PERFORMANCE OF A PATIENT-DEDICATED, ON-DEMAND BLOOD-GAS MONITOR IN MEDICAL ICU PATIENTS

被引:16
|
作者
MAHUTTE, CK
SASSE, SA
CHEN, PA
HOLODY, M
机构
[1] THREE M HEALTHCARE,CDI,TUSTIN,CA
[2] UNIV CALIF IRVINE,IRVINE,CA
关键词
D O I
10.1164/ajrccm.150.3.8087363
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
We examined the performance characteristics of a new bedside blood gas monitor. This monitor's fluorescent pH, PCO2, and PO2 sensors are embedded in a cassette, which is calibrated in vitro and then inserted into the patient's radial artery tubing set. In 50 medical ICU patients, 683 paired monitor and conventional blood gas analyzer values were obtained. Performance was assessed via calculations of bias (mean monitor and analyzer difference) and its standard deviation (SD), plots of monitor and analyzer differences against the means (of monitor and analyzer), and linear regression analysis of the sequential changes in monitor values versus the corresponding sequential changes in analyzer values. The ex vivo calibration, assessed using the initial paired blood samples, showed a bias +/- SD of 0.02 +/- 0.02 for pH, -0.1 +/- 1.9 mm Hg for PCO2, and 4.3 +/- 6.0 mm Hg for PO2. For all paired samples (n = 683), the biases +/- SD were 0.004 +/- 0.023 for pH, 0.6 +/- 2.4 mm Hg for PCO2, and 2.7 +/- 6.4 mm HG for PO2. The PO2 bias increased as PO2 increased. The standard deviations (imprecision) of both PCO2 and PO2 also increased as the magnitudes of these variables increased. Sequential changes in monitor values versus the corresponding sequential changes in analyzer values revealed regression lines close to the line of identity. Serum sodium had no effect on pH bias. Daily drift of the sensors was inconsequential, with values of -0.01/d for pH, 1.7 mm Hg/d for PCO2, and 1.1 mm Hg/d for PO2. We conclude that the performance of this monitor is comparable to that of conventional blood gas analyzers.
引用
收藏
页码:865 / 869
页数:5
相关论文
共 11 条
  • [1] DEVELOPMENT OF A PATIENT-DEDICATED, ON-DEMAND, BLOOD-GAS MONITOR
    MAHUTTE, CK
    HOLODY, M
    MAXWELL, TP
    CHEN, PA
    SASSE, SA
    AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 1994, 149 (04) : 852 - 859
  • [2] Evaluation of a patient-dedicated blood gas analyser
    JJ Fox
    TH Clutton-Brock
    Critical Care, 19 (Suppl 1):
  • [3] COMPARISON OF CONVENTIONAL BLOOD-GAS ANALYSIS AND AN ON-DEMAND BEDSIDE BLOOD-GAS MONITOR IN THE DETERMINATION OF MIXED VENOUS-BLOOD GAS VALUES
    PERUZZI, WT
    FRANKLIN, ML
    SHAPIRO, BA
    MOEN, SG
    ANESTHESIOLOGY, 1994, 81 (3A) : A620 - A620
  • [4] Performance of an on-demand blood gas monitor in surgical intensive care unit (SICU) patients
    Rothenberg, DM
    Columbo, JA
    Albu, LL
    McCarthy, RJ
    CRITICAL CARE MEDICINE, 1999, 27 (01) : A114 - A114
  • [5] VARIABILITY OF ARTERIAL BLOOD-GAS VALUES OVER TIME IN STABLE MEDICAL ICU PATIENTS
    SASSE, SA
    CHEN, PA
    MAHUTTE, CK
    CHEST, 1994, 106 (01) : 187 - 193
  • [6] PERFORMANCE OF AN INVIVO, CONTINUOUS BLOOD-GAS MONITOR WITH DISPOSABLE PROBE
    MILLER, WW
    YAFUSO, M
    YAN, CF
    HUI, HK
    ARICK, S
    CLINICAL CHEMISTRY, 1987, 33 (09) : 1538 - 1542
  • [7] EVALUATION OF AN INTRAARTERIAL BLOOD-GAS MONITOR IN MECHANICALLY VENTILATED PATIENTS
    BARRERA, R
    LOIACONO, J
    GROEGER, JS
    CARLON, GC
    ANESTHESIOLOGY, 1994, 81 (3A) : A615 - A615
  • [8] Implementation Of An On-Demand Strategy For Routine Blood Testing In Icu Patients
    Iosfina, I.
    Merkeley, H.
    Cessford, T.
    Geller, G.
    Amiri, N.
    Baradaran, N.
    Norena, M.
    Ayas, N.
    Dodek, P. M.
    AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 2013, 187
  • [9] CLINICAL-PERFORMANCE OF A BLOOD-GAS MONITOR - A PROSPECTIVE, MULTICENTER TRIAL
    SHAPIRO, BA
    MAHUTTE, CK
    CANE, RD
    GILMOUR, IJ
    CRITICAL CARE MEDICINE, 1993, 21 (04) : 487 - 494
  • [10] Evaluation of an on-demand, ex vivo bedside blood gas monitor on pulmonary artery blood gas determinations
    Franklin, ML
    Peruzzi, WT
    Moen, SG
    Shapiro, BA
    ANESTHESIA AND ANALGESIA, 1996, 83 (03): : 500 - 504