PHYSIOLOGICAL CHROMIUM DETERMINATION IN SERUM BY ZEEMAN GRAPHITE-FURNACE ATOMIC-ABSORPTION SPECTROMETRY - A SERIOUS CHALLENGE

被引:8
|
作者
CHAPPUIS, P
POUPON, J
DESCHAMPS, JF
GUILLAUSSEAU, PJ
ROUSSELET, F
机构
[1] HOP LARIBOISIERE,HEMOBIOL TRANSFUS LAB,F-75475 PARIS 10,FRANCE
[2] HOP LARIBOISIERE,SERV MED INTERNE,F-75475 PARIS 10,FRANCE
关键词
CHROMIUM; SERUM; ATOMIC ABSORPTION SPECTROMETRY;
D O I
10.1007/BF02784591
中图分类号
Q5 [生物化学]; Q7 [分子生物学];
学科分类号
071010 ; 081704 ;
摘要
Several authors have already underlined chromium implication in glucose and lipids metabolism of humans. In this field, physiological chromium determination in serum could be helpful, but the discrepancies reported in numerous papers are confusing. Here we report some results obtained by Zeeman correction Graphite Furnace Atomic Absorption Spectrometry. This technique includes a dilution of serum with 12.5 mM ultrapure nitric acid and 0.25% Triton X-100 (final concentrations). Some main features can be outlined: (1) the contamination constitutes a serious drawback and (2) the sensitivity of the technique is critical (characteristic mass found: 1.76 pg/0.0044 A.s). Our results obtained from 27 healthy subjects (2.01 +/- 0.77 nmol/L) agree with most recent studies and indicate that serum chromium level does not seem to be sex-related.