THE ACCURACY OF A 3-COMPONENT LASER DOPPLER-VELOCIMETER SYSTEM USING A SINGLE-LENS APPROACH

被引:1
|
作者
CHEVRIN, PA [1 ]
PETRIE, HL [1 ]
DEUTSCH, S [1 ]
机构
[1] PENN STATE UNIV,APPL RES LAB,STATE COLL,PA 16804
关键词
D O I
10.1115/1.2910097
中图分类号
TH [机械、仪表工业];
学科分类号
0802 ;
摘要
An analysis of the effects of misalignment on the on-axis velocity component measurements made with a three-component LDV system using two colors and five beams issuing from a single lens is discussed. A slight misalignment of the center beam from the optical axis is shown to cause large changes in the measured ''on-axis'' component of velocity. The effect of misalignment on the ensemble averaged velocity statistics in turbulent flows is also considered. The error analysis of Neti and Clark (1979) is applicable to on-axis component of the present system and it shows that even when the system is perfectly aligned, the relative uncertainty of the on-axis component of velocity is much larger than that of the other components of velocity. Measurements taken in a laminar pipe flow support this conclusion.
引用
收藏
页码:142 / 147
页数:6
相关论文
共 50 条
  • [1] 3-COMPONENT OPTICAL DOPPLER-VELOCIMETER
    KULYBIN, VM
    [J]. IZVESTIYA VYSSHIKH UCHEBNYKH ZAVEDENII FIZIKA, 1975, (10): : 82 - 86
  • [2] A CORRECTION LENS FOR LASER DOPPLER-VELOCIMETER MEASUREMENTS IN A CYLINDRICAL TUBE
    DURRETT, RP
    GOULD, RD
    STEVENSON, WH
    THOMPSON, HD
    [J]. AIAA JOURNAL, 1985, 23 (09) : 1387 - 1391
  • [3] 2 COMPONENT LASER DOPPLER-VELOCIMETER FOR PERIODIC FLOW FIELDS
    SULLIVAN, JP
    EZEKIEL, S
    [J]. JOURNAL OF PHYSICS E-SCIENTIFIC INSTRUMENTS, 1974, 7 (04): : 272 - 274
  • [4] REMOTE MEASUREMENT OF THE TRANSVERSE WIND VELOCITY COMPONENT USING A LASER DOPPLER-VELOCIMETER
    KENNEDY, LZ
    BILBRO, JW
    [J]. APPLIED OPTICS, 1979, 18 (17): : 3010 - 3013
  • [5] MEASUREMENT OF IMPACT LOADS USING A LASER DOPPLER-VELOCIMETER
    BIRCH, RS
    JONES, N
    [J]. PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART C-JOURNAL OF MECHANICAL ENGINEERING SCIENCE, 1990, 204 (01) : 1 - 8
  • [6] DUAL SINGLE-BEAM FIBER LASER DOPPLER-VELOCIMETER
    SASAKI, O
    WATANABE, T
    [J]. APPLIED OPTICS, 1981, 20 (23): : 3990 - 3991
  • [7] A COMPACT DIFFERENTIAL LASER DOPPLER-VELOCIMETER USING A SEMICONDUCTOR-LASER
    JENTINK, HW
    VANBEURDEN, JAJ
    HELSDINGEN, MA
    DEMUL, FFM
    SUICHIES, HE
    AARNOUDSE, JG
    GREVE, J
    [J]. JOURNAL OF PHYSICS E-SCIENTIFIC INSTRUMENTS, 1987, 20 (10): : 1281 - 1283
  • [8] LASER DOPPLER-VELOCIMETER AS AN OPTOELECTRONIC DATA-PROCESSING SYSTEM
    DUBNITSHEV, YN
    KORONKEVITCH, VP
    SOBOLEV, VS
    STOLPOVSKI, AA
    VASILENKO, YG
    UTKIN, EN
    [J]. APPLIED OPTICS, 1975, 14 (01): : 180 - 184
  • [9] VERIFICATION OF 3-DIMENSIONAL LASER DOPPLER-VELOCIMETER MEASUREMENTS
    MEHTA, RD
    [J]. EXPERIMENTS IN FLUIDS, 1990, 8 (06) : 354 - 356
  • [10] ACCURACY IN BLOOD VELOCITY-MEASUREMENT BY LASER DOPPLER-VELOCIMETER WITH OPTICAL FIBER
    MITO, K
    HOKI, N
    TOMONAGA, G
    KAGIYAMA, M
    HIRAMATSU, O
    KAJIYA, F
    [J]. BIORHEOLOGY, 1982, 19 (1-2) : 376 - 377