CONTINUOUS DIRECT AND INDIRECT BLOOD-PRESSURE MEASUREMENT (FINAPRES) IN THE CRITICALLY ILL

被引:20
|
作者
FARQUHAR, IK [1 ]
机构
[1] UNIV NOTTINGHAM HOSP, DEPT ANAESTHESIA, NOTTINGHAM NG7 2UH, ENGLAND
关键词
MEASUREMENT TECHNIQUES; PLETHYSMOGRAPHY; EQUIPMENT; FINAPRES;
D O I
10.1111/j.1365-2044.1991.tb09922.x
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Continuous noninvasive blood pressure measurement is of great potential use in the critically ill. This study was designed to find out whether measurements of blood pressure by a Finapres accurately represented intra‐arterial pressure. Comparisons were made between readings from a radial artery line and from a Finapres finger cuff in 10 critically ill patients. There was an unpredictable but usually stable difference between measurements from the two sources; this difference varied in magnitude and direction in individual patients. A t present the Finapres cannot be recommended as a substitute for intra‐arterial blood pressure measurement in the critically ill. Copyright © 1991, Wiley Blackwell. All rights reserved
引用
收藏
页码:1050 / 1055
页数:6
相关论文
共 50 条
  • [1] DIRECT VERSUS INDIRECT BLOOD-PRESSURE MEASUREMENTS IN CRITICALLY ILL PATIENTS
    VENUS, B
    MATHRU, M
    SMITH, RA
    PHAM, CG
    HEART & LUNG, 1985, 14 (03): : 228 - 231
  • [2] DIRECT VERSUS INDIRECT MEASUREMENT OF BLOOD-PRESSURE
    RAFTERY, EB
    JOURNAL OF HYPERTENSION, 1991, 9 : S10 - S12
  • [3] ACCURACY OF OSCILLOMETRIC BLOOD-PRESSURE MEASUREMENT IN CRITICALLY ILL NEONATES
    GEVERS, M
    VANGENDERINGEN, HR
    LAFEBER, HN
    HACK, WW
    PEDIATRIC RESEARCH, 1994, 36 (01) : A15 - A15
  • [4] BLOOD-PRESSURE MEASUREMENT DURING TRANSPORT - A COMPARISON OF DIRECT AND OSCILLOTONOMETRIC READINGS IN CRITICALLY ILL PATIENTS
    RUNCIE, CJ
    REEVE, WG
    REIDY, J
    DOUGALL, JR
    ANAESTHESIA, 1990, 45 (08) : 659 - 665
  • [5] MEASUREMENT OF BLOOD-PRESSURE IN OBESE PERSONS - COMPARISON OF INTRAARTERIAL AND FINGER ARTERIAL BLOOD-PRESSURE (FINAPRES)
    PONCELET, P
    DURAND, P
    LECHANTRE, R
    MOUNIERVEHIER, C
    FOUQUOIRE, B
    PETETIN, N
    CARRE, A
    ARCHIVES DES MALADIES DU COEUR ET DES VAISSEAUX, 1992, 85 (08): : 1165 - 1168
  • [6] 3 METHODS FOR INDIRECT BLOOD-PRESSURE MEASUREMENT IN ANIMALS REVIEWED CRITICALLY
    WERNER, J
    KLEINTIERPRAXIS, 1980, 25 (08): : 485 - &
  • [7] AN EVALUATION OF 2 INDIRECT METHODS OF BLOOD-PRESSURE MEASUREMENT IN ILL PATIENTS
    REBENSONPIANO, M
    HOLM, K
    FOREMAN, MD
    KIRCHHOFF, KT
    NURSING RESEARCH, 1989, 38 (01) : 42 - 45
  • [8] COMPARISON OF DIRECT AND INDIRECT BLOOD-PRESSURE MEASUREMENT IN ANESTHETIZED DOGS
    SAWYER, DC
    BROWN, M
    STRILER, EL
    DURHAM, RA
    LANGHAM, MA
    RECH, RH
    LABORATORY ANIMAL SCIENCE, 1991, 41 (02): : 134 - 138
  • [9] VARIABILITY IN INDIRECT BLOOD-PRESSURE MEASUREMENT
    BOTTINI, PB
    CARR, AA
    ARCHIVES OF INTERNAL MEDICINE, 1994, 154 (13) : 1536 - 1536
  • [10] INDIRECT METHODS OF BLOOD-PRESSURE MEASUREMENT
    MOSS, AJ
    PEDIATRIC CLINICS OF NORTH AMERICA, 1978, 25 (01) : 3 - 14