Design and testing of an irradiation room with low room-scattering for neutron calibration

被引:0
|
作者
Liu, Haixia [1 ]
Sun, Tao [1 ]
Li, Chongwei [1 ]
Wang, Dongxi [1 ]
Guan, Xian [1 ]
Zhou, Hongzhao [1 ]
机构
[1] State Key Lab NBC Protect Civilian, Beijing 102205, Peoples R China
关键词
Room-scattering; Neutron calibration; Irradiation room; Material prescription; Maximum available calibration distance; Shadow cone; FACILITY; FIELD;
D O I
10.1016/j.apradiso.2024.111402
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
Introduction: This laboratory plans to establish neutron reference radiation fields with three neutron sources to calibrate neutron-measuring devices. To perform calibration at multiple dose rates, neutron ambient dose equivalent rate (center dot) H & lowast; (10 ) needs to range 1 mu Sv/h to 10 mSv/h. The lower limit requires that the maximum available calibration distance should be at least 4.5 m. Methods: To reduce room-scattered neutrons and extend the available calibration distance, MC simulations were conducted to determine the material and size of the irradiation room. A 3 '' Bonner sphere and a LB6411 environmental neutron dosimeter were used to characterize the irradiation room. Results: A 14.32 x 14.32 x 12.00 m 3 irradiation room was built based on simulation results. Floor, roof, and walls are made of 75 cm concrete covered by a coating layer of 2 cm BPE and 3 cm PE. Experimental maximum available calibration distance reaches 4.65 m. The range of (center dot) H & lowast; (10 ) for calibration covers 1 mu Sv/h to 10 mSv/h. Neutron and photon (center dot) H & lowast; (10 ) outside the room are within 0.19 mu Sv/h and 0.22 mu Sv/h, respectively.
引用
收藏
页数:6
相关论文
共 50 条
  • [1] Calibration of neutron personal dosimeters in a realistically simulated neutron irradiation room
    Kim, J. H.
    Park, H.
    Choi, K. -O.
    RADIATION MEASUREMENTS, 2010, 45 (10) : 1544 - 1547
  • [2] Determination of neutron room scattering corrections in RCL's calibration facility at KAERI
    Yoon, SC
    Chang, SY
    Kim, JS
    Kim, JL
    RADIATION PROTECTION DOSIMETRY, 1997, 70 (1-4) : 349 - 351
  • [3] Radiation protection aspects in the design of a Boron Neutron CaptureTherapy irradiation room
    Magni, Chiara
    Ferrarini, Michele
    Postuma, Ian
    Vercesi, Valerio
    Fatemi, Setareh
    Ramos, Ricardo Luis
    Bortolussi, Silva
    RADIATION PHYSICS AND CHEMISTRY, 2024, 218
  • [4] Calibration of Spherical Neutron Measuring Devices with a 252Cf Source in the Neutron Calibration Room of the KAERI
    Jun, Sun-Mee
    Kim, Bong-Hwan
    Kim, Jang-Lyul
    Lee, Jai Ki
    JOURNAL OF NUCLEAR SCIENCE AND TECHNOLOGY, 2008, : 286 - 289
  • [5] Shielding design of a treatment room for an accelerator-based epithermal neutron irradiation facility for BNCT
    Evans, JF
    Blue, TE
    HEALTH PHYSICS, 1996, 71 (05): : 692 - 699
  • [7] Design of performance testing room for refrigerating container
    Wu Ruofei
    Liu Jianhua
    IEEE/SOLI'2008: PROCEEDINGS OF 2008 IEEE INTERNATIONAL CONFERENCE ON SERVICE OPERATIONS AND LOGISTICS, AND INFORMATICS, VOLS 1 AND 2, 2008, : 1636 - 1639
  • [8] EFFECT OF LOW DOSE NEUTRON IRRADIATION ON TENSILE BEHAVIOR OF HT-9 STEEL AT ROOM TEMPERATURE
    Sarkar, A.
    Alsabbagh, A. H.
    Murty, K. L.
    TMS 2014 SUPPLEMENTAL PROCEEDINGS, 2014, : 147 - 154
  • [10] Experimental Study of a New Design of a Freezer Testing Room
    Panyakunanont, Pacharapong
    Wongsuwan, Wipawadee
    Anankawanich, Narong
    2022 7TH INTERNATIONAL CONFERENCE ON BUSINESS AND INDUSTRIAL RESEARCH (ICBIR2022), 2022, : 295 - 300