DIRECT MICROTITRE PLATE ENZYME-IMMUNOASSAY OF FOLIC-ACID WITHOUT HEAT DENATURATION OF SERUM

被引:1
|
作者
DASSARMA, J
DUTTAGUPTA, C
ALI, E
DHAR, TK
机构
[1] INDIAN INST CHEM BIOL,CALCUTTA 700032,W BENGAL,INDIA
[2] INDIAN STAT INST,CALCUTTA 700035,W BENGAL,INDIA
关键词
FOLIC ACID; DIRECT ENZYME IMMUNOASSAY; MICROBIOLOGICAL ASSAY; FOLIC ACID-FREE SERUM;
D O I
暂无
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A new and simple method for enzyme immunoassay of folic acid (FA) has been developed, which does not require extraction or heat denaturation of serum. FA-free serum for standards was prepared by a new immunosorbent technique as conventional methods were unsuccessful. The detection limit of the assay is 0.05 ng/ml. Intra- and interassay variabilities ranged between 5-13.3%. Analytical recoveries obtained after spiking with different amounts of FA ranged between 93-110%. We eliminated the interference of endogenous folate binding protein - a major problem in direct FA assay by incubating serum samples (or standards) with FA-HRP conjugate in antibody coated plates at 50 degrees C. Comparison of our data with results obtained by microbiological assay and also by heating samples in alkaline buffer showed good correlation.