Evaluation of the Long-Term Stability for the Cylindrical Ionization Chambers

被引:0
|
作者
Rah, Jeong Eun [1 ]
Hong, Ju Young [1 ]
Kim, Gwe Ya [2 ]
Lim, Chun Il [2 ]
Jeong, Hee Kyo [2 ]
Shin, Dong Oh [3 ]
Suh, Tea Suk [1 ]
机构
[1] Catholic Univ Korea, Coll Med, Dept Biomed Engn, Seoul, South Korea
[2] Korea Food & Drug Adm, Radiat Stand Div, Seoul, South Korea
[3] Kyung Hee Univ, Coll Med, Dept Radiat Oncol, Seoul, South Korea
来源
RADIATION ONCOLOGY JOURNAL | 2006年 / 24卷 / 02期
关键词
N-k calibration factor; N-D; N-W calibration factor; Farmer type cylindrical chamber; Long-term stability;
D O I
暂无
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: To analyze the long-term stability of Farmer-type cylindrical ionization chambers by calibration factor provided from the KFDA (Korea Food Drug Administration) Materials and Methods: The cylindrical ionization chambers used in this study were the PTW 30001 (30006), 30013, 30002, 30004, 23333, the Capintec PR06C, the NE 2571, the Exradin A12 and the Wellhofer FC65G (IC70). We were analyzed that the N-k and N-D,N-W calibration factor for the cylindrical chambers and compared between the measured N-D,N-W and calculated N-D,N-W calibration factor. Results: We have observed that the long-term stability of the PTW 30013 (30006), the Wellhofer FC65G (IC70) and the NE 2571 has varied within 0.2%. The measured N-D,N-W calibration factor was about 1.0% higher than the calculated N-D,N-W that determined by the N-k calibration factor. Conclusion: The study has evaluated that the long-term stability of the cylindrical chambers through analysis for the N-k and N-D,N-W calibration factor. It has contributed to the improvement of clinical electron dosimetry in radiotherapy centers.
引用
收藏
页码:138 / 143
页数:6
相关论文
共 50 条
  • [1] Evaluation of the long-term stability for the cylindrical ionization chambers
    Lah, J.
    Lim, C.
    Suh, T.
    [J]. MEDICAL PHYSICS, 2007, 34 (06) : 2452 - 2452
  • [2] Stability of cylindrical shells subject to long-term damage
    Khoroshun L.P.
    Babich D.V.
    [J]. International Applied Mechanics, 2009, 45 (10) : 1092 - 1100
  • [3] Numerical evaluation of long-term stability
    Giovannozzi, M
    Scandale, W
    Todesco, E
    [J]. BEAM STABILITY AND NONLINEAR DYNAMICS, 1997, (405): : 25 - 39
  • [4] EVALUATION OF LONG-TERM MAGNET STABILITY
    ZINGERY, WL
    WHALLEY, WB
    ROMBERG, EB
    WHEELER, FW
    [J]. JOURNAL OF APPLIED PHYSICS, 1966, 37 (03) : 1101 - &
  • [5] Evaluation of the Long-Term Stability of Sheath Plication
    Nathan, Karim
    Nathan, Bassem
    [J]. PLASTIC AND RECONSTRUCTIVE SURGERY, 2013, 131 (05): : 851E - 852E
  • [6] OPTIMUM SHAPE FOR CYLINDRICAL IONIZATION CHAMBERS
    CLARK, RK
    BRAR, SS
    [J]. NUCLEONICS, 1954, 12 (08): : 28 - 29
  • [7] LONG-TERM STABILITY OF 2 TYPES OF HOT CATHODE IONIZATION GAUGES
    WOOD, SD
    TILFORD, CR
    [J]. JOURNAL OF VACUUM SCIENCE & TECHNOLOGY A-VACUUM SURFACES AND FILMS, 1985, 3 (03): : 542 - 545
  • [8] Reply: Evaluation of the Long-Term Stability of Sheath Plication
    Mestak, Ondrej
    Kullac, Robert
    Mestak, Jan
    Nosek, Alex
    Krajcova, Aneta
    Sukop, Andrej
    [J]. PLASTIC AND RECONSTRUCTIVE SURGERY, 2013, 131 (05): : 852E - 852E
  • [9] STABILITY IMPROVEMENT OF IONIZATION CHAMBERS
    YAROM, A
    SABBAH, B
    YAIR, R
    [J]. NUCLEAR INSTRUMENTS & METHODS, 1972, 104 (02): : 257 - &