ERRORS ARISING FROM INCORRECT ORIENTATION OF E-TEST STRIPS

被引:1
|
作者
HAMILTONMILLER, JMT
SHAH, S
YAM, TS
机构
关键词
D O I
10.1128/JCM.33.7.1966-1967.1995
中图分类号
Q93 [微生物学];
学科分类号
071005 ; 100705 ;
摘要
Major errors arise if E test strips are placed upside down. Asymmetric zones, or no zone at all, may result. MICs indicated by upside-down tests were almost always considerably higher than true values. This situation differs markedly from that for conventional testing, where orientation of disks is not important.