEEDINGS, 2018, 5 (05) : 11900 - 11905
  • [8] Biomechanical influence of thread form on stress distribution over short implants (≤6 mm) using finite element analysis
    Liu, Fan
    Mao, Zhi-Hong
    Peng, Wujie
    Wen, Sheng
    Biomedizinische Technik, 2022, 67 (01): : 53 - 60
  • [9] Biomechanical influence of thread form on stress distribution over short implants (≤6 mm) using finite element analysis
    Liu, Fan
    Mao, Zhi-Hong
    Peng, Wujie
    Wen, Sheng
    BIOMEDICAL ENGINEERING-BIOMEDIZINISCHE TECHNIK, 2022, 67 (01): : 53 - 60
  • [10] Biomechanical finite element analysis of small diameter and short dental implants: extensive study of commercial implants
    Bourauel, Christoph
    Aitlahrach, Maria
    Heinemann, Friedhelm
    Hasan, Istabrak
    BIOMEDICAL ENGINEERING-BIOMEDIZINISCHE TECHNIK, 2012, 57 (01): : 21 - 32