SLURRY SAMPLE INTRODUCTION WITH MICROWAVE-INDUCED PLASMA ATOMIC EMISSION-SPECTROMETRY

被引:16
|
作者
MATUSIEWICZ, H [1 ]
STURGEON, RE [1 ]
机构
[1] NATL RES COUNCIL CANADA,INST ENVIRONM CHEM,OTTAWA K1A 0R9,ONTARIO,CANADA
关键词
Atomic emission spectroscopy (AES) - Microwave induced plasma (MIP) - Microwave plasma cavity - Powdered suspensions - Slurry nebulization - V-groove Babington nebulizer;
D O I
10.1016/0584-8547(93)80075-6
中图分类号
O433 [光谱学];
学科分类号
0703 ; 070302 ;
摘要
The successful direct introduction of aqueous slurry samples into a highly efficient TE101 microwave plasma has been demonstrated. Slurry samples from a spray chamber are fed directly into the cavity with no desolvation apparatus. A V-groove, clog-free Babington-type nebulizer was evaluated for use with high solids content solutions. Slurry concentrations up to 10% m v were used for the microwave induced plasma work with calibration by the standard additions method. Results are presented for the analysis of two NRCC Standard Reference Materials, i.e. TORT-1 (Lobster Hepatopancreas) and PACS-21 (Marine Sediment). Agreement between analytical results and certified values for the test elements Cd, Cu, Fe and Zn (in the range of 28-850 μg g) was good. No memory effects were evident and the nebulizer system had a rapid clean-out time. © 1993.
引用
收藏
页码:723 / 727
页数:5
相关论文
共 50 条