Compact Acetylene Detecting System Based on Tunable Diode Laser Absorption Spectroscopy

被引:3
|
作者
Jiang Lijun [1 ]
Qiu Xuanbing [1 ]
Zhou Qinghong [2 ,3 ]
Shao Ligang [1 ]
Yang Wen [1 ]
Wei Jilin [1 ]
Li Chuanliang [1 ]
Ma Weiguang [4 ]
机构
[1] Taiyuan Univ Sci & Technol, Appl Sci Coll, Taiyuan 030021, Shanxi, Peoples R China
[2] Southwest Univ Sci & Technol, Lab Extreme Condit Matter Properties, Res Ctr Laser Fus, China Acad Engn Phys, Mianyang 621010, Sichuan, Peoples R China
[3] Southwest Univ Sci & Technol, Sch Sci, Mianyang 621010, Sichuan, Peoples R China
[4] Shanxi Univ, Inst Laser Spect, State Key Lab Quantum Opt & Quantum Opt Devices, Taiyuan 030006, Shanxi, Peoples R China
关键词
spectroscopy; detecting instrument of acetylene; tunable diode laser absorption spectroscopy; wavelength modulation-multi-pass absorption technique; detecting sensitivity; compact;
D O I
10.3788/LOP55.033002
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
A detecting system of acetylene (C2H2) gas in near-infrared region is developed based on the tunable diode laser absorption spectroscopy. The wavelength modulation spectroscopy in combination of multi-pass absorption cell is used to improve detceting sensitivity of the system. To optimize detecting conditions of the system, spectral signals of standard C2H2 gas with volume fraction of 5X 10(-3) are detected under the conditions of pressure of 5.3-12.0 kPa and modulated amplitude of 0.010-0.035 V. The second harmonic signals of C2H2 with different concentrations arc detected at the total pressure of 10.7 kPa. To verify stability of the system, spectral signals for 60 s are acquired, and the most suitable detection time as well as the minimum detection limit is obtained by the Allan variance analysis. The results show that the signal intensity of the second harmonic signals is the greatest at pressure of 10.7 kPa and modulation amplitude of 0.030 V. The volume fraction of C2H2 is with a good linear relationship with the amplitude of the second harmonic signals, and the detecting deviation is less than +/- 2% in the range of volume fraction from 1 X 10(-6) to 5 X 10(-3). The most suitable detection time of the experimental system is 7.2 s, and the minimum detection limit is 2.8 X 10(-11). The instrument adopts a laser driver and a digital lock-in amplifier based on an embedded system, so the instrument has the characteristics of simple mechanical structure, small size, and easy integration with other systems. Therefore, the compact detecting system for C2H2 gas is suitable for industrial process and gas transmission.
引用
收藏
页数:6
相关论文
共 18 条
  • [1] Chen J Y, 2007, J ATMOSPHERIC ENV OP, V2, P146
  • [2] Cheng Y, 2013, ACTA OPT SINICA, V33
  • [3] GUO YQ, 2013, CHINESE J LASERS, V40
  • [4] Online Detection System on Acetylene with Tunable Diode Laser Absorption Spectroscopy Method
    He Qi-xin
    Liu Hui-fang
    Li Bin
    Pan Jiao-qing
    Wang Li-jun
    Zheng Chuan-tao
    Wang Yi-ding
    [J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2016, 36 (11) : 3501 - 3505
  • [5] The Development of Acetylene On-Line Monitoring Technology Based on Laser Absorption Spectrum
    He Ying
    Zhang Yu-jun
    Kan Rui-feng
    Xia Hui
    Wang Min
    Cui Yiao-juan
    Chen Jiu-ying
    Chen Dong
    Liu Wen-qing
    Liu Jian-guo
    [J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2008, 28 (10) : 2228 - 2231
  • [6] Theoretical and experimental study of measuring gas temperature in vacuum environment using tunable diode laser absorption spectroscopy
    Lan Li-Juan
    Ding Yan-Jun
    Jia Jun-Wei
    Du Yan-Jun
    Peng Zhi-Min
    [J]. ACTA PHYSICA SINICA, 2014, 63 (08)
  • [7] Pressure-Dependent Detection of Carbon Monoxide Employing Wavelength Modulation Spectroscopy Using a Herriott-Type Cell
    Li, Chuanliang
    Wu, Yingfa
    Qiu, Xuanbing
    Wei, Jilin
    Deng, Lunhua
    [J]. APPLIED SPECTROSCOPY, 2017, 71 (05) : 809 - 816
  • [8] Li Fei, 2011, Chinese Journal of Theoretical and Applied Mechanics, V43, P1061
  • [9] Ultra-high sensitive acetylene detection using quartz-enhanced photoacoustic spectroscopy with a fiber amplified diode laser and a 30.72 kHz quartz tuning fork
    Ma, Yufei
    He, Ying
    Zhang, Ligong
    Yu, Xin
    Zhang, Jingbo
    Sun, Rui
    Tittel, Frank K.
    [J]. APPLIED PHYSICS LETTERS, 2017, 110 (03)
  • [10] CO concentration and temperature sensor for combustion gases using quantum-cascade laser absorption near 4.7 μm
    Ren, W.
    Farooq, A.
    Davidson, D. F.
    Hanson, R. K.
    [J]. APPLIED PHYSICS B-LASERS AND OPTICS, 2012, 107 (03): : 849 - 860