INTRAVASCULAR MEMBRANE-OXYGENATION AND CARBON-DIOXIDE REMOVAL WITH IVOX - CAN IMPROVED DESIGN AND PERMISSIVE HYPERCAPNIA ACHIEVE ADEQUATE RESPIRATORY SUPPORT DURING SEVERE RESPIRATORY-FAILURE

被引:6
|
作者
ZWISCHENBERGER, JB
CARDENAS, VJ
TAO, WK
NIRANJAN, SC
CLARK, JW
BIDANI, A
机构
[1] UNIV TEXAS,MED BRANCH,DEPT INTERNAL MED,GALVESTON,TX 77550
[2] RICE UNIV,DEPT ELECT & COMP ENGN,HOUSTON,TX 77251
关键词
PERMISSIVE HYPERCAPNIA; ACUTE RESPIRATORY FAILURE; INTRACORPOREAL GAS EXCHANGE; IVOX; ARDS;
D O I
10.1111/j.1525-1594.1994.tb03332.x
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The intravenacaval oxygenator and carbon dioxide removal device (IVOX) conceived by Mortensen at CardioPulmonics is a diffusion-limited device capable of removing 30% of CO2 production of an adult at normocapnia with minimal reduction in ventilator requirements. Through mathematical modeling, an ex vivo venovenous bypass circuit to model the vena cava and animal models of severe smoke inhalation injury, the practice of permissive hypercapnia has been established to enhance CO2 removal by IVOX. By allowing the blood Pco2 to rise gradually, the CO2 excretion by IVOX can be linearly increased in a 1:1 relationship. Experimental and clinical studies have shown that CO2 removal by IVOX increased from 30-40 ml/min at normal blood Pco2 to 80-90 ml/min at Pco2 of 90 mm Hg. In addition, IVOX with permissive hypercapnia allowed a significant reduction in minute ventilation and peak airway pressure. Design changes could also improve the performance of IVOX. Increased surface area and mixing with more fibers and crimping in new prototypes of IVOX significantly increased CO2 removal and oxygen transfer. Active mixing in the blood to decrease the boundary layer resistance can further enhance gas exchange of IVOX. In conclusion, gas exchange by the current design of IVOX is limited, and improvements in design are needed for it to become a more clinically applicable device. Permissive hypercapnia can significantly enhance CO2 removal by IVOX as well as significantly reduce ventilator requirements.
引用
收藏
页码:833 / 839
页数:7
相关论文
共 5 条
  • [1] INTRAVENACAVAL MEMBRANE-OXYGENATION AND CARBON-DIOXIDE REMOVAL IN SEVERE ACUTE RESPIRATORY-FAILURE
    CONRAD, SA
    EGGERSTEDT, JM
    GRIER, LR
    MORRIS, VF
    ROMERO, MD
    CHEST, 1995, 107 (06) : 1689 - 1697
  • [2] INTRAVASCULAR MEMBRANE-OXYGENATION AND CARBON-DIOXIDE REMOVAL - A NEW APPLICATION FOR PERMISSIVE HYPERCAPNIA
    ZWISCHENBERGER, JB
    COX, CS
    GRAVES, D
    BIDANI, A
    THORACIC AND CARDIOVASCULAR SURGEON, 1992, 40 (03): : 115 - 120
  • [3] Extracorporeal carbon dioxide removal may prevent the need for extracorporeal membrane oxygenation in severe respiratory failure A cohort study
    Pestana, David
    Gomis, Antonio
    de Pablo, Raul
    Tenorio, Maria T.
    EUROPEAN JOURNAL OF ANAESTHESIOLOGY, 2020, 37 (10) : 950 - 952
  • [4] SEVERE ACUTE RESPIRATORY-FAILURE MANAGED WITH CONTINUOUS POSITIVE AIRWAY PRESSURE AND PARTIAL EXTRACORPOREAL CARBON-DIOXIDE REMOVAL BY AN ARTIFICIAL MEMBRANE LUNG - A CONTROLLED, RANDOMIZED ANIMAL STUDY
    BORELLI, M
    KOLOBOW, T
    SPATOLA, R
    PRATO, P
    TSUNO, K
    AMERICAN REVIEW OF RESPIRATORY DISEASE, 1988, 138 (06): : 1480 - 1487
  • [5] Reduced ventilator pressure and improved P/F ratio during percutaneous arteriovenous carbon dioxide removal for severe respiratory failure
    Alpard, SK
    Zwischenberger, JB
    Tao, WK
    Deyo, DJ
    Bidani, A
    ANNALS OF SURGERY, 1999, 230 (02) : 215 - 224