Calibrated intraoral scan protocol (CISP) for full-arch implant impressions: An in vitro comparison to conventional impression, intraoral scan, and intraoral scan with scan-aid

被引:0
|
作者
Li, Junying [1 ]
Chen, Zhaozhao [2 ]
Nava, Paolo [2 ]
Yang, Shengtao [1 ,3 ]
Calatrava, Javier [2 ,4 ]
Wang, Hom-Lay [2 ]
机构
[1] Univ Michigan, Sch Dent, Dept Biol & Mat Sci, Div Prosthodont, 1011 North Univ Ave, Ann Arbor, MI 48109 USA
[2] Univ Michigan, Sch Dent, Dept Periodont & Oral Med, Ann Arbor, MI USA
[3] Sichuan Univ, West China Hosp Stomatol, Natl Ctr Stomatol,Dept Dent Technol, Natl Clin Res Ctr Oral Dis,State Key Lab Oral Dis, Chengdu, Peoples R China
[4] Univ Complutense, Fac Odontol, Sect Grad Periodontol, Madrid, Spain
关键词
accuracy; dental implant; dental impression; full-arch; intraoral scanner; ACCURACY; PROSTHESES; FIT;
D O I
10.1111/cid.13338
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Objective To assess a newly developed intraoral scan protocol in enhancing the accuracy of complete-arch implant impressions. Materials and Methods Four impression approaches were applied to the same maxillary edentulous model with 6 implants: (1) intraoral scan (IOS), (2) intraoral scan with scan aid (IOS-SA), (3) calibrated intraoral scan protocol (CISP), and (4) conventional splinted open-tray impression (CONV). Each approach was repeated 10 times, and a direct scan of the model with a desktop scanner was used as a reference model. The alignment of scans and the reference model was conducted by two methods: (a) aligning all scan bodies to evaluate the overall fit, and (b) aligning the first and second scan bodies to simulate the Sheffield fit test for passive fitting of multiple implant-supported prostheses. Linear deviations from the reference model (trueness) and within each group (precision) were analyzed using Python scripts. Results When aligned by all scan bodies, the CISP group exhibited comparable mean trueness (38.33 mu m) and precision (45.97 mu m) to the CONV group (44.30 and 47.92 mu m respectively), both of which significantly outperformed the IOS group (86.82 and 83.17 mu m, respectively). Furthermore, in the virtual Sheffield fit test, the CISP group achieved the highest levels of mean trueness at the end span (121.7 mu m), making a linear deviation reduction of 36.7%, 60%, and 41.4% when compared to the CONV, the IOS, and the IOS-SA groups, respectively. Moreover, the CISP group (104.3 mu m) displayed a remarkable 65, 182, and 86 mu m advantage in precision over the CONV, IOS, and IOS-SA groups, respectively. Conclusion CISP demonstrated comparable accuracy to the gold standard, the conventional splinted open-tray impression. Furthermore, it excelled in the virtual passive fitting test.
引用
收藏
页数:10
相关论文
共 50 条
  • [1] Evaluation of Intraoral Full-Arch Scan versus Conventional Preliminary Impression
    Janosi, Kinga Maria
    Cerghizan, Diana
    Martha, Krisztina Ildiko
    Elekes, Eva
    Szakacs, Brigitta
    Elekes, Zoltan
    Kovacs, Alpar
    Szasz, Andrea
    Muresan, Izabella
    Hantoiu, Liana Georgiana
    [J]. JOURNAL OF CLINICAL MEDICINE, 2023, 12 (17)
  • [2] Accuracy of edentulous full-arch implant impression: An in vitro comparison between conventional impression, intraoral scan with and without splinting, and photogrammetry
    Cheng, Jing
    Zhang, Haidong
    Liu, Hailin
    Li, Junying
    Wang, Hom-Lay
    Tao, Xian
    [J]. CLINICAL ORAL IMPLANTS RESEARCH, 2024, 35 (05) : 560 - 572
  • [3] Effects of the intraoral scanner and implant library on the trueness of digital impressions in the full-arch implant scan: A comparative in vitro study
    Hauschild, Uli
    Lerner, Henriette
    Weigl, Paul
    Porra, Tommaso
    Admakin, Oleg
    Mangano, Francesco Guido
    [J]. JOURNAL OF DENTISTRY, 2024, 150
  • [4] The Accuracy of Full-Arch Intraoral Optical Impressions (IOS): Clinical Pilot Study of the Influence of the Scan Strategy, Operator, and Intraoral Scanner
    Cordaro, Matteo
    Sailer, Irena
    Zarauz, Cristina
    Liu, Xinran
    Karasan, Duygu
    [J]. INTERNATIONAL JOURNAL OF PROSTHODONTICS, 2023, 36 (06) : 689 - 696
  • [5] Comparison of the accuracy of intraoral scans between complete-arch scan and quadrant scan
    Moon, Youn-Gyeong
    Lee, Kyung-Min
    [J]. PROGRESS IN ORTHODONTICS, 2020, 21 (01)
  • [6] Comparison of the accuracy of intraoral scans between complete-arch scan and quadrant scan
    Youn-Gyeong Moon
    Kyung-Min Lee
    [J]. Progress in Orthodontics, 21
  • [7] Impact of Different Scan Bodies and Scan Strategies on the Accuracy of Digital Implant Impressions Assessed with an Intraoral Scanner: An In Vitro Study
    Motel, Constantin
    Kirchner, Elena
    Adler, Werner
    Wichmann, Manfred
    Matta, Ragai Edward
    [J]. JOURNAL OF PROSTHODONTICS-IMPLANT ESTHETIC AND RECONSTRUCTIVE DENTISTRY, 2020, 29 (04): : 309 - 314
  • [8] Accuracy of intraoral optical scan versus stereophotogrammetry for complete-arch digital implant impression: An in vitro study
    Pozzi, Alessandro
    Agliardi, Enrico
    Lio, Fabrizio
    Nagy, Katalin
    Nardi, Alessandra
    Arcuri, Lorenzo
    [J]. JOURNAL OF PROSTHODONTIC RESEARCH, 2024, 68 (01) : 172 - 180
  • [9] Conventional open-tray impression versus intraoral digital scan for implant-level complete-arch impression
    Kim, Kyoung Rok
    Seo, Kyoung-young
    Kim, Sunjai
    [J]. JOURNAL OF PROSTHETIC DENTISTRY, 2019, 122 (06): : 543 - 549
  • [10] Feasibility of using an intraoral scanner for a complete arch digital scan, part 2: A comparison of scan strategies
    Son, KeunBaDa
    Jin, Myoung-Uk
    Lee, Kyu-Bok
    [J]. JOURNAL OF PROSTHETIC DENTISTRY, 2023, 129 (02): : 341 - 349