CHARACTERISTICS OF SODIUM ION-SELECTIVE ELECTRODES BASED ON NEUTRAL CARRIERS AND THEIR APPLICATION TO ANALYSIS OF PHYSIOLOGICAL FLUIDS

被引:6
|
作者
SHIBATA, Y
MARUIZUMI, T
MIYAGI, H
机构
[1] Central Research Raboratory, Hitachi Ltd.
关键词
D O I
10.1246/nikkashi.1992.961
中图分类号
O6 [化学];
学科分类号
0703 ;
摘要
Characteristics of sodium ion-selective electrodes (Na-ISEs) with polymeric membranes incorporating cyclic and acyclic compounds as ionophores were studied and their applicability to analysis of physiological fluids was examined. In this study, bis[(12-crown-4)methyl] methyl(dodecyl)malonate [bis(12-crown-4)] and polyether-amide compounds (ETH 157, ETH 2120) were used as ionophores and adipic esters [bis(2-ethylhexyl) adipate, bis(1-butylpentyl) adipate] were used as plasticizers. These Na-ISEs showed Nernstian slope of 54 approximately 57 mV/decade and response time of < 10 s. Among three ionophores examined, bis(12-crown-4) was the most selective to Na+ in the presence of drugs [1, 1-dimethyl-5-methoxy-3-(di-2-thienylmethylene)piperidinium bromide monohydrate and [2-(alpha, alpha-dicyclopentyl acetoxy)ethyl] triethylammonium bromide] and lipophilic quaternary ammonium salts such as a cationic surfactant (trimethyloctadecylammonium salt) as well as cations. The cause of interference to three Na-ISEs by the above two drugs was studied. Then, the migration of interfering substances into the sensing membrane was suggested by the measurement of potential response characteristics. In the analysis of control serum samples, good reproducibility (C. V. = 0.3 approximately 0.4%) was obtained by all Na-ISEs. But accuracy in the analysis of patient serum samples was varied with the type of ionophore of Na-ISE. The deviation of error (S(yx) was as
引用
收藏
页码:961 / 967
页数:7
相关论文
共 50 条
  • [1] Hydrogen ion-selective membrane electrodes based on alkyldibenzylamines as neutral carriers
    Cho, DH
    Chung, KC
    Park, MY
    TALANTA, 1998, 47 (04) : 815 - 821
  • [2] APPLICATION OF AZOLES AS NEUTRAL CARRIERS IN LIQUID MEMBRANE ION-SELECTIVE PH ELECTRODES
    CHOJNACKI, J
    BIERNAT, JF
    JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 1990, 277 (1-2): : 159 - 164
  • [3] DESIGN OF ION CARRIERS AND THEIR APPLICATION IN ION-SELECTIVE ELECTRODES
    PRETSCH, E
    AMMANN, D
    SIMON, W
    RESEARCH-DEVELOPMENT, 1974, 25 (03): : 20 - 24
  • [4] Sodium ion-selective electrodes based on silicone-rubber membranes covalently incorporating neutral carriers
    Tsujimura, Y
    Sunagawa, T
    Yokoyama, A
    Kimura, K
    ANALYST, 1996, 121 (11) : 1705 - 1709
  • [5] KINETIC CONSIDERATIONS OF ION SELECTIVITY COEFFICIENTS OF ION-SELECTIVE ELECTRODES BASED ON NEUTRAL CARRIERS
    OKADA, T
    HIRATANI, K
    SUGIHARA, H
    ANALYTICA CHIMICA ACTA, 1989, 221 (01) : 117 - 129
  • [6] Ion-selective electrodes based on molecular tweezer-type neutral carriers
    Shim, JH
    Jeong, IS
    Lee, MH
    Hong, HP
    On, JH
    Kim, KS
    Kim, HS
    Kim, BH
    Cha, GS
    Nam, H
    TALANTA, 2004, 63 (01) : 61 - 71
  • [7] NEUTRAL CARRIER BASED ION-SELECTIVE ELECTRODES
    AMMANN, D
    MORF, WE
    ANKER, P
    MEIER, PC
    PRETSCH, E
    SIMON, W
    ION-SELECTIVE ELECTRODE REVIEWS, 1983, 5 (01): : 3 - 92
  • [8] Polyvinyl chloride prasticized ion-selective electrodes based on the neutral carriers for determinations of novocaine
    Nemilova, MY
    Shvedene, NV
    Kochulina, OV
    Pronina, IA
    Pletniov, IV
    VESTNIK MOSKOVSKOGO UNIVERSITETA SERIYA 2 KHIMIYA, 1997, 38 (02): : 99 - 103
  • [9] Anion Selectivities of Dibutyltin Dihalides as Neutral Carriers for Ion-Selective Electrodes
    Takayo Moriuchi-Kawakami
    Hiroshi Aoto
    Naotake Takeda
    Keiichi Fujimori
    Yasuhiko Shibutani
    Analytical Sciences, 2000, 16 : 303 - 306
  • [10] Anion selectivities of dibutyltin dihalides as neutral carriers for ion-selective electrodes
    Moriuchi-Kawakami, T
    Aoto, H
    Takeda, N
    Fujimori, K
    Shibutani, Y
    ANALYTICAL SCIENCES, 2000, 16 (03) : 303 - 306