FLOW-INJECTION ANALYSIS OF MYOGLOBIN AND HEMOGLOBIN AT TOLUIDINE BLUE CHEMICALLY MODIFIED ELECTRODE

被引:17
|
作者
ZHOU, JX [1 ]
WANG, E [1 ]
机构
[1] CHINESE ACAD SCI,CHANGCHUN INST APPL CHEM,ELECTROANALYT CHEM,CHANGCHUN 130022,PEOPLES R CHINA
关键词
D O I
10.1002/elan.1140030311
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Electrodeposition of the phenothiazine mediator titrant toluidine blue onto a glassy carbon substrate at an appropriate potential was used to construct a toluidine blue chemically modified electrode (CME) exhibiting electrocatalytic reduction for myoglobin and hemoglobin. The CME catalyzed the hemoprotein electroreduction at the reduction potential of the mediator molecule. When the CME as used as a detector for flow injection analysis at a constant applied potential of -0.30 V vs. a saturated calomel electrode, it gave detection limits of 20 and 50 ng (1.2 and 0.78 pmol) injected myoglobin and hemoglobin, respectively, with a dynamic linear concentration range over 2 orders of magnitude. After a brief equilibration period, the CME retained nearly 90% of its initial myoglobin response over 8 hours of continuous exposure to the flow-through system.