QUALITY ASSURANCE OF ULTRASOUND IMAGING INSTRUMENTS BY MONITORING THE MONITOR

被引:1
|
作者
WALKER, JB [1 ]
THORNE, GC [1 ]
HALLIWELL, M [1 ]
机构
[1] BRISTOL GEN HOSP,DEPT MED PHYS & BIOENGN,BRISTOL BS1 6SY,AVON,ENGLAND
来源
PHYSICS IN MEDICINE AND BIOLOGY | 1993年 / 38卷 / 11期
关键词
D O I
10.1088/0031-9155/38/11/005
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Ultrasound quality assurance (QA) is a means of assuring the constant performance of an ultrasound instrument. A novel 'ultrasound image analyser' has been developed to allow objective, accurate and repeatable measurement of the image displayed on the ultrasound screen, i.e. as seen by the operator. The analyser uses a television camera/framestore combination to digitize and analyse this image. A QA scheme is described along with the procedures necessary to obtain a repeatable measurement of the image so that comparisons with earlier good images can be made. These include repositioning the camera and resetting the video display characteristics. The advantages of using the analyser over other methods are discussed. It is concluded that the analyser has distinct advantages over subjective image assessment methods and will be a valuable addition to current ultrasound QA programmes.
引用
收藏
页码:1601 / 1609
页数:9
相关论文
共 50 条
  • [1] Diagnostic Ultrasound Imaging Quality Assurance
    Zagzebski, J.
    [J]. MEDICAL PHYSICS, 2016, 43 (06) : 3800 - 3800
  • [2] Quality assurance for the ultrasound
    Bolvig, Lars
    [J]. ULTRASCHALL IN DER MEDIZIN, 2006, 27 (06): : 597 - 597
  • [3] Reference materials - Instruments of quality assurance
    Nehls, I
    [J]. NACHRICHTEN AUS DER CHEMIE, 2005, 53 (09) : 925 - 927
  • [4] AN EASY METHOD TO MONITOR THE CALIBRATION OF ULTRASOUND INSTRUMENTS
    HAUGEN, JP
    STAMPER, RL
    [J]. OPHTHALMIC SURGERY AND LASERS, 1987, 18 (05): : 398 - 398
  • [5] INTRAVASCULAR ULTRASOUND IMAGING AS A TOOL FOR QUALITY ASSURANCE DURING CAROTID ENDARTERECTOMY
    KARNIK, R
    AMMERER, HP
    WINKLER, WB
    VALENTIN, A
    SLANY, J
    [J]. STROKE, 1993, 24 (03) : 500 - 500
  • [6] QUALITY ASSURANCE IN OBSTETRIC ULTRASOUND
    DUDLEY, NJ
    POTTER, R
    [J]. BRITISH JOURNAL OF RADIOLOGY, 1993, 66 (790): : 865 - 870
  • [7] Quality assurance in diagnostic ultrasound
    Sipila, Outi
    Mannila, Vilma
    Vartiainen, Eija
    [J]. EUROPEAN JOURNAL OF RADIOLOGY, 2011, 80 (02) : 519 - 525
  • [8] A COMPARISON OF 2 QUALITY ASSESSMENT INSTRUMENTS - MONITOR AND SENIOR MONITOR
    TOMALIN, DA
    REDFERN, SJ
    NORMAN, IJ
    OLIVER, S
    [J]. INTERNATIONAL JOURNAL OF NURSING STUDIES, 1994, 31 (03) : 299 - 311
  • [9] The validity of two quality assessment instruments: Monitor and Senior Monitor
    Norman, IJ
    Redfern, SJ
    [J]. INTERNATIONAL JOURNAL OF NURSING STUDIES, 1996, 33 (06) : 660 - 668
  • [10] Status and evaluation of internal quality assurance instruments
    Schönberg, F
    [J]. ZEITSCHRIFT FUR GERONTOLOGIE UND GERIATRIE, 2004, 37 : 33 - 33